Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMPROSAFE® Blood Collection Sets with Pre-attached to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROSAFE® Multi-Sample Needles are intended to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROVACUTER® Blood Collection Sets are intended to be used with evacuated blood collection tube for the collection of venous blood.
    The IMPROSAFE® Blood Collection Sets are intended to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROVACUTER® Blood Collection Sets with Pre-attached holder are intended to be used with evacuated blood collection tube for the collection of venous blood.
    The IMPROSAFE® Multi-Sample Needles (Flashback) are intended to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROVACUTER® Multi-Sample Needles are used with holder and vacuum blood collection tube for the venous blood collection.

    Device Description

    The proposed devices are blood collection sets form a channel between patient's vein and the evacuated blood collection tube intended for collection of blood. They are available in different configurations as follows:
    IMPROSAFE® Blood Collection Set with Pre-attached holder: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, Pre-attached Holder, Safety Mechanism, Visible Flash Back
    IMPROSAFE® Blood Collection Set: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, N.A., Safety Mechanism, Visible Flash Back
    IMPROVACUTER® Blood Collection Set with Pre-attached Holder: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, Pre-attached Holder, N.A., Visible Flash Back
    IMPROVACUTER® Blood Collection Set: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, N.A., N.A., Visible Flash Back
    IMPROSAFE® Multi Sample Needle: Patient-End Needle, Non-Patient End Needle, N.A., Pre-attached Holder, Safety Mechanism, N.A.
    IMPROSAFE® Multi Sample Needle (Flashback): Patient-End Needle, Non-Patient End Needle, N.A., Pre-attached Holder, Safety Mechanism, Visible Flash Back
    IMPROVACUTER® Multi Sample Needle: Patient-End Needle, Non-Patient End Needle, N.A., N.A., N.A., N.A.

    AI/ML Overview

    The provided text is a 510(k) Summary for several blood collection devices. It aims to demonstrate substantial equivalence to previously cleared predicate devices, rather than proving a device meets specific acceptance criteria through an independent clinical study for AI software.

    Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, type of ground truth, training set information) are not applicable or not provided in this document, as it pertains to medical hardware (needles and collection sets) and not AI software.

    However, I can extract information related to non-clinical tests and the general equivalence claims.

    Here's the breakdown of the information provided in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in a quantitative, pre-defined manner for performance metrics. Instead, it refers to conformity with established international and ASTM standards. The "reported device performance" is essentially that the device "conforms" to these standards.

    Acceptance Criteria (Implied by Standards)Reported Device Performance (Conformity)
    Material (Stainless steel needle tubing): Conforms to ISO 9626:1991+AMENDMENT 1 2001Conforms to ISO 9626:1991 AMD 2001 (and ISO 7864:1993 for sterile hypodermic needles)
    Sterile Hypodermic Needles: Conforms to ISO 7864: 1993Conforms to ISO 7864: 1993 (and ISO 9626:1991 AMD 2001 for needle tubing)
    Ethylene Oxide Sterilization residuals: Conforms to ISO 10993-7:2008Conforms with ISO 10993
    Seal Strength of Flexible Barrier Materials: Conforms to ASTM F 88/F88M-09Conforms to ASTM F 88/F88M-09
    Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration: Conforms to ASTM F1929-12Conforms to ASTM F1929-12
    Bacterial Endotoxins Limit: Conforms to USP38-NF33 <85>Conforms to USP38-NF33 <85>
    Tests for in vitro cytotoxicity: Conforms to ISO 10993-5:2009Conform with ISO 10993
    Tests for irritation and delayed-type hypersensitivity: Conforms to ISO 10993-10:2010Conform with ISO 10993
    Tests for systemic toxicity: Conforms to ISO 10993-11:2006Conform with ISO 10993
    Assessment of hemolytic properties of materials: Conforms to ASTM F 756-13Conform to ASTM F 756-13
    Sharps Injury Prevention Features: Conforms to Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention FeaturesConforms to Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention Features
    Labeling Requirements: Conforms to 21 CFR Part 801Conform with 21 CFR Part 801
    Sterility: EO sterilizedEO sterilized
    Single Use: Single UseSingle Use

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not provided. This document does not detail specific sample sizes for non-clinical tests or data provenance. The tests are general compliance tests for medical devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable/Not provided. This is not an AI device or a diagnostic device relying on expert interpretation for "ground truth." The tests are for physical/biological properties and compliance with engineering standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not provided. This concept is not relevant to the non-clinical testing of blood collection devices described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a blood collection set/needle, not an AI-powered diagnostic tool, so MRMC studies are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. The "ground truth" for this type of device is compliance with specific material, sterility, performance (e.g., seal strength, endotoxin limits), and biocompatibility standards, as measured by established laboratory test methods.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not a machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no "training set" or corresponding ground truth for this type of device.

    Summary of the Study (Non-Clinical Test Conclusion) from the document:

    The non-clinical tests were conducted to verify that the proposed devices met all design specifications and were Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:

    • ISO 9626:1991+AMENDMENT 1 2001 (Stainless steel needle tubing)
    • ISO 7864: 1993 (Sterile hypodermic needles for single use)
    • ISO 10993-7:2008 (Biological evaluation - Ethylene oxide sterilization residuals)
    • ASTM F 88/F88M-09 (Seal strength of flexible barrier materials)
    • ASTM F1929-12 (Detecting seal leaks in porous medical packaging by dye penetration)
    • USP38-NF33 <85> (Bacterial Endotoxins Limit)
    • ISO 10993-5:2009 (Biological evaluation - Tests for in vitro cytotoxicity)
    • ISO 10993-10:2010 (Biological evaluation - Test for irritation and delayed-type hypersensitivity)
    • ISO 10993-11:2006 (Biological evaluation - Tests for systemic toxicity)
    • ASTM F 756-13 (Assessment of hemolytic properties of materials)
    • Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention Features

    Clinical Test Conclusion:
    "No clinical study is included in this submission."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1