510(k) Data Aggregation

    K Number
    K153703

    Validate with FDA (Live)

    Device Name
    Guiding Catheter
    Date Cleared
    2016-05-12

    (141 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Guiding Catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vasculature.

    Device Description

    The TXM Guiding Catheter is a guiding catheter, for application in intravascular vessels. It consists of a lubricous inner liner made from Teflon, a stainless steel braid over the inner liner. The outer shaft consists of durometer hardness and different lengths of colored Polyether block amide. A Polycarbonate hub is attached to the proximal end of the Guiding Catheter. A strain relief is placed at the proximal end of the Guiding Catheter and over the distal end of the device shafts will have an outer diameter of 5 French sizes (0.070") and 6 French sizes (0.082) respectively for each catheter. The 5 French sizes will have an inner diameter of 0.058" and the 6 French sizes will have an inner diameter of 0.070". The device is available in length of 100cm. The device is supplied sterile and is intended for single use.

    AI/ML Overview

    This document is a 510(k) premarket notification for a guiding catheter, indicating its substantial equivalence to a predicate device. It doesn't describe the evaluation of an AI/ML powered device, but rather a traditional medical device (catheter). Therefore, most of the requested information regarding AI/ML device evaluation (like sample sizes for test/training sets, expert ground truth, MRMC studies, standalone performance) is not applicable or cannot be extracted from this document.

    However, I can extract the acceptance criteria and results for the bench testing and biocompatibility testing as presented for this traditional medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    For this medical device, the "acceptance criteria" are implied by meeting the specified standards or internal test methods, and the "reported device performance" is a simple "Pass" or "Fail" against those criteria.

    Test NameApplicable Standard or Internal Test MethodAcceptance Criteria (Implied by standard and "Pass" result)Reported Device Performance
    Air LeakISO 10555Meet ISO 10555 requirementsPass
    In-Vitro Track ForceInternal Test MethodMeet internal method requirementsPass
    StiffnessInternal Test MethodMeet internal method requirementsPass
    Tensile StrengthISO 10555Meet ISO 10555 requirementsPass
    Gravity Flow RateISO 10555Meet ISO 10555 requirementsPass
    Liquid LeakISO 10555Meet ISO 10555 requirementsPass
    Power Injector (Static Burst Pressure)ISO 10555Meet ISO 10555 requirementsPass
    Tip Shape RetentionInternal Test MethodMeet internal method requirementsPass
    Dimensional & Physical AttributesISO 10555Meet ISO 10555 requirementsPass
    Kink TestingInternal Test MethodMeet internal method requirementsPass
    Buckling TestInternal Test MethodMeet internal method requirementsPass
    Corrosion ResistanceISO 10555Meet ISO 10555 requirementsPass
    RadiopacityASTM-F640-12Meet ASTM-F640-12 requirementsPass
    Torque ResponseInternal Test MethodMeet internal method requirementsPass
    Torque to FailureInternal Test MethodMeet internal method requirementsPass
    OvalizationInternal Test MethodMeet internal method requirementsPass
    Simulated UseInternal Test MethodMeet internal method requirementsPass
    Seal Strength (Pouch)ASTM F88/F88M-09Meet ASTM F88/F88M-09 requirementsPass
    Dye Penetration TestASTM-F1929-12Meet ASTM-F1929-12 requirementsPass
    Female Luer VerificationISO 594Meet ISO 594 requirementsPass
    Shipping and TransportationISTA 3AMeet ISTA 3A requirementsPass
    Accelerated AgingASTM F1980-07Meet ASTM F1980-07 requirementsPass
    Biocompatibility Tests
    CytotoxicityISO10993-5Meet ISO10993-5 requirementsPass
    IrritationISO10993-10Meet ISO10993-10 requirementsPass
    SensitizationISO10993-10Meet ISO10993-10 requirementsPass
    Systemic ToxicityISO10993-11Meet ISO10993-11 requirementsPass
    PyrogenicityUSP<151>, ISO10993-11, USP <85>Meet specified USP and ISO requirementsPass
    HemocompatibilityISO10993-4Meet ISO10993-4 requirementsPass
    Toxicological HazardISO10993-17Meet ISO10993-17 requirements (Risk Assessment)Pass
    Sterilization Test
    Ethylene Oxide GasRelevant validation standards (implicit)Sterility Assurance Level (SAL) of 10-6Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document refers to "non-aged devices (T=0)" and "devices subject to 3 years of accelerated aging (T=3)" for bench testing, but the specific number of devices tested for each test is not detailed. The data provenance is also not specified, as this is a device submission focused on bench and biocompatibility testing, not clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or expert evaluation for establishing ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or expert adjudication for test sets.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's evaluation is primarily based on meeting the specifications outlined in recognized international and national standards (e.g., ISO 10555, ASTM F640-12, ISO 10993 series, USP) and internal test methods. For example, for tensile strength, the ground truth is simply whether the device withstood the specified force according to the standard. For biocompatibility, it's whether the biological responses fall within acceptable limits defined by the ISO 10993 series.

    8. The sample size for the training set

    Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or training sets.

    9. How the ground truth for the training set was established

    Not applicable. This is a traditional medical device (catheter) and does not involve AI/ML or training sets.