Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151963

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-07-31

    (15 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartSite® Vented Vial Access Device is intended for use by healthcare professionals in a wide variety of healthcare environments including hospitals, healthcare facilities, and pharmacies for reconstitution or dispensing of medication. The SmartSite® Vented Vial Access Device is indicated for use with rubber-stopper medication vials for reconstitution or dispensing of medications, including chemotherapy agents.

    Device Description

    The 20mm SmartSite® Vented Vial Access Device is a stand alone, sterile, single-use, disposable device which permits access to a medication vial without the use of a needle. It consists of a spike, locking shroud, hydrophobic filter, and SmartSite® needle-free access valve for Luer access.

    The SmartSite® Vented Vial Access Device is microbiologically closed. When used in a USP<797> compliant pharmaceutical compounding and storage environment, the SmartSite® Vented Vial Access Device is capable of maintaining the sterility of vial medications for up to7 days.

    AI/ML Overview

    This document is a 510(k) summary for the SmartSite® Vented Vial Access Device, which is a medical device for accessing medication vials. The summary primarily focuses on demonstrating substantial equivalence to a predicate device and does not involve AI/ML performance evaluation. Therefore, many of the requested categories related to AI/ML studies are not applicable.

    Here's the information extracted and categorized based on your request, with "Not Applicable" for sections where the document does not provide the requested AI/ML related details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" against which a device's AI/ML performance is measured, as this is a traditional medical device submission, not an AI/ML one. However, it lists various tests performed to demonstrate the device's functional and safety characteristics, primarily against a predicate device.

    Test Performed (Performance Characteristic)Device TestedResultNote/Interpretation
    Attachment ForceSubject & PredicateMet specified requirementsDemonstrated equivalency during testing.
    Filter Misuse PressureSubject & PredicateMet specified requirementsDemonstrated equivalency during testing.
    Residual Fluid VolumeSubject & PredicateMet specified requirementsDemonstrated equivalency during testing.
    Flow RateSubject & PredicateMet specified requirementsDemonstrated equivalency during testing.
    Tests Performed on Subject Device Only:
    Packaging maintains sterile barrierSubjectMet specified requirementsPerformed to establish safety and effectiveness.
    Device packaging opens with gloved handsSubjectMet specified requirementsPerformed to establish usability.
    Packaging materials are radiation sterilizableSubjectMet specified requirementsPerformed to establish manufacturing compatibility.
    Device materials are radiation sterilizableSubjectMet specified requirementsPerformed to establish manufacturing compatibility.
    Vial septum visibilitySubjectMet specified requirementsPerformed to establish usability.
    Device usable with gloved handsSubjectMet specified requirementsPerformed to establish usability.
    Priming VolumeSubjectMet specified requirementsPerformed to establish functional characteristic.
    SmartSite bond strengthSubjectMet specified requirementsPerformed to establish structural integrity.
    Stopper coringSubjectMet specified requirementsPerformed to establish functional characteristic and prevent contamination.
    Device locks with >= 2 fingersSubjectMet specified requirementsPerformed to establish usability and security.
    Spike breakageSubjectMet specified requirementsPerformed to establish durability.
    Shroud breakageSubjectMet specified requirementsPerformed to establish durability.
    Vertical and horizontal detachment forceSubjectMet specified requirementsPerformed to establish security of attachment.
    Negative pressure leakageSubjectMet specified requirementsPerformed to establish functional integrity.
    Filter prevents >=99.9%SubjectMet specified requirementsPerformed to establish filtration effectiveness.
    Shelf life (accelerated)SubjectMet specified requirementsPerformed to establish product lifespan.
    Storage temperatureSubjectMet specified requirementsPerformed to establish storage conditions.
    Biocompatibility ISO 10993 (Cytotoxicity, Intracutaneous Reactivity, Sensitization, Acute Systemic Toxicity, Hemocompatibility)SubjectMet specified requirementsPerformed to establish biological safety.
    Chemotherapy and Hazardous Drug CompatibilitySubjectMet specified requirementsPerformed to establish chemical compatibility and safety.
    Tests Leveraged from Predicate (SmartSite® needle free valve):
    Needle-free Valve Activations >45Subject & PredicatePreviously evaluated & clearedDemonstrated robust performance of the shared component.
    Needle-free Valve ISO-594 Luer CompatibilitySubject & PredicatePreviously evaluated & clearedDemonstrated standard compatibility of the shared component.
    Microbial ingressSubject & PredicatePreviously evaluated & clearedDemonstrated integrity of the shared component.
    LeakageSubject & PredicatePreviously evaluated & clearedDemonstrated integrity of the shared component.
    Additional Test for Subject Device Specifics:
    Maintains drug sterility for 7 daysSubjectMet specified requirementsAddresses the "microbiologically closed" characteristic of the subject device.

    2. Sample size used for the test set and the data provenance

    Not Applicable (This is a traditional medical device; no "test set" in the context of AI/ML data). The document refers to "tests conducted" on the device itself and the predicate device. The specific sample sizes for these engineering and biological tests are not detailed in this summary. Data provenance is not described as it relates to physical device testing rather than clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not Applicable (No AI/ML system involved in a clinical study to establish ground truth).

    4. Adjudication method for the test set

    Not Applicable (No AI/ML system involved in a clinical study requiring adjudication).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable (This device is a physical medical device, not an AI/ML diagnostic or assistive tool).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not Applicable (No AI/ML algorithm involved).

    7. The type of ground truth used

    Not Applicable (No AI/ML system requiring a ground truth in the context of expert consensus, pathology, or outcomes data). The "ground truth" for this device would be its adherence to established engineering, safety, and performance specifications as measured by laboratory testing and previous FDA clearances for its components.

    8. The sample size for the training set

    Not Applicable (No AI/ML model with a training set).

    9. How the ground truth for the training set was established

    Not Applicable (No AI/ML model with a training set).