Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151912

    Validate with FDA (Live)

    Date Cleared
    2016-02-03

    (205 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AIR-FLOW handy 3.0 PLUS Dental Handpiece is a dental handpiece intended for use in the cleaning and polishing of teeth by the projection of a mixture of water, air, and EMS prophylaxis powder onto the tooth surface. The device removes soft deposits and areas of discoloration and can be used to prepare teeth for dental procedures such as the placement of composite fillings, porcelain inlays, and laminate veneers. The device can be used to clean implant abutments and to clean teeth prior to treatments such as shade matching, fluoridation, and bleaching. The device can also be used to degrease crowns and bridges prior to placement and clean fixed bands and brackets on orthodontic appliances.

    The AIR-FLOW handy 3.0 PLUS is intended for patients suffering from periodontal disease.

    The AIR-FLOW handy 3.0 PLUS is indicated for the non-surgical removal of subgingival plaque in pockets up to 4 mm after initial periodontal treatment.

    Device Description

    The proposed AIR-FLOW handy 3.0 PLUS device is similar in design and materials to the predicates AIR-FLOW handy 3.0 (K132478) and AIR-FLOW handy 3.0 PERIO (K132480). The proposed AIR-FLOW handy 3.0 PLUS device and the predicates connect to a standard turbine connection on a dental operative unit and deliver a mixture of water, air, and prophylaxis powder to a treatment site. The handpiece component of the proposed AIR-FLOW handy 3.0 PLUS (PLUS) is to be commercialized as an additional optional accessory to the predicate AIR-FLOW handy 3.0 PERIO (PERIO) handpiece component. Therefore, the proposed PLUS and the predicate PERIO share the same EMS reference number FT-221#. The purpose for adding this additional PLUS handpiece is so that the patients can have both supragingival treatments and subgingival treatments by the same AIR-FLOW PERIO prophylaxis powder that is previously cleared in AIR-FLOW MASTER (K082791) and thus create more convenience for the users. There are no changes to the powder in this submission.

    The AIR-FLOW handy 3.0 (STANDARD) and the proposed device, AIR-FLOW handy 3.0 PLUS (PLUS) handpieces, have the same air and water ports to enable supragingival use.

    The difference between the proposed PLUS and the predicate PERIO is the handpiece, which has the same body, powder chamber and handpiece body with the only difference being the nozzle. The geometric shape of the integrated nozzle of the PLUS handpiece is relatively thin and slim which enables the supragingival use and the subgingival use in pockets up to 4 mm after initial periodontal treatment. The integrated nozzle of PLUS handpiece has no direct contact with the subgingival shallow pocket. The separate slim nozzle of PERIO that has direct contract with the subgingival pocket enables the subgingival use in pockets up to 5mm after initial periodontal treatment.

    The PLUS handpiece should not be used with the STANDARD powder chamber. To prevent this, a different connector is used so that it can only be used with the PERIO powder and PERIO powder chamber. The PLUS handpiece uses the same PERIO polishing powder as the PERIO handpiece.

    The proposed AIR-FLOW handy 3.0 PLUS device is therefore substantially equivalent to the previously FDA-cleared AIR-FLOW handy 3.0 PERIO (K132480) and AIR-FLOW handy 3.0 (K132478).

    AI/ML Overview

    The provided text describes the regulatory clearance for the "AIR-FLOW handy 3.0 PLUS," a dental handpiece. The information focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than providing detailed acceptance criteria and a study to prove meeting those criteria in the context of an AI/algorithm-based medical device.

    Based on the document, here's an analysis of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of predetermined acceptance criteria for performance metrics (like sensitivity, specificity, accuracy) that would be typical for an AI/algorithm-based device. Instead, the "acceptance criteria" are implied by the verification of compliance with established medical device standards and the comparison testing against predicate devices.

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with relevant medical device standards (e.g., sterilization, usability, biological evaluation, quality management, risk management, labeling, packaging).The AIR-FLOW handy 3.0 PLUS underwent testing to verify compliance with: - ANSI/AAMI ST79: 2010/A4:2013 (Steam sterilization) - ANSI/AAMI/ISO 17665-1:2013 (Moist heat sterilization) - IEC 62366 (2007+AMD1:2014 CSV) (Usability engineering) - ISO 10993-1 (2009/Cor1:2010) (Biological evaluation) - ISO 13485 (2003) + CORR. 1: 2009 (Quality Management Systems) - ISO 14971 (2007) (Risk Management) - ISO 15223-1 (2012) (Symbols for labels) - ISO 17664 (2004) (Sterilization information by manufacturer) - ISTA 2A (2011) (Preshipment packaging tests)
    Performance equivalence to predicate devices (AIR-FLOW handy 3.0 and AIR-FLOW handy 3.0 PERIO) in terms of powder flow rate, powder residue, powder working time, and cleaning efficiency.Comparison testing was performed for the subject device and predicate devices using currently available EMS powders. This test suite included: - Powder flow rate - Powder residue - Powder working time - Cleaning efficiency The document states: "The test suite results passed." and "The results of this testing confirm that the AIR-FLOW handy 3.0 PLUS is as safe and as effective for the intended use described in Section 6." It also concludes that it is "substantially equivalent to the predicates."
    Operational characteristics (water/air pressure, operating/storage/transport conditions) match or are within acceptable ranges of predicate.Table 1 shows identical Function (Air-polishing) and Mechanism of action (Projection of water/air/powder mixture). Operational characteristics (service pressure, operating/storage/transport condition ranges) are listed for the proposed device and are implicitly considered equivalent or acceptable compared to predicates since substantial equivalence is claimed.
    Intended Use capabilities (supragingival and subgingival cleaning up to 4mm) are supported.The device combines functionalities of the two predicates, allowing for both supragingival and subgingival cleaning in pockets up to 4mm. The testing results are used to support this combined indication.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide details on the sample sizes for the "comparison testing" of performance metrics (powder flow, residue, working time, cleaning efficiency). It simply states "comparison testing of the subject device and the predicate device as used with currently available EMS powders was performed." There is no mention of data provenance (e.g., country of origin) or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided in the document. The device is a dental handpiece, not an AI/algorithm that requires expert adjudication for ground truth. The testing performed was related to physical device performance and compliance with standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided in the document. Adjudication methods are typically associated with expert review of medical images or data for AI/algorithm validation, not for the physical performance testing of a dental handpiece.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic or clinical decision support tools. The device described is a physical dental handpiece.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone (algorithm only) performance study was not done. This concept is not applicable to a physical dental handpiece.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The concept of "ground truth" as typically applied to AI/algorithm development (e.g., expert consensus, pathology, outcomes data) is not directly applicable here. The "ground truth" for the performance testing would be the measured physical and operational characteristics of the device and its predicates, compared against engineering specifications and industry standards, as well as the observed cleaning efficiency, powder flow, etc.

    8. The sample size for the training set

    This information is not applicable and not provided. The device is not an AI/algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. The device is not an AI/algorithm that requires a training set or ground truth for training.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1