Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150942

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-11-25

    (231 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Contour Next USB blood glucose monitoring system is an over the counter (OTC) device utilized by persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for single-patient use only and should not be shared.

    The system consists of a Contour Next USB blood glucose meter, Contour Next test strips and Contour Next control solutions.

    The Contour Next USB blood glucose monitoring system is indicated for use with fresh capillary whole blood samples drawn from the fingertip and palm only. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).

    The Contour Next USB blood glucose monitoring system is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

    The Contour Next Test Strips are for use with the Contour Next USB blood glucose monitoring system for the quantitative measurement of glucose in whole blood.

    Device Description

    The Contour Next USB Blood Glucose Meter consists of a small handheld blood glucose meter that utilizes dry reagent test strips for the measurement of glucose in capillary whole blood by persons with diabetes. Liquid control solution is used to check the performance of the system. The meter, together with the test strips and control solutions, is referred to as the Contour Next USB Blood Glucose Monitoring System.

    The chemical principle utilized for both the predicate and modified devices is based on measurement of electrical current caused by the reaction of glucose in the blood with chemicals on the reagent strip. The blood sample is drawn into the tip of the reagent strip through capillary action. Glucose in the sample reacts with FAD glucose dehydrogenase (FAD-GDH) enzyme on the reagent strip. The electrons generated by this reaction are shuttled to an electrode by a mediator chemical, producing a current that is proportional to the glucose in the sample. After a fixed reaction time, the glucose concentration in the sample is calculated and displayed.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Contour Next USB Blood Glucose Monitoring System, which primarily describes modifications to an existing predicate device (K121087) rather than providing a detailed study of the modified device's performance against specific acceptance criteria. The modifications focus on improved error detection capabilities.

    Therefore, much of the requested information regarding acceptance criteria, specific performance metrics, sample sizes, expert involvement, and ground truth for a full performance study is not explicitly detailed in the provided text. The submission focuses on demonstrating substantial equivalence to the predicate device due to specific enhancements.

    Here's an attempt to extract the available information based on your request, with noted limitations where data is not provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria in a table format for blood glucose measurement accuracy for the modified device. Instead, it describes functional improvements related to error detection. The "reported device performance" focuses on the functionality of these new error checks and equivalency to the predicate device.

    Acceptance Criteria (Inferred from improvements)Reported Device Performance (Modified Device)
    Detect test strips exposed to degrading chemicalError message given for exposed strips
    Detect un-mixed control solutionError message given for un-mixed control solution
    Detect sample 'perturbation' during a testError message given for disturbed samples
    For overall glucose measurement accuracy:
    Not statistically different from predicate device (K121087)Equivalency testing confirmed performance not statistically different.
    No good results classified as errors by modificationsSoftware verification ensured no good results were misclassified.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not specified. The document mentions "bench testing" and an "internal user study" but does not give sample sizes for either.
    • Data Provenance: Not specified, but likely proprietary internal data from Bayer Healthcare. It doesn't specify country of origin or whether it's retrospective or prospective. Given the context of 510(k) submissions, such studies are typically prospective tests conducted by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable or not specified. This type of submission for a blood glucose meter does not typically involve expert consensus to establish ground truth in the same way an imaging or diagnostic AI device would. The "ground truth" for a blood glucose meter is usually established by a laboratory reference method. The document mentions "internal user study," suggesting lay users may have been involved in some testing, but no details on expert involvement for ground truth are provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable or not specified. Adjudication methods like 2+1 are typically used for subjective assessments (e.g., image interpretation) where multiple experts review and reconcile differences. For a quantitative device like a blood glucose meter, ground truth is established by a reference method, not by expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are relevant for devices that assist human readers (e.g., AI for medical imaging). This is a standalone diagnostic device. The study mentions an "internal user study" but not in the context of an MRMC design or comparing human performance with/without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in essence. The bench testing and software verification of the error detection mechanisms would fall under "standalone" algorithm testing to ensure the meter correctly identifies errors and does not misclassify good results. The equivalency testing would also implicitly assess the algorithm's accuracy compared to the predicate.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For blood glucose meters, the ground truth for accuracy studies is typically established by a laboratory reference method (e.g., YSI glucose analyzer) for whole blood glucose concentration. While not explicitly stated for this specific submission, this is the standard for blood glucose meter validation. The text focuses on the functional correctness of its new error checks rather than primary accuracy data against a reference method, as it's demonstrating equivalence to an already cleared device.

    8. The sample size for the training set

    Not applicable. As a medical device (blood glucose meter) and not an AI/ML algorithm that requires a "training set" in the conventional sense, this information is not relevant or provided. The "training" for such a device is its design and manufacturing process, and its performance is validated through testing.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the context of AI/ML for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1