Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153741

    Validate with FDA (Live)

    Date Cleared
    2016-03-18

    (81 days)

    Product Code
    Regulation Number
    862.3870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Clungene Methamphetamine Tests are immunochromatographic assays for the qualitative determination of dmethamphetamine in human urine at cut-off concentration of 1000 ng/mL. The calibrator is d-methamphetamine. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format. The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

    Clungene Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-Nor-△9-Tetrahydrocannabinol-9-COOH in human urine at cut-off concentration of 50 ng/mL. The calibrator is 11-Nor-△9-Tetrahydrocannabinol-9-COOH. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format. The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

    Clungene Morphine Tests are immunochromatographic assays for the qualitative determination of Morphine in human urine at cut-off concentration of 300 ng/mL. The callbrator is Morphine. The tests are available in a Cassette format, a Easy Cup format, a Split Key Cup format and a Dip Card format. The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. This test is intended for over the counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling. For in vitro diagnostic use only.

    Device Description

    The CLUNGENE Methamphetamine Tests, CLUNGENE Morphine Tests, and CLUNGENE Marijuana Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Methamphetamine, Morphine and 11-Nor-A9-Tetrahydrocannabinol-9-COOH (target analytes) in human urine. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Clungene Methamphetamine, Morphine, and Marijuana Tests, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in a separate section with specific numerical thresholds for accuracy, sensitivity, or specificity that the device must meet for approval. Instead, it describes performance characteristics and then presents the results of those studies. The implicit acceptance criteria appear to be the demonstrated performance of the device showing appropriate qualitative detection relative to the cut-off concentrations when compared to GC/MS, and user readability for OTC use.

    Performance Characteristics Summary (Implicit Acceptance Criteria and Reported Performance)

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    PrecisionConsistent results across different lots and concentrations around cut-off.Across 3 lots and 9 different concentrations (-100% to +100% cut-off), the devices consistently showed 100% agreement for concentrations far from the cut-off. For samples at the cut-off, there was variability in positive/negative calls (e.g., for Methamphetamine Cassette, Lot 1: 22-/28+, Lot 2: 29-/21+, Lot 3: 29-/21+ at cut-off). This implies appropriate performance around the threshold.
    Cut-off AccuracySamples below -25% cut-off should be negative, and samples above +25% cut-off should be positive.All samples at and below -25% Cut-Off were negative. All samples at and above +25% Cut-Off were positive for Methamphetamine, Morphine, and Marijuana.
    InterferenceNo interference from common physiological/pathological substances or drugs at specified concentrations.Numerous substances (listed in the document) showed no interference at 100µg/mL.
    Specificity (Cross-reactivity)Limited cross-reactivity with structurally similar compounds; identified cross-reactive compounds should be listed with their reactivity percentage.Detailed tables provided showing cross-reactivity percentages for various related compounds (e.g., Methamphetamine: MDEA 5%, D/L-Methamphetamine 33%; Morphine: O6-Acetylmorphine 100%, Codeine 100%; Marijuana: 11-Hydroxy-Δ9-Tetrahydrocannabinol 1%).
    Effect of Urine Specific Gravity & pHNo significant impact on results within specified ranges (pH 4-9, SG 1.000-1.035).All samples within these ranges showed expected positive/negative results when spiked at +/- 25% cut-off.
    Method Comparison (Professional User)High agreement with GC/MS results, especially for samples far from the cut-off. Discordant results should primarily occur near the cut-off.Methamphetamine, Morphine, Marijuana: High agreement for negative, low negative, high positive samples. Discordant results were primarily observed for samples very close to the cut-off (e.g., GC/MS result 936 ng/mL for Methamphetamine vs. 1000 ng/mL cut-off).
    Lay-user Study (Accuracy)High percentage of correct results by lay users, particularly for samples clearly below or above the cut-off.For concentrations -100%, -75%, -50% cut-off, lay users showed 100% correct negative results. For +50%, +75% cut-off, lay users showed 100% correct positive results. For -25% and +25% cut-off, correctness ranged from 90-95%.
    Lay-user Study (Usability)Instructions should be easily understood by lay users.All lay users indicated the device instructions can be easily followed. Flesch-Kincaid Grade Level of 7.

    2. Sample Size Used for the Test Set and Data Provenance

    • Professional User (Method Comparison) Test Set:
      • Sample Size: 80 unaltered clinical urine samples per drug (40 negative, 40 positive).
      • Data Provenance: Retrospective, described as "in-house" studies using "unaltered clinical samples." The country of origin is not explicitly stated, but the manufacturer is based in China.
    • Lay-user Study Test Set:
      • Sample Size:
        • Participants: 1680 lay persons.
        • Samples: 1 device per participant, with one blind-labeled urine sample.
        • For each drug (Methamphetamine, Morphine, Marijuana) and each device format (Cassette, Dip Card, Split-Key Cup, Easy Cup), there were a total of 7 concentrations tested. For each concentration, 20 samples were used.
        • Total samples per drug (e.g., Methamphetamine): 7 concentrations * 20 samples/concentration = 140 samples.
        • Total samples for all 3 drugs across all 4 device formats: 3 drugs * 4 device formats * (7 concentrations * 20 samples/concentration) = 3 * 4 * 140 = 1680 samples. This matches the number of lay persons, implying each lay person tested one sample with one device.
      • Data Provenance: Prospective, as it involved lay persons testing samples. The study was performed "at three intended user sites," but the country of origin for these sites is not specified. Samples were prepared by spiking known drug concentrations into drug-free pooled urine.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Professional User (Method Comparison) Test Set:
      • Number of "Experts": Three "laboratory assistants" were involved in running the visual tests for comparison. Their qualifications are not specified beyond being "laboratory assistants."
      • Ground Truth: The ground truth for the clinical samples was established by Gas Chromatography/Mass Spectrometry (GC/MS) results. This is considered the gold standard for drug quantification.
    • Lay-user Study Test Set:
      • Number of Experts: No explicit "experts" were used to establish the ground truth for reading the device results in this specific study, as the lay users were the test subjects. The ground truth for the sample concentrations was established by GC/MS.

    4. Adjudication Method for the Test Set

    • Professional User (Method Comparison) Test Set: No formal adjudication method (like 2+1 or 3+1) is described for the visual interpretation by the three laboratory assistants. Each assistant's interpretation was recorded and compared to the GC/MS ground truth. Discordant results between the device and GC/MS were tabulated for each viewer individually.
    • Lay-user Study Test Set: No adjudication. Each lay user read their own device result.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such MRMC comparative effectiveness study was done. The device is a qualitative immunochromatographic assay for drug detection, not an AI interpretation system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a manual, visually interpreted immunochromatographic assay, not an algorithm or AI system. Its performance is inherently "standalone" in that it produces a visual result without external algorithm assistance for interpretation, but it always requires human observation and interpretation.

    7. The Type of Ground Truth Used

    • For all studies (Precision, Cut-off, Specificity, Interference, Method Comparison, Lay-user): The primary ground truth for the presence and concentration of drugs in urine samples was Gas Chromatography/Mass Spectrometry (GC/MS). This is considered a highly accurate and quantitative analytical method.

    8. The Sample Size for the Training Set

    This information is not provided. The document describes performance testing of the finished device. For immunochromatographic assays, there isn't typically a "training set" in the machine learning sense. Instead, development involves R&D to optimize reagents, membrane materials, and assay parameters. The performance studies described serve to validate the finalized design.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a "training set" in the AI sense is not relevant for this type of device. During the development and optimization of the assay, standard reference materials of known drug concentrations (confirmed by methods like GC/MS) would be used to evaluate and refine the test's linearity, sensitivity, and specificity.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1