Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153016

    Validate with FDA (Live)

    Date Cleared
    2016-10-20

    (371 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    aoni Natural Rubber Latex Condom [Smooth/Dotted/Combined]

    a) Smooth 52mm/54mm/55mm natural

    b) Dotted 52mm natural

    c) Combined [smooth+girdling] 52mm natural

    d) Combined [dotted+girdling] 52mm natural

    The Condom is used for contraception and for prophylactic purpose ( to help prevent pregnancy and the transmission of sexually transmitted infections.)

    Device Description

    aoni Natural Rubber Latex Condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. These condoms conform to ISO 4074:2002 Natural Rubber Latex Condoms – Requirements and Test Methods, ASTM D3492-14 Standard Specification for Rubber Contraceptives (Male Condom), and ISO 10993-1 Biological Evaluation of Medical Devices - Part 1(2009). These condoms have repeated contact potential for a duration of <24 hours and contact the mucosal membrane. They are for home use.

    aoni Natural Rubber Latex Condom include four types as following:

    [Smooth]: aoni Natural rubber latex condom [Smooth] are straight walled bulbous ended with smooth surface, having a nominal length of 180mm, a nominal width of 52mm/54mm/55mm, a nominal thickness of 0.036-0.080mm, Natural color

    [Dotted]: aoni Natural rubber latex condom [Dotted] are straight walled bulbous ended with Dotted surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

    [Combined: (smooth+girdling)]: aoni Natural rubber latex condom [Combined: (smooth+girdling)] are girdling walled bulbous ended with smooth surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

    [Combined: (dotted+girdling)]: aoni Natural rubber latex condom [Combined: (dotted+girdling)] are girdling walled bulbous ended with dotted surface, having a nominal length of 180mm, a nominal width of 52mm, a nominal thickness of 0.036-0.080mm, Natural color.

    AI/ML Overview

    Acceptance Criteria and Device Performance for aoni Natural Rubber Latex Condom

    This document summarizes the acceptance criteria and reported device performance for the "aoni Natural Rubber Latex Condom," as described in the provided 510(k) summary (K153016).


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the aoni Natural Rubber Latex Condom are based on recognized international and national standards for rubber contraceptives and biocompatibility. The reported device performance indicates that the product met these specifications.

    Table 1: Acceptance Criteria and Reported Device Performance

    Parameter / TestAcceptance CriteriaReported Device PerformanceStandard(s) Applied
    Physical Performance Tests
    Water Leakage (Freedom from Holes)Met specifications of ISO 4074:2002 and ASTM D3492-14"met the specifications"ISO 4074:2002, ASTM D3492-14
    Burst Volume and PressureMet specifications of ISO 4074:2002 and ASTM D3492-14"met the specifications"ISO 4074:2002, ASTM D3492-14
    Dimensions (Length, Width, Thickness)Within specified ranges (e.g., Nominal Length 180mm, Width 52/54/55mm, Thickness 0.036-0.080mm)Conformed to nominal dimensions and thickness rangesISO 4074:2002, ASTM D3492-14
    Biocompatibility Tests
    CytotoxicityMet acceptance criteria of ISO 10993-5:2009"passed each biocompatibility test"ISO 10993-5:2009
    Skin IrritationMet acceptance criteria of ISO 10993-10:2010"passed each biocompatibility test"ISO 10993-10:2010
    Vaginal IrritationMet acceptance criteria of ISO 10993-10:2010"passed each biocompatibility test"ISO 10993-10:2010
    SensitizationMet acceptance criteria of ISO 10993-10:2010"passed each biocompatibility test"ISO 10993-10:2010
    Acute Systemic ToxicityMet acceptance criteria of ISO 10993-11:2006"passed each biocompatibility test"ISO 10993-11:2006
    Shelf Life5 years (60 months)"has a five years (60 month) shelf life based on the results of real time storage and testing"21 CFR 801.435, ISO 4074:2012 (Note: Document specifies 2012, while performance tests mention 2002 version for ISO 4074)

    Note on Standards: The document refers to ISO 4074:2002 for performance testing and ISO 4074:2012 for shelf life. Both are relevant versions of the standard for natural rubber latex condoms.


    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical testing for performance and biocompatibility. It does not refer to a "test set" in the context of clinical studies with human subjects.

    • Sample Size for performance and biocompatibility testing: The specific sample sizes for each non-clinical test (e.g., number of condoms for burst testing, number of animals for toxicity tests) are not explicitly stated in the provided 510(k) summary. The summary only broadly states that "Each variant of aoni Natural Rubber Latex Condom was tested".
    • Data Provenance: The testing was conducted by Guangzhou Daming United Rubber Products Ltd. or its contracted laboratories. The information does not specify the country of origin for the data for individual tests beyond the manufacturer's location in Guangzhou, China. All testing appears to be retrospective (i.e., conducted on finished products to demonstrate compliance with standards) rather than prospective clinical trials.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This section is not applicable as the evaluation described is for ex-vivo device performance and biocompatibility testing against established standards, not a clinical study requiring expert assessment of "ground truth" for a test set (e.g., diagnosis of disease). The "ground truth" in this context is defined by the technical specifications and requirements outlined in the referenced ISO and ASTM standards.


    4. Adjudication Method for the Test Set

    This section is not applicable. The evaluation described involves objective physical and chemical testing against pre-defined performance criteria in established standards, not interpretation by human adjudicators.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not performed as this device (a condom) is not a diagnostic device that relies on human interpretation for an outcome. The evaluation is based on objective physical and chemical tests.


    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    This section is not applicable. The device is a physical medical device (condom), not an algorithm or AI-driven system. Therefore, standalone algorithm performance is not relevant.


    7. Type of Ground Truth Used

    The "ground truth" for demonstrating device compliance is defined by:

    • Performance Specifications: Quantitative and qualitative requirements outlined in ISO 4074: Natural Rubber Latex Condoms – Requirements and Test Methods and ASTM D3492-14 Standard Specification for Rubber Contraceptives (Male Condom). These standards define acceptable ranges for parameters like burst volume and pressure, freedom from holes, and dimensions.
    • Biocompatibility Standards: Criteria for various biological responses (cytotoxicity, irritation, sensitization, systemic toxicity) outlined in ISO 10993-1: Biological Evaluation of Medical Devices - Part 1, and its sub-parts (e.g., ISO 10993-5, ISO 10993-10, ISO 10993-11).

    In essence, the ground truth is derived from established scientific and engineering standards and accepted laboratory test methods.


    8. Sample Size for the Training Set

    This section is not applicable. As a physical device, there is no "training set" in the context of machine learning or AI models. The manufacturing process is validated (and products routinely tested) to ensure consistent quality and adherence to specifications.


    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the reasons stated in point 8. The "ground truth" for manufacturing a condom to meet specifications is established through the development and validation of manufacturing processes to consistently produce a product that passes the non-clinical tests detailed above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1