510(k) Data Aggregation

    K Number
    K150275

    Validate with FDA (Live)

    Date Cleared
    2015-03-09

    (33 days)

    Product Code
    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10ng/mL. The assay is intended for use in laboratories for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers. This assay is calibrated against 6-Acetylmorphine. This in-vitro diagnostic device is for prescription use only.

    The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result. particularly when preliminary positive results are used.

    Immunalysis 6-Acetylmorphine Urine Controls: The Immunalysis 6-Acetylmorphine Urine Controls are used as control materials in Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay.

    Immunalysis 6-Acetylmorphine Urine Calibrator: The Immunalysis 6-Acetylmorphine Urine Calibrator is used as a calibrator in the Immunalysis Urine Enzyme Immunoassay for the qualitative determination of 6-Acetylmorphine in urine on automated clinical chemistry analyzers.

    Device Description

    The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes recombinant antibodies to 6-Acetylmorphine. glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in HEPES buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes 6-AcetyImorphine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in HEPES buffer with Sodium Azide as a preservative. Calibrator and controls are sold separately. Reagents are liquid, ready to use

    The 6-AcetyImorphine calibrator and controls consist of a cutoff calibrator at 10ng/mL, a LOW control at 7.5ng/mL for the 10ng/mL cutoff and a HIGH control at 12.5ng/mL for the 10ng/mL cutoff.

    AI/ML Overview

    The Immunalysis 6-acetylmorphine Urine Enzyme Immunoassay is intended for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers, with a cutoff of 10ng/mL. The study performed to establish substantial equivalence included tests for precision/cutoff characterization, specificity and cross-reactivity, interference, and method comparison.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a dedicated table. However, performance can be inferred from the results presented. The device aims to accurately classify samples as negative or positive for 6-Acetylmorphine based on the 10ng/mL cutoff.

    Test CategoryAcceptance Criteria (Inferred)Reported Device Performance
    Precision/Cutoff CharacterizationAll samples at -25% to -100% of cutoff should be negative. All samples at +25% to +100% of cutoff should be positive. Samples at cutoff (10ng/mL) should show a distribution around 50% positive/negative, demonstrating the cutoff as a boundary.Table 1: Qualitative Analysis (for 10ng/mL cutoff)- 0 to 7.5 ng/mL ( -100% to -25% of cutoff): 80/80 Negative results for each concentration.- 10 ng/mL (Cutoff): 43 Negative / 37 Positive (demonstrates the cutoff as a boundary).- 12.5 to 20 ng/mL (+25% to +100% of cutoff): 80/80 Positive results for each concentration. Conclusion: Meets inferred criteria.
    Specificity and Cross-ReactivityStructurally similar compounds should show expected cross-reactivity or non-cross-reactivity. Structurally dissimilar compounds should not interfere.Table 2: Structurally Related Compounds (for 10 ng/mL cutoff) - Qualitative- 6-Acetylmorphine (10 ng/mL): 100% cross-reactivity (Positive result).- 6-Acetylcodeine (600 ng/mL): 1.7% cross-reactivity (Positive result).- Heroin (1,375 ng/mL): 0.7% cross-reactivity (Positive result).- Hydromorphone (325,000 ng/mL): 0.003% cross-reactivity (Positive result).- Morphine (285,000 ng/mL): 0.000035% cross-reactivity (Positive result).- Nalorphine (80,000 ng/mL): 0.0125% cross-reactivity (Positive result).- Naloxone (300,000 ng/mL): 0.00333% cross-reactivity (Positive result).- Naltrexone (390,000 ng/mL): 0.00256% cross-reactivity (Positive result).- Normorphine (250,000 ng/mL): 0.004% cross-reactivity (Positive result).- Oxymorphone (360,000 ng/mL): 0.00277% cross-reactivity (Positive result).- Other listed compounds showed "Negative" results or N.D. (Not Determined). Conclusion: Meets inferred criteria, with predictable cross-reactivity for related opioids at higher concentrations, and no unexpected cross-reactivity for others.
    InterferenceStructurally non-similar compounds, endogenous compounds, and variations in pH and specific gravity should not cause interference, maintaining the correct classification of samples at ±25% of the cutoff. Exceptions should be noted.Tables 3 & 4: Structurally Non-Similar Compounds & Endogenous Compounds (for 10ng/mL cutoff)- Majority of compounds (e.g., Acetaminophen, Alprazolam, Caffeine, Ibuprofen) showed "No Interference" at concentrations up to 500,000 ng/mL for -25% and +25% cutoff samples.- Interferences noted: Ascorbic Acid (1.5 g/dL) caused interference at +25% cutoff (negative result when it should be positive), and at -50% cutoff (negative result when it should be positive). Sodium Chloride (6.0 g/dL) caused interference at +25% cutoff (negative result when it should be positive). Boric Acid (1% w/v) caused interference at +25% and +50% cutoffs (negative result when it should be positive).- Table 8: Effect of pH- pH 3.0 caused interference at +25% cutoff (negative result when it should be positive), and at +50% cutoff (negative result when it should be positive). pH values from 4.0 to 11.0 showed no interference. - Table 10: Effect of Specific Gravity- No interference was observed for specific gravity values from 1.000 to 1.030. Conclusion: Meets inferred criteria for most compounds, pH, and specific gravity. Identifiable interferences (Ascorbic Acid, Sodium Chloride, Boric Acid, low pH) are explicitly noted.
    Method Comparison100% agreement between the test device (qualitative results) and the confirmatory method (LC/MS) for samples around the cutoff.Tables 11: Method Comparison (for 10ng/mL cutoff) - Qualitative- LC/MS Confirmation: 40 Positive, 40 Negative.- Test Device: 40 Positive, 40 Negative.- Agreement: 100% agreement (40 True Positives, 40 True Negatives).- Qualitative/Positive: 100% agreement (4 samples 10-15 ng/mL, 36 samples >15 ng/mL correctly identified as positive).- Qualitative/Negative: 100% agreement (36 samples <5 ng/mL, 4 samples 5.0-9.9 ng/mL correctly identified as negative). Conclusion: Meets inferred criteria with 100% agreement with LC/MS.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision/Cutoff Characterization Study: N=80 determinations for each concentration level (0, 2.5, 5, 7.5, 10, 12.5, 15, 17.5, 20 ng/mL). The study was performed for 20 days, 2 runs per day in duplicate. The document does not specify the provenance of the samples (e.g., country of origin) or if they were retrospective or prospective.
    • Specificity and Cross-Reactivity Study: Structurally similar compounds were spiked into drug-free urine. The sample size for these individual tests is not explicitly stated but implies multiple replicates to confirm the result.
    • Interference Study: Structurally non-similar compounds, endogenous compounds, pH effects, and specific gravity effects were evaluated by spiking potential interferents into drug-free urine containing the target analyte at ±25% of the cutoff. Sample sizes are not explicitly stated for each individual interferent but imply sufficient replicates to confirm results. Additional testing for ascorbic acid, sodium chloride, and boric acid was done at ±50% of the cutoff, and pH 3.0 at ±50% of the cutoff.
    • Method Comparison Study: 80 clinical urine samples were used (40 LC/MS confirmed positive, 40 LC/MS confirmed negative). These were "Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories." The provenance (country of origin) is not specified, but the samples were retrospective from clinical testing laboratories.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    No human experts were used to establish the ground truth for the test set. The ground truth for the method comparison study was established by Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS), which are considered preferred confirmatory chemical methods.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    No human adjudication method was used for the test set, as the ground truth was based on chemical analysis (LC/MS).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is an in-vitro diagnostic device (immunoassay) for the qualitative analysis of 6-Acetylmorphine in human urine. It is not an AI-assisted diagnostic device, nor does it involve human readers interpreting images or data in a way that an MRMC study would be applicable. Therefore, no MRMC comparative effectiveness study was performed, and no effect size for human reader improvement with AI assistance is provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented are essentially a standalone performance evaluation of the "algorithm" (the immunoassay kit run on automated clinical chemistry analyzers). The device's performance is measured directly against predefined cutoffs and confirmatory methods without human interpretation influencing the primary result. The results are qualitative (positive/negative) based on the instrument's optical density readings compared to a calibrator.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The primary ground truth used for relevant studies (e.g., Method Comparison) was Liquid Chromatography / Mass Spectrometry (LC/MS) confirmation. For other studies like specificity and interference, the ground truth was established by spiking known concentrations of compounds into drug-free urine.

    8. The sample size for the training set

    This document describes a pre-market notification (510(k)) for an immunoassay kit. Immunoassays are not "trained" in the same way machine learning algorithms are. Therefore, there is no "training set" in the context of an AI/ML model for this device. The development process involves reagent formulation and optimization, followed by validation studies as described.

    9. How the ground truth for the training set was established

    As there is no "training set" in the context of AI/ML, this question is not applicable. The assay is calibrated against 6-Acetylmorphine using specific calibrator materials. The calibrator and control materials themselves have their values assigned by mass spectrometry, ensuring traceability to commercially available standards.