Search Results
Found 1 results
510(k) Data Aggregation
Dafilon Nonabsorbable Polyamide Surgical Sutures are indicated for use in all types of general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, microsurgery and neurological procedures.
Dafilon is a nonabsorbable monofilament surgical suture which is supplied sterile. Dafilon is composed of the long-chain aliphatic polymers polyamide 6 and/or 6,6. The Dafilon suture is offered undyed and dyed with FDA approved colorants Logwood extract in accordance with Title 21 CRF, §73.1410 or [Phthalocyaninato(2-)] copper in accordance with Title 21 CFR, §74.3045. The Dafilon suture will be offered in diameters ranging from USP size 11-0 through 5 and will be available in a variety of cut lengths with out needles attached.
This document is a 510(k) premarket notification for a surgical suture, not a device that involves performance criteria typically associated with AI/ML algorithms or interpretative medical devices. Therefore, many of the requested sections regarding acceptance criteria and study design are not applicable in the context of this document. This submission primarily focuses on demonstrating substantial equivalence to existing predicate devices based on material composition, indications for use, and mechanical/biocompatibility testing, rather than an AI model's diagnostic or predictive performance.
However, based on the provided text, I can extract the relevant "acceptance criteria" and "performance data" for this suture device where applicable.
Here's an interpretation of the document in the context of your request, focusing on the available information:
Device: Aesculap Dafilon Nonabsorbable Polyamide Surgical Suture
Since this is a surgical suture, the "acceptance criteria" are related to its physical properties, sterility, and biocompatibility, rather than diagnostic accuracy or sensitivity/specificity of a predictive model.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criterion (Suture Device) | Reported Device Performance |
|---|---|
| Physical Characteristics: | |
| Diameter | Meets USP Requirements. |
| Length | Meets USP Requirements. |
| Needle Attachment | Meets USP Requirements. |
| Tensile Strength | Meets USP Requirements. |
| Biocompatibility: | |
| Cytotoxicity | Testing conducted. Results demonstrate safety and effectiveness. |
| Sensitization | Testing conducted. Results demonstrate safety and effectiveness. |
| Intracutaneous Irritation | Testing conducted. Results demonstrate safety and effectiveness. |
| Systemic and Muscle Implantation (4-week) | Testing conducted. Results demonstrate safety and effectiveness. |
| Material Equivalency: | |
| Chemical Characterization (Polyamide raw material) | Results demonstrate substantial equivalence to predicate devices. |
| Sterilization & Shelf Life: | |
| Sterilization (Gamma Irradiation or EO) | Validated. |
| Shelf Life (Accelerated Aging) | Data generated to support submission. |
2. Sample size used for the test set and the data provenance
The document specifies "tests were conducted for diameter, tensile strength, and needle attachment" and "biocompatibility testing." However, it does not specify the sample sizes for these tests. The provenance is internal testing by Aesculap, Inc. against USP (United States Pharmacopeia) requirements and ISO (International Organization for Standardization) standards. The tests are likely prospective, conducted on manufactured samples of the device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to a nonabsorbable surgical suture device. The "ground truth" for suture performance is established by standardized physical and biological tests (USP, ISO standards), not by human expert interpretation or consensus in the way it would be for an AI diagnostic device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable for a surgical suture. Adjudication methods are relevant for subjective interpretations of data, typically in diagnostic studies. Here, the "truth" is determined by objective measurements against established standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a surgical suture, not an AI-assisted diagnostic tool. No human reader involvement or AI assistance is relevant to its performance evaluation for this 510(k) submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical medical device (suture), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" is based on established industry standards and regulatory requirements:
- USP Requirements: For physical characteristics such as diameter, length, needle attachment, and tensile strength.
- ISO 10993-1: For biocompatibility testing (Cytotoxicity, Sensitization, Intracutaneous Irritation, Systemic and Muscle Implantation).
- 21 CFR §73.1410 and §74.3045: For colorant approval.
8. The sample size for the training set
Not applicable. This device does not involve a "training set" in the context of machine learning or AI.
9. How the ground truth for the training set was established
Not applicable. As there is no training set, there is no ground truth establishment for it.
Ask a specific question about this device
Page 1 of 1