Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K143409

    Validate with FDA (Live)

    Date Cleared
    2015-03-26

    (118 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siemens SOMATOM Definition AS Open systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

    (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

    Device Description

    New software version syngo® VA48 (SOMARIS/7 VA48) is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation that will be available on the SOMATOM Definition AS Open Computed Tomography systems. syngo® VA48 (SOMARIS/7 VA48) is a further development to the SOMARIS/7 operating software cleared as part of the predicate device.

    syngo® VA48 is scanner platform software that supports the following device features:

    • 1). New system scanner software version synqo® VA48 (SOMARIS/7 VA48) which includes:
      • -Respiratory Analysis of Respiratory Rate & Pitch Adjustment -FAST 3D Reconstruction (FAST 3D Align)
      • -Multiphase reconstruction with extended Field of View
      • -FAST DE Results (Dual Energy PACS-ready images)
      • -FAST contact
      • -Iterative Reconstruction with extended Field of View
      • -OEM Varian RGSC Online Mode
      • -Full 4D Lung Scan
      • -Applications at CT syngo.via client
      • -TrueD 4D Viewer
    • 2). ADMIRE Iterative Reconstruction (Option)
    • 3). iMAR Improved Metal Artifact Reduction (Option)

    There are no modifications to the hardware of the device.

    AI/ML Overview

    This document, a 510(k) Summary for the Siemens SOMATOM Definition AS Open CT system with software version syngo® VA48, primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study demonstrating improved human reader performance with AI assistance. It describes software updates and their verification and validation against technical standards.

    Therefore, many of the requested details, such as specific acceptance criteria for AI performance, clinical study design for improved human reader performance, sample sizes for test sets in an MRMC study, expert qualifications for ground truth in a clinical context, or the effect size of AI assistance on human reader performance, are not explicitly available within this document. This submission is for a computed tomography x-ray system, and the "AI" or "machine learning" components mentioned (e.g., ADMIRE Iterative Reconstruction, iMAR Improved Metal Artifact Reduction) are features of the imaging system and reconstruction algorithms, not typically standalone AI interpretation tools that would undergo an MRMC study in the way a diagnostic AI would.

    Given the information provided, here's what can be extracted and inferred, with limitations noted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a table of performance acceptance criteria in the sense of a diagnostic AI's clinical metrics (e.g., sensitivity, specificity, AUC). Instead, it refers to acceptance criteria for software specifications and conformance to technical standards.

    Acceptance Criterion (Inferred from Document)Reported Device Performance (Inferred from Document)
    Conformance to IEC 60601-2-44"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to IEC 61223-3-5"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to NEMA XR-25"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to IEC 61223-2-6"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to NEMA PS 3.1 3.18 (DICOM)"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to IEC 62304 Ed. 1.0"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to IEC 60601-1"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to ISO 14971 (Risk Management)"The Risk analysis was completed, and risk control implemented, to mitigate identified hazards." (Implies acceptance criteria for risk mitigation were met)
    Conformance to NEMA XR-29"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Conformance to ISO/IEC 10918-1 (JPEG)"SOMATOM Definition AS/AS+ configured with software version syngo® VA48 is designed to fulfill the requirements... The test results show that all the software specifications have met the acceptance criteria."
    Software specifications functionality"The test results show that all the software specifications have met the acceptance criteria."
    Software verification and validation acceptability"Verification and validation testing of the device was found acceptable to support the claims of substantial equivalence."
    Performance as intended"The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document refers to "Performance tests" and "Non-clinical tests (integration and functional)" but does not specify a sample size for a clinical test set of patient data, nor its provenance (country, retrospective/prospective). This type of information would be expected for a diagnostic AI device, not typically for a CT system software update focusing on features like iterative reconstruction or metal artifact reduction.

    3. Number of Experts and Qualifications for Ground Truth:

    Not applicable to this type of submission. There is no mention of human experts establishing ground truth for a diagnostic test set in the context of this 510(k). The "ground truth" here likely refers to technical specifications and expected performance characteristics of the CT system and its software, validated through engineering and phantom testing, rather than clinical interpretation.

    4. Adjudication Method for the Test Set:

    Not applicable. No clinical adjudication process is described as there isn't a stated clinical test set requiring human interpretation for ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC study was explicitly not done or described in this document. This regulatory submission is for a computed tomography x-ray system, specifically a software update (syngo® VA48) that includes features like iterative reconstruction (ADMIRE) and metal artifact reduction (iMAR). These are image processing and acquisition technologies, not AI-driven diagnostic assistance tools designed to change human reader performance in a comparative effectiveness study.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    The document describes "Performance tests" to "test the functionality of the SOMATOM Definition AS Open configured with software version syngo VA48." While not using the term "standalone performance" in the context of a diagnostic AI, the testing described appears to be algorithm-only, focused on the technical performance and functional verification of the software features themselves (e.g., image reconstruction quality, artifact reduction effectiveness) against technical standards. "The results of these tests demonstrate that the subject device performs as intended."

    7. Type of Ground Truth Used:

    The ground truth for this submission appears to be based on:

    • Technical specifications and engineering standards: Conformance to various IEC and NEMA standards (e.g., IEC 60601-2-44, IEC 61223-3-5, NEMA XR-25, DICOM).
    • Internal software specifications and functional requirements: "The test results show that all the software specifications have met the acceptance criteria."
    • Risk analysis and mitigation: "The Risk analysis was completed, and risk control implemented, to mitigate identified hazards." (The "ground truth" here is the identified hazards and their successful mitigation).

    There is no mention of expert consensus, pathology, or outcomes data as a ground truth for clinical performance in this document.

    8. Sample Size for the Training Set:

    Not applicable. This document is not describing a machine learning model that was "trained" on a dataset in the way a diagnostic AI would be. The software updates described (like ADMIRE and iMAR) are based on algorithms and iterative processes, but the traditional concept of a "training set" for a deep learning model isn't presented here.

    9. How Ground Truth for the Training Set Was Established:

    Not applicable, as there is no described training set for an AI model in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1