K Number
K150275

Validate with FDA (Live)

Date Cleared
2015-03-09

(33 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10ng/mL. The assay is intended for use in laboratories for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers. This assay is calibrated against 6-Acetylmorphine. This in-vitro diagnostic device is for prescription use only.

The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result. particularly when preliminary positive results are used.

Immunalysis 6-Acetylmorphine Urine Controls: The Immunalysis 6-Acetylmorphine Urine Controls are used as control materials in Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay.

Immunalysis 6-Acetylmorphine Urine Calibrator: The Immunalysis 6-Acetylmorphine Urine Calibrator is used as a calibrator in the Immunalysis Urine Enzyme Immunoassay for the qualitative determination of 6-Acetylmorphine in urine on automated clinical chemistry analyzers.

Device Description

The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes recombinant antibodies to 6-Acetylmorphine. glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in HEPES buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes 6-AcetyImorphine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in HEPES buffer with Sodium Azide as a preservative. Calibrator and controls are sold separately. Reagents are liquid, ready to use

The 6-AcetyImorphine calibrator and controls consist of a cutoff calibrator at 10ng/mL, a LOW control at 7.5ng/mL for the 10ng/mL cutoff and a HIGH control at 12.5ng/mL for the 10ng/mL cutoff.

AI/ML Overview

The Immunalysis 6-acetylmorphine Urine Enzyme Immunoassay is intended for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers, with a cutoff of 10ng/mL. The study performed to establish substantial equivalence included tests for precision/cutoff characterization, specificity and cross-reactivity, interference, and method comparison.

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria in a dedicated table. However, performance can be inferred from the results presented. The device aims to accurately classify samples as negative or positive for 6-Acetylmorphine based on the 10ng/mL cutoff.

Test CategoryAcceptance Criteria (Inferred)Reported Device Performance
Precision/Cutoff CharacterizationAll samples at -25% to -100% of cutoff should be negative. All samples at +25% to +100% of cutoff should be positive. Samples at cutoff (10ng/mL) should show a distribution around 50% positive/negative, demonstrating the cutoff as a boundary.Table 1: Qualitative Analysis (for 10ng/mL cutoff)- 0 to 7.5 ng/mL ( -100% to -25% of cutoff): 80/80 Negative results for each concentration.- 10 ng/mL (Cutoff): 43 Negative / 37 Positive (demonstrates the cutoff as a boundary).- 12.5 to 20 ng/mL (+25% to +100% of cutoff): 80/80 Positive results for each concentration. Conclusion: Meets inferred criteria.
Specificity and Cross-ReactivityStructurally similar compounds should show expected cross-reactivity or non-cross-reactivity. Structurally dissimilar compounds should not interfere.Table 2: Structurally Related Compounds (for 10 ng/mL cutoff) - Qualitative- 6-Acetylmorphine (10 ng/mL): 100% cross-reactivity (Positive result).- 6-Acetylcodeine (600 ng/mL): 1.7% cross-reactivity (Positive result).- Heroin (1,375 ng/mL): 0.7% cross-reactivity (Positive result).- Hydromorphone (325,000 ng/mL): 0.003% cross-reactivity (Positive result).- Morphine (285,000 ng/mL): 0.000035% cross-reactivity (Positive result).- Nalorphine (80,000 ng/mL): 0.0125% cross-reactivity (Positive result).- Naloxone (300,000 ng/mL): 0.00333% cross-reactivity (Positive result).- Naltrexone (390,000 ng/mL): 0.00256% cross-reactivity (Positive result).- Normorphine (250,000 ng/mL): 0.004% cross-reactivity (Positive result).- Oxymorphone (360,000 ng/mL): 0.00277% cross-reactivity (Positive result).- Other listed compounds showed "Negative" results or N.D. (Not Determined). Conclusion: Meets inferred criteria, with predictable cross-reactivity for related opioids at higher concentrations, and no unexpected cross-reactivity for others.
InterferenceStructurally non-similar compounds, endogenous compounds, and variations in pH and specific gravity should not cause interference, maintaining the correct classification of samples at ±25% of the cutoff. Exceptions should be noted.Tables 3 & 4: Structurally Non-Similar Compounds & Endogenous Compounds (for 10ng/mL cutoff)- Majority of compounds (e.g., Acetaminophen, Alprazolam, Caffeine, Ibuprofen) showed "No Interference" at concentrations up to 500,000 ng/mL for -25% and +25% cutoff samples.- Interferences noted: Ascorbic Acid (1.5 g/dL) caused interference at +25% cutoff (negative result when it should be positive), and at -50% cutoff (negative result when it should be positive). Sodium Chloride (6.0 g/dL) caused interference at +25% cutoff (negative result when it should be positive). Boric Acid (1% w/v) caused interference at +25% and +50% cutoffs (negative result when it should be positive).- Table 8: Effect of pH- pH 3.0 caused interference at +25% cutoff (negative result when it should be positive), and at +50% cutoff (negative result when it should be positive). pH values from 4.0 to 11.0 showed no interference. - Table 10: Effect of Specific Gravity- No interference was observed for specific gravity values from 1.000 to 1.030. Conclusion: Meets inferred criteria for most compounds, pH, and specific gravity. Identifiable interferences (Ascorbic Acid, Sodium Chloride, Boric Acid, low pH) are explicitly noted.
Method Comparison100% agreement between the test device (qualitative results) and the confirmatory method (LC/MS) for samples around the cutoff.Tables 11: Method Comparison (for 10ng/mL cutoff) - Qualitative- LC/MS Confirmation: 40 Positive, 40 Negative.- Test Device: 40 Positive, 40 Negative.- Agreement: 100% agreement (40 True Positives, 40 True Negatives).- Qualitative/Positive: 100% agreement (4 samples 10-15 ng/mL, 36 samples >15 ng/mL correctly identified as positive).- Qualitative/Negative: 100% agreement (36 samples <5 ng/mL, 4 samples 5.0-9.9 ng/mL correctly identified as negative). Conclusion: Meets inferred criteria with 100% agreement with LC/MS.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Precision/Cutoff Characterization Study: N=80 determinations for each concentration level (0, 2.5, 5, 7.5, 10, 12.5, 15, 17.5, 20 ng/mL). The study was performed for 20 days, 2 runs per day in duplicate. The document does not specify the provenance of the samples (e.g., country of origin) or if they were retrospective or prospective.
  • Specificity and Cross-Reactivity Study: Structurally similar compounds were spiked into drug-free urine. The sample size for these individual tests is not explicitly stated but implies multiple replicates to confirm the result.
  • Interference Study: Structurally non-similar compounds, endogenous compounds, pH effects, and specific gravity effects were evaluated by spiking potential interferents into drug-free urine containing the target analyte at ±25% of the cutoff. Sample sizes are not explicitly stated for each individual interferent but imply sufficient replicates to confirm results. Additional testing for ascorbic acid, sodium chloride, and boric acid was done at ±50% of the cutoff, and pH 3.0 at ±50% of the cutoff.
  • Method Comparison Study: 80 clinical urine samples were used (40 LC/MS confirmed positive, 40 LC/MS confirmed negative). These were "Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories." The provenance (country of origin) is not specified, but the samples were retrospective from clinical testing laboratories.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

No human experts were used to establish the ground truth for the test set. The ground truth for the method comparison study was established by Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS), which are considered preferred confirmatory chemical methods.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

No human adjudication method was used for the test set, as the ground truth was based on chemical analysis (LC/MS).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is an in-vitro diagnostic device (immunoassay) for the qualitative analysis of 6-Acetylmorphine in human urine. It is not an AI-assisted diagnostic device, nor does it involve human readers interpreting images or data in a way that an MRMC study would be applicable. Therefore, no MRMC comparative effectiveness study was performed, and no effect size for human reader improvement with AI assistance is provided.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the studies presented are essentially a standalone performance evaluation of the "algorithm" (the immunoassay kit run on automated clinical chemistry analyzers). The device's performance is measured directly against predefined cutoffs and confirmatory methods without human interpretation influencing the primary result. The results are qualitative (positive/negative) based on the instrument's optical density readings compared to a calibrator.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The primary ground truth used for relevant studies (e.g., Method Comparison) was Liquid Chromatography / Mass Spectrometry (LC/MS) confirmation. For other studies like specificity and interference, the ground truth was established by spiking known concentrations of compounds into drug-free urine.

8. The sample size for the training set

This document describes a pre-market notification (510(k)) for an immunoassay kit. Immunoassays are not "trained" in the same way machine learning algorithms are. Therefore, there is no "training set" in the context of an AI/ML model for this device. The development process involves reagent formulation and optimization, followed by validation studies as described.

9. How the ground truth for the training set was established

As there is no "training set" in the context of AI/ML, this question is not applicable. The assay is calibrated against 6-Acetylmorphine using specific calibrator materials. The calibrator and control materials themselves have their values assigned by mass spectrometry, ensuring traceability to commercially available standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

IMMUNALYSIS CORPORATION JOSEPH GINETE REGULATORY AFFAIRS SPECIALIST 829 TOWNE CENTER DR. POMONA CA 91767

March 9, 2015

Re: K150275

Trade/Device Name: Immunalysis 6-acetylmorphine Urine Enzyme Immunoassay, Immunalysis 6-acetylmorphine Urine Calibrator, Immunalysis 6-acetylmorphine Urine Controls

Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, DKB, DIF Dated: January 20, 2015 Received: February 4, 2015

Dear Mr. Ginete:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -A

For : Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150275

Device Name

Immunalysis 6-Acetylmorphine Urine Enzyme Immunalysis 6-Acetylmorphine Urine Controls and Immunalysis 6-Acetylmorphine Urine Calibrator

Indications for Use (Describe)

The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10ng/mL. The assay is intended for use in laboratories for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers. This assay is callbrated against 6-Acetylmorphine. This in-vitro diagnostic device is for prescription use only.

The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result. particularly when preliminary positive results are used.

Immunalysis 6-Acetylmorphine Urine Controls: The Immunalysis 6-Acetylmorphine Urine Controls are used as control materials in Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay.

Immunalysis 6-Acetylmorphine Urine Calibrator: The Immunalysis 6-Acetylmorphine Urine Calibrator is used as a calibrator in the Immunalysis Urine Enzyme Immunoassay for the qualitative determination of 6-Acetylmorphine in urine on automated clinical chemistry analyzers.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and appears to be in bold font. The red background is a solid color, and the white letters stand out against it.

A. Contact Information

510(k) SUMMARY (K150275)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).

    1. Manufacturer: Immunalysis Corporation 2. Contact Name: Joseph Ginete Regulatory Affairs Specialist 3. Contact Title: 4. Address: 829 Towne Center Drive Pomona, CA 91767 (909) 482-0840 5. Phone: (909) 482-0850 6. Fax: 7. Email: iginete@immunalysis.com 8. Summary prepared on: March 6, 2015 B. Device Information 1. Trade Name: Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay Immunalysis 6-AcetyImorphine Urine Controls Immunalysis 6-Acetylmorphine Urine Calibrator 2. Common Name: Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay Immunalysis 6-AcetyImorphine Urine Controls Immunalysis 6-AcetyImorphine Urine Calibrator 3. Device Classification: II 4. Regulation Number: 21 CFR 862.3650 Opiate Test System 21 CFR 862.3200 Clinical Toxicology Calibrator 21 CFR 862.3280 Clinical Toxicology Control Material 5. Panel: Toxicology(91) 6. Product Code: DJG DKB DIF C. Legally Marketed Device to Which We are Claiming Equivalence (807.92(A)(3))
      1. Predicate Device: Emit® II Plus 6-AcetyImorphine Assay Emit® II Plus 6-AM/ Ecstasy Calibrators Emit® II Plus 6-AM/ Ecstasy Controls 2. Predicate Company: Siemens Healthcare Diagnostics Inc. 3. Predicate K Number: K102779

{4}------------------------------------------------

D. Device Description

The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes recombinant antibodies to 6-Acetylmorphine. glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in HEPES buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes 6-AcetyImorphine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in HEPES buffer with Sodium Azide as a preservative. Calibrator and controls are sold separately. Reagents are liquid, ready to use

The 6-AcetyImorphine calibrator and controls consist of a cutoff calibrator at 10ng/mL, a LOW control at 7.5ng/mL for the 10ng/mL cutoff and a HIGH control at 12.5ng/mL for the 10ng/mL cutoff.

  • E. Intended Use
    The Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10ng/mL. The assay is intended for use in laboratories for the qualitative analysis of 6-Acetylmorphine in human urine with automated clinical chemistry analyzers. This assay is calibrated against 6-Acetylmorphine. This in-vitro diagnostic device is for prescription use only.

The Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liguid Chromatography / Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

lmmunalysis 6-Acetylmorphine Urine Controls: The Immunalysis 6-Acetylmorphine Urine Controls are used as control materials in Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay.

Immunalysis 6-AcetyImorphine Urine Calibrator: The Immunalysis 6-AcetyImorphine Urine Calibrator is used as a calibrator in the Immunalysis Urine Enzyme Immunoassay for the qualitative determination of 6-Acetylmorphine in urine on automated clinical chemistry analyzers.

6-Acetylmorphine Assay K102779Immunalysis 6-Acetylmorphine Urine EIA
Intended UseThe qualitative analysis of 6-Acetylmorphine in human urine withautomated clinical chemistryanalyzersSame
Type of ProductAnalytical ReagentsSame
Measured Analytes6-AcetylmorphineSame
Test MatrixUrineSame
Cutoff Levels10ng/mL of 6-AcetylmorphineSame
Test SystemHomogeneous EnzymeImmunoassaySame
MaterialsLiquid Ready-to-Use R1 Reagentand Lyophilized R2 ReagentLiquid Ready-to-Use Antibody/SubstrateReagents and Liquid Ready-to-Use EnzymeLabeled Conjugate
  • F. Comparison of the new device with the predicate device

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "IMMUNALYS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects or details in the image. The font is sans-serif and the letters are bold.

6-Acetylmorphine Assay K102779Immunalysis 6-Acetylmorphine Urine ElA
Mass SpectroscopyConfirmationRequired for preliminary positiveanalytical resultsSame
AntibodyMouse monoclonal antibody to 6-AcetylmorphineRecombinant antibody to 6-Acetylmorphine
Storage2 – 8°C until expiration dateSame
Calibrator FormLiquidSame
Calibrator LevelFour (4) Levels (5, 10, 15 and 20ng/mL)One (1) Level (10ng/mL)
Control LevelsTwo (2) Levels (7.5 and 12.5ng/mL)Same
  • G. The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay to the predicate
      1. Precision/Cutoff Characterization Study was performed for 20 days, 2 runs per day in duplicate (N=80) on concentration of ±25%, ±50%, ±75%, and ±100% of the cutoff. The study verified that the cutoff serves as a boundary between a negative and positive interpretation of a qualitative result. In addition, it also verified product performance relative to the ability of the device to produce the same value during repeated measurements. The instruments used for this was Beckman Coulter AU 400e. y table of the Qualitativ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 1. Qualitative Analysis (for 10ng/mL cutoff)
The following is a summary table of the Qualitative Analysis for the 10ng/mL cutoff test data results.
Table 1 - Qualitative Analysis (for 10ng/mL cutoff)
Concentration (ng/mL)% of cutoff# of determinationsResult
0-100%8080 Negative
2.5-75%8080 Negative
5-50%8080 Negative
7.5-25%8080 Negative
10Cutoff8043 Neg / 37 Pos
12.5+25%8080 Positive
15+50%8080 Positive
17.5+75%8080 Positive
20+100%8080 Positive
    1. Specificity and Cross-Reactivity Structurally similar compounds were spiked into drug free urine at levels that will yield a result that is equivalent to the cutoff. The instrument used for this test was a Beckman Coulter AU 400e.
      The qualitative result summary table for the 10ng/mL cutoff is outlined below:
Table 2 - Structurally Related Compounds (for 10 ng/mL cutoff) - Qualitative
CompoundConcentration Tested (ng/mL)ResultCross-Reactivity (%)
6-Acetylmorphine10Positive100
6-Acetylcodeine600Positive1.7
Buprenorphine1,000,000NegativeN.D.
Codeine1,000,000NegativeN.D.
Dextromethorphan1,000,000NegativeN.D.
Dihydrocodeine1,000,000NegativeN.D.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters with a drop shadow. The word is set against a red background. The text appears to be bolded and is the central focus of the image.

Table 2 - Structurally Related Compounds (for 10 ng/mL cutoff) - Qualitative
Ethylmorphine1,000,000NegativeN.D.
Heroin1,375Positive0.7
Hydrocodone1,000,000NegativeN.D.
Hydromorphone325,000Positive0.003
Imipramine1,000,000NegativeN.D.
Levorphanol1,000,000NegativeN.D.
Meperidine1,000,000NegativeN.D.
Morphine285,000Positive0.000035
Morphine 3-D-glucuronide1,000,000NegativeN.D.
Morphine 6-D-glucuronide1,000,000NegativeN.D.
Nalorphine80,000Positive0.0125
Naloxone300,000Positive0.00333
Naltrexone390,000Positive0.00256
Naproxen1,000,000NegativeN.D.
Norbuprenorphine100,000NegativeN.D.
Norcodeine1,000,000NegativeN.D.
Normorphine250,000Positive0.004
Oxycodone1,000,000NegativeN.D.
Oxymorphone360,000Positive0.00277
  1. Interference - Structurally non-similar compounds, endogenous compounds, the effect of pH and the effect of specific gravity was evaluated by spiking the potential interferent into drug free urine containing the target analyte at ±25% of the cutoff. The instrument used for this test was a Beckman Coulter AU 400e. a. The following is a summary table of the structurally non-similar
compounds for the 10ng/mL cutoff :
Table 3 - Structurally Non-Similar Compounds (for 10ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-25% Cutoff (7.5ng/mL)ResultInterference?+25% Cutoff (12.5ng/mL)ResultInterference?
4-Bromo-2,5,Dimethoxyphenethylamine100,000NegativeNoPositiveNo
Acetaminophen500,000NegativeNoPositiveNo
Acetylsalicyclic Acid500,000NegativeNoPositiveNo
Alprazolam100,000NegativeNoPositiveNo
7-Aminoclonazepam100,000NegativeNoPositiveNo
7-Aminoflurnitrazepam100,000NegativeNoPositiveNo
7-Aminonitrazepam100,000NegativeNoPositiveNo
Amitriptyline100,000NegativeNoPositiveNo
Amobarbital100,000NegativeNoPositiveNo
S-(+) Amphetamine100,000NegativeNoPositiveNo
Benzoylecgonine500,000NegativeNoPositiveNo
Benzylpiperazine100,000NegativeNoPositiveNo
Bromazepam100,000NegativeNoPositiveNo
Bupropion100,000NegativeNoPositiveNo
Butabarbital100,000NegativeNoPositiveNo
Butalbital100,000NegativeNoPositiveNo
Caffeine500,000NegativeNoPositiveNo
Carbamazepine100,000NegativeNoPositiveNo
Chlorpromazine100,000NegativeNoPositiveNo
cis Tramadol100,000NegativeNoPositiveNo
Table 3 - Structurally Non-Similar Compounds (for 10ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-25% Cutoff (7.5ng/mL)+25% Cutoff (12.5ng/mL)
ResultInterference?ResultInterference?
Clobazam100,000NegativeNoPositiveNo
Clomipramine100,000NegativeNoPositiveNo
Clonazepam100,000NegativeNoPositiveNo
Cannabidiol100,000NegativeNoPositiveNo
Cannabinol100,000NegativeNoPositiveNo
Carisoprodol100,000NegativeNoPositiveNo
Chlordiazepoxide100,000NegativeNoPositiveNo
Cocaine100,000NegativeNoPositiveNo
Cotinine100,000NegativeNoPositiveNo
Cyclobenzaprine100,000NegativeNoPositiveNo
Digoxin100,000NegativeNoPositiveNo
Desalkyflurazepam100,000NegativeNoPositiveNo
Dehydronorketamine50,000NegativeNoPositiveNo
Demoxepam100,000NegativeNoPositiveNo
Delta-9-THC100,000NegativeNoPositiveNo
Desipramine100,000NegativeNoPositiveNo
N-desmethyltapentadol100,000NegativeNoPositiveNo
Diazepam100,000NegativeNoPositiveNo
Diphenhydramine500,000NegativeNoPositiveNo
Doxepin100,000NegativeNoPositiveNo
Ecgonine100,000NegativeNoPositiveNo
Ecgonine methyl ester100,000NegativeNoPositiveNo
EDDP100,000NegativeNoPositiveNo
1R,2S(-)-Ephedrine100,000NegativeNoPositiveNo
1S,2R(+)-Ephedrine100,000NegativeNoPositiveNo
Ethyl glucuronide100,000NegativeNoPositiveNo
Fenfluramine100,000NegativeNoPositiveNo
Fentanyl100,000NegativeNoPositiveNo
Flunitrazepam100,000NegativeNoPositiveNo
Fluoxetine100,000NegativeNoPositiveNo
Flurazepam100,000NegativeNoPositiveNo
Haloperidol100,000NegativeNoPositiveNo
Hexobarbital100,000NegativeNoPositiveNo
11-hydroxy-delta-9-THC100,000NegativeNoPositiveNo
Ibuprofen500,000NegativeNoPositiveNo
Ketamine100,000NegativeNoPositiveNo
Lamotrignine100,000NegativeNoPositiveNo
Lidocaine100,000NegativeNoPositiveNo
Lorazepam100,000NegativeNoPositiveNo
Lorazepam Glucuronide50,000NegativeNoPositiveNo
Lormetazepam100,000NegativeNoPositiveNo
LSD100,000NegativeNoPositiveNo
Maprotiline100,000NegativeNoPositiveNo
(+)-MDA100,000NegativeNoPositiveNo
MDEA100,000NegativeNoPositiveNo
MDMA100,000NegativeNoPositiveNo
Meprobamate100,000NegativeNoPositiveNo
Methadone100,000NegativeNoPositiveNo
S(+)-Methamphetamine100,000NegativeNoPositiveNo
Methaquolone100,000NegativeNoPositiveNo
Methoxetamine100,000NegativeNoPositiveNo
Methylphenidate100,000NegativeNoPositiveNo
Midazolam100,000NegativeNoPositiveNo

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and slightly blurred, giving them a three-dimensional appearance. The red background is a solid color, and the overall image has a simple, clean design.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and slightly blurred, giving them a soft appearance. The red background is a solid color, providing a strong contrast to the white text.

Table 3 - Structurally Non-Similar Compounds (for 10ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-25% Cutoff (7.5ng/mL)+25% Cutoff (12.5ng/mL)
ResultInterference?ResultInterference?
Nitrazepam100,000NegativeNoPositiveNo
Nordiazepam100,000NegativeNoPositiveNo
Norketamine100,000NegativeNoPositiveNo
Norpropoxyphene100,000NegativeNoPositiveNo
Norpseudoephedrine100,000NegativeNoPositiveNo
Nortriptyline100,000NegativeNoPositiveNo
Oxazepam100,000NegativeNoPositiveNo
Oxazepam glucuronide100,000NegativeNoPositiveNo
PCP100,000NegativeNoPositiveNo
Pentazocine100,000NegativeNoPositiveNo
Phentermine100,000NegativeNoPositiveNo
Pentobarbital100,000NegativeNoPositiveNo
Phenobarbital100,000NegativeNoPositiveNo
Phenylephedrine100,000NegativeNoPositiveNo
Phenylpropanolamine100,000NegativeNoPositiveNo
Phenytoin100,000NegativeNoPositiveNo
PMA100,000NegativeNoPositiveNo
Prazepam100,000NegativeNoPositiveNo
Proproxyphene100,000NegativeNoPositiveNo
Propranolol100,000NegativeNoPositiveNo
Protriptyline100,000NegativeNoPositiveNo
R,R(-)-Pseudoephedrine100,000NegativeNoPositiveNo
S,S(+)-Pseudoephedrine100,000NegativeNoPositiveNo
Ranitidine100,000NegativeNoPositiveNo
Ritalinic Acid100,000NegativeNoPositiveNo
Salicylic Acid100,000NegativeNoPositiveNo
Secobarbital100,000NegativeNoPositiveNo
Sertraline100,000NegativeNoPositiveNo
Sufentanil Citrate50,000NegativeNoPositiveNo
11-nor-9 carboxy THC100,000NegativeNoPositiveNo
Temazepam100,000NegativeNoPositiveNo
Theophylline100,000NegativeNoPositiveNo
Thioridazine100,000NegativeNoPositiveNo
Triazolam100,000NegativeNoPositiveNo
Trifluoromethylphenyl-piperazine100,000NegativeNoPositiveNo
Trimipramine100,000NegativeNoPositiveNo
Trazodone100,000NegativeNoPositiveNo
Verapamil100,000NegativeNoPositiveNo
Venlafaxine100,000NegativeNoPositiveNo
Zolpidem Tartrate100,000NegativeNoPositiveNo

b. The following is a summary table of the endogenous compounds results for the 10ng/mL cutoff:

Table 4 - Endogenous Compounds (for 10ng/mL cutoff)
Concentration-25% Cutoff (7.5ng/mL)+25% Cutoff (12.5ng/mL)
CompoundTested (ng/mL)ResultInterference?ResultInterference?
Acetone1.0 g/dLNegativeNoPositiveNo
Ascorbic Acid1.5 g/dLNegativeNoNegativeYes
Bilirubin0.002 g/dLNegativeNoPositiveNo
Creatinine0.5 g/dLNegativeNoPositiveNo
Ethanol1.0 g/dLNegativeNoPositiveNo
Galactose0.01 g/dLNegativeNoPositiveNo
y-Globulin0.5 g/dLNegativeNoPositiveNo

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and slightly shadowed, giving them a three-dimensional appearance. The red background is a solid color, and the word is centered.

Table 4 - Endogenous Compounds (for 10ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-25% Cutoff (7.5ng/mL)+25% Cutoff (12.5ng/mL)
ResultInterference?ResultInterference?
Glucose2.0 g/dLNegativeNoPositiveNo
Hemoglobin0.115 g/dLNegativeNoPositiveNo
Human Serum Albumin0.5 g/dLNegativeNoPositiveNo
Oxalic Acid0.1 g/dLNegativeNoPositiveNo
Riboflavin0.0075 g/dLNegativeNoPositiveNo
Sodium Azide1% w/vNegativeNoPositiveNo
Sodium Chloride6.0 g/dLNegativeNoNegativeYes
Sodium Fluoride1% w/vNegativeNoPositiveNo
Urea6.0 g/dLNegativeNoPositiveNo

c. The following is a summary table of the additional ascorbic acid and sodium chloride testing at ±50% of the cutoff:

Table 5 - Endogenous Compounds (for 10ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-50% Cutoff (5ng/mL)+50% Cutoff (15ng/mL)
ResultInterference?ResultInterference?
Ascorbic Acid1.5 g/dLNegativeNoNegativeYes
Sodium Chloride6.0 g/dLNegativeNoPositiveNo

d. The following is a summary table of Boric Acid for the 10ng/mL cutoff results:

Table 6 - Boric Acid (for 10ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-25% Cutoff (7.5ng/mL)+25% Cutoff (12.5ng/mL)
Boric Acid1% w/vResultInterference?ResultInterference?
NegativeNoNegativeYes

e. The following is a summary table of the additional boric acid testing at +50% of the cutoff:

Table 7 - Boric Acid (for 10ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-50% Cutoff (5ng/mL)+50% Cutoff (15ng/mL)
ResultInterference?ResultInterference?
Boric Acid1% w/vNegativeNoNegativeYes

f. The following is a summary table of the effect of pH results for the 10ng/mL cutoff:

Table 8 - Effect of pH (for 10ng/mL cutoff)
Test ParameterValue-25% Cutoff (7.5ng/mL)+25% Cutoff (12.5ng/mL)
ResultInterference?ResultInterference?
pH3.0NegativeNoNegativeYes
pH4.0NegativeNoPositiveNo
pH5.0NegativeNoPositiveNo
pH6.0NegativeNoPositiveNo
pH7.0NegativeNoPositiveNo
pH8.0NegativeNoPositiveNo
pH9.0NegativeNoPositiveNo
pH10.0NegativeNoPositiveNo
pH11.0NegativeNoPositiveNo

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color, and the white letters stand out clearly against it. The font is sans-serif and appears to be bolded.

Table 9 - Effect of pH (for 10ng/mL cutoff)
Test ParameterValue-50% Cutoff (5ng/mL)+50% Cutoff (15ng/mL)
ResultInterference?ResultInterference?
pH3.0NegativeNoPositiveNo
  • g. The following is a summary table of the additional pH 3.0 testing at ±50%
  • h. The following is a summary table of the effect of specific gravity results for 10ng/mL cutoff:
Table 10 - Effect of Specific Gravity (for 10ng/mL cutoff)
Test ParameterValue-25% Cutoff (7.5ng/mL)+25% Cutoff (12.5ng/mL)
Specific Gravity1.000ResultInterference?ResultInterference?
Specific Gravity1.000NegativeNoPositiveNo
Specific Gravity1.002NegativeNoPositiveNo
Specific Gravity1.005NegativeNoPositiveNo
Specific Gravity1.010NegativeNoPositiveNo
Specific Gravity1.015NegativeNoPositiveNo
Specific Gravity1.020NegativeNoPositiveNo
Specific Gravity1.025NegativeNoPositiveNo
Specific Gravity1.030NegativeNoPositiveNo
  • i. Ascorbic acid, sodium chloride, boric acid and pH of 3 are considered interferences
    1. Method Comparison Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories were analyzed with the test device. The study verified that the product performance can be verified by Mass Spectrometry. The instrument used for this test was a Beckman Coulter AU 400e and an Agilent 6430 Liquid Chromatography Tandem Mass Spectrometry.
    • a. The following is a comparison table of qualitative assay performance for the 10ng/mL cutoff:

Table 11 - Method Comparison (for 10ng/mL cutoff) - Qualitative

LC/MS Confirmation
(+)(-)
Test Device(+)400
(-)040
  • b. The following is a summary table of qualitative assay performance for the 10ng/mL cutoff:
Type6-Acetylmorphine ConcentrationAgreement (%)
< 5ng/mL5.0 ~ 9.9 ng/mL10 ~ 15 ng/mL> 15 ng/mL
Qualitative/ Positive00436100
Qualitative/ Negative36400100
    1. Stability -
      A closed accelerated stability study was performed on reagents, calibrator and controls at 25°C to establish the initial expiration dating. The stability study supported an initial expiration date of 1 year for reagents. This stability study supported an initial expiration date of 12 months for calibrator and controls. The instrument used for this test was a Beckman Coulter AU 400e. Real stability studies are ongoing.
    1. Calibrator and Control Traceability all components of the calibrator and controls have been traced to a commercially available standard solution.
    1. Calibrator and Control Stability An open accelerated stability study was performed at 25°C to establish the initial open vial expiration dating. The stability

{11}------------------------------------------------

study supported an initial open vial expiration date of 12 months. The instrument used for this test was a Beckman Coulter AU 400e. All calibrator level(s) (10ng/mL) and control level(s) (7.5ng/mL and 12.5ng/mL) were within specifications for Day 0, 2, 8, 16, 24, 30, 32, and 40. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.

  • Calibrator and Control Value Assignment calibrator and controls are 8. manufactured and are tested by mass spectrometry. If any of the analytes are out of the acceptable range, then the calibrator and control is adjusted and retested. Values are assigned to the calibrator and controls once the mass spectrometry results are within the acceptable ranges.
  • H. Conclusion

The information provided in this pre-market notification demonstrates that the Immunalysis 6-AcetyImorphine Urine Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its general intended use.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).