510(k) Data Aggregation

    K Number
    K153233

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-01-06

    (58 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Galaxy UNYCO System is intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.

    Temporary stabilization of the tibia and foot in conditions and procedures, such as:

    • · comminuted open or closed tibial fractures
    • · polytrauma patient
    • · damage control orthopedics for fractures with severe soft tissue injuries
    • peri-prosthetic or peri-implant fractures
    • · joint dislocations, intra- and extra-articular injuries where spanning fixation is needed
    • · intra-operative fracture reduction
    • · intermediate stabilization in staged surgery
    • · infected non-union pending second stage treatment bone-loss or other reconstructive procedures

    The Galaxy UNYCO System is compatible with Galaxy Fixation System and bicortical screws. Galaxy Fixation System and bicortical screws must be used when Galaxy UNYCO is not indicated or available. The product is indicated for non-weight-bearing use.

    Device Description

    The Orthofix Galaxy UNYCO System consists of a series of sterile kits that include UNYCO Cancellous Screws, Large Multiscrew Clamps for UNYCO Screws, Rods Ø 12mm, a radiolucent Foot Unit and specific application tools. External fixation systems are modular, therefore different frame configurations are possible. The Orthofix components in the Galaxy UNYCO System are not intended to replace normal healthy bone or to withstand the stresses of weight bearing.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Galaxy UNYCO System (K153233)

    The provided documentation describes the acceptance criteria and study findings for the Orthofix Galaxy UNYCO System, a device intended for temporary bone stabilization.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device PerformanceComments
    Mechanical PerformanceMechanical strength as per ASTM F 1541-02Found to be equivalent or better than the predicate devices.Testing met or exceeded requirements.
    Material CompatibilityNot explicitly stated as a numerical criterion, but implied by equivalence to predicate.Utilizes similar materials to predicate devices.Part of establishing substantial equivalence.
    Intended UseTemporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.Device is intended for these uses.Confirmed through indications for use.
    SafetyRisk Management Plan, adherence to applicable standards.Potential hazards evaluated and controlled; testing met or exceeded requirements.No clinical data was deemed necessary for safety and effectiveness.
    MRI CompatibilityAdherence to ASTM F2182 "Standard test method for measurement of radio frequency induced heating near passive implants during magnetic resonance imaging"MRI compatibility testing was conducted per ASTM F2182.Specific results not detailed, but testing was performed.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily relies on mechanical testing and material comparison for demonstrating substantial equivalence. It does not mention a "test set" in the context of clinical data or patient data.

    • Test Set (for mechanical and MRI testing): The specific number of devices/components tested for mechanical performance (ASTM F 1541) and MRI compatibility (ASTM F2182) is not explicitly stated in the provided summary.
    • Data Provenance: The studies are described as non-clinical technical assessments (mechanical and MRI compatibility testing) rather than clinical studies with patient data. Therefore, there is no mention of country of origin or retrospective/prospective nature regarding patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not applicable as the studies were non-clinical mechanical and MRI compatibility tests. Ground truth in this context would be defined by the technical specifications of the standards (ASTM F 1541, ASTM F2182).

    4. Adjudication Method for the Test Set:

    This information is not applicable as the studies were non-clinical mechanical and MRI compatibility tests. Adjudication methods are typically associated with human reviewer assessments in clinical or diagnostic studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was mentioned. The submission explicitly states: "Clinical data was not needed to support the safety and effectiveness of the Subject Device."

    6. Standalone Performance (Algorithm Only):

    No standalone performance study for an algorithm was mentioned. The device is a physical external fixation system, not a software algorithm.

    7. Type of Ground Truth Used:

    The "ground truth" for the evaluation of the Galaxy UNYCO System was based on:

    • Engineering Standards: Specifically, the requirements and methodologies outlined in ASTM F 1541 for mechanical performance and ASTM F2182 for MRI compatibility.
    • Predicate Device Characteristics: Comparison to the established design, materials, and mechanical performance of legally marketed predicate devices (Synthes Large External Fixation, MR Conditional (K082650) and Orthofix Galaxy Unyco Diaphyseal Tibia Kit (K142052)).

    8. Sample Size for the Training Set:

    This information is not applicable. The device is a physical medical device, not an algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reasons stated in point 8.