510(k) Data Aggregation

    K Number
    K152465

    Validate with FDA (Live)

    Date Cleared
    2015-10-29

    (59 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes.

    Device Description

    The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, pediatric and adult sizes.

    The proposed device is disposable blood pressure cuff, which contains Cuff with bladder and Air Hose. Air hose has single tube and double tube.

    During the operation process, the integrated inflatable bladder will wrapped around the patient's limb and secured by hook and loop closure, and the air hose will connected to the noninvasive blood pressure measurement systems.

    The disposable blood pressure cuff is single use device, and which is made of non-woven fabrics & PVC (Cuff) and PVC (Air Hose), thereinto, the non-woven fabrics & PVC (Cuff) is the material used to contacting with the patient.

    The disposable blood pressure cuff has 26 models with different size for different population with different arm size, and the 13 models of 26 are use single tube of air hose and other 13 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.

    The disposable blood pressure cuff is provided as non-sterile.

    AI/ML Overview

    The provided document describes the 510(k) premarket notification for a Disposable Blood Pressure Cuff (K152465) by Shenzhen Vistar Medical Supplies Co., Ltd. The submission aims to demonstrate substantial equivalence to a legally marketed predicate device (K151290).

    Here's an analysis of the acceptance criteria and the study that proves the device meets these criteria, based on the provided text:

    Key Takeaways:

    • This submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel clinical effectiveness.
    • The primary means of demonstrating compliance is through non-clinical testing against recognized industry standards and a comparison of specifications with the predicate device.
    • No clinical study was included in this submission. Therefore, there is no information about human reader performance (MRMC studies) or standalone algorithm performance, as these are typically part of clinical trials for AI/software devices.
    • The device is a physical medical device (blood pressure cuff), not a software or AI-driven diagnostic tool. Thus, concepts like "ground truth" in the AI sense, expert panels for ground truth establishment, or training sets are not directly applicable in the way they would be for an AI device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from recognized standards and comparison with the predicate device. The performance is reported as compliance with these standards or equivalence to the predicate.

    Acceptance Criteria / SpecificationReported Device Performance (Proposed Device)
    Intended Use"The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes." (Identical to Predicate)
    Patient PopulationsAdults/Pediatrics (Identical to Predicate)
    Tube ConfigurationOne or two tube (Identical to Predicate)
    Principles of Operation"Bladder is wrapped around the patient's limb and secured by hook and loop closure Air hose is connected to the noninvasive blood pressure measurement systems" (Identical to Predicate)
    SizeConform to AHA bladder sizes recommendations. Has similar sizes and more size specifications than the predicate for different patient arm circumferences. Neonatal 1 (3-6 cm) Neonatal 2 (4-8 cm) Neonatal 3 (6-11 cm) Neonatal 4 (7-13 cm) Neonatal 5 (8-15 cm) Infant (9-14.8 cm) Child (13.8-21.5 cm) Small Adult (20.5-28.5 cm) Adult (27.5-36.5 cm) Adult Long (27.5-36.5 cm) Large Adult (35.5-46 cm) Large Adult (35.5 - 46 cm) Thigh (45 - 56.5 cm)
    Single UseYes (Identical to Predicate)
    SterileNo (Identical to Predicate)
    Pressure Range0-300 mmHg (Identical to Predicate)
    Max. Pressure400 mmHg (Predicate: >350 mmHg)
    Tubing Size (mm)OD = 8.0; ID=4.0; L=200; (Identical to Predicate)
    Max. Leakage< 4mm Hg/ min. (Identical to Predicate)
    Materialnon-woven fabrics & PVC (Cuff) and PVC (Air Hose) (Identical to Predicate)
    Biocompatibility - CytotoxicityUnder the conditions of the study, not cytotoxicity effect (Complies with ISO 10993-5)
    Biocompatibility - IrritationUnder the conditions of the study, not an irritant (Complies with ISO 10993-10)
    Biocompatibility - SensitizationUnder conditions of the study, not a sensitizer. (Not explicitly stated for predicate in comparison table, but overall concluded SE)
    Performance StandardComplies with ISO 81060-1:2012 Non-Invasive Sphygmomanometers - Part 1: Requirements And Test Methods For Non-Automated Measurement Type

    2. Sample Size Used for the Test Set and the Data Provenance

    Since this is a non-clinical device and no human clinical trials were conducted for this 510(k) submission, the concept of a "test set" in the context of patient data is not applicable. The testing was performed on the device itself (e.g., individual cuffs, materials).

    • Sample Size for Test Set: Not specified in terms of number of units tested, but physical units of the device were subjected to laboratory tests.
    • Data Provenance: Not applicable in the context of patient data. The tests were laboratory-based (e.g., material testing, pressure testing) conducted by the manufacturer. The company is based in China: Shenzhen Vistar Medical Supplies Co., Ltd. (Shenzhen, China). The data is retrospective in the sense that it's generated for the submission but not from a prospective clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This question is not directly applicable to this type of device submission.

    • "Ground truth" for a physical device is established by meeting engineering specifications and performance standards. These are defined by regulatory bodies and industry standards (e.g., ISO, FDA guidance documents). The "experts" are the engineers and scientists who developed and validated the test methods to ensure compliance with these standards, not external clinicians establishing diagnostic ground truth from patient data.
    • No specific number or qualifications of experts are mentioned for establishing ground truth, as it's not a diagnostic AI device requiring expert consensus on images or patient outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies, especially for evaluating diagnostic performance where disagreements among readers need resolution to establish a reference standard (ground truth). This submission relies on objective laboratory tests and engineering comparisons.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. This is a blood pressure cuff, not an AI diagnostic tool that assists human readers.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a blood pressure cuff, which is a physical accessory for a non-invasive blood pressure measurement system. It does not contain an algorithm nor is it intended for standalone diagnostic performance in the way an AI software device would be.


    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on engineering specifications, industry standards, and the performance characteristics of the predicate device.

    • Engineering Specifications: E.g., pressure range, maximum leakage, tubing dimensions.
    • Biocompatibility Standards: ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Irritation and Skin Sensitization). Compliance with these standards is considered the "ground truth" for material safety.
    • Performance Standards: ISO 81060-1:2012 (Non-Invasive Sphygmomanometers - Part 1: Requirements And Test Methods For Non-Automated Measurement Type). Compliance with this standard is the "ground truth" for the cuff's functional performance.
    • Predicate Device Characteristics: The characteristics of the legally marketed predicate device (K151290) serve as a de facto "ground truth" for substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not a machine learning model. There is no "training set."


    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for a physical device.