Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K143041

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-06-04

    (225 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ebony® PTA 0.014" over the wire Catheter is intended for dilatation of lesions in the femoral, popliteal and infrapopliteal arteries. The Ebony® PTA 0.014" over the wire Catheter is contraindicated for use in coronary arteries or neuro-vasculature.

    The Ebony® PTA 0.018" over the wire Catheter is intended for dilatation of lesions in the femoral, iliac, popliteal, infrapopliteal and renal arteries. The Ebony® PTA 0.018" over the wire Catheter is not for use in coronary arteries or neuro-vasculature.

    Device Description

    The Ebony ® PTA 0.014" over the wire catheter is a standard catheter with a balloon near the distal tip and a Y-connector at the proximal end. The guide wire lumen (inner tubing) permits the use of guide wires to facilitate advancement of the catheter to and through the stenosis to be dilated. Maximum guide wire diameter is 0.014" (0.36 mm). The usable catheter length for OTW is 120 cm and 150 cm.

    In order to correctly position the balloon under fluoroscopy, two radiopaque markers are placed on the shaft under the balloon itself, defining its cylindrical area. The catheter includes a smooth, soft and atraumatic tip to facilitate advancement of the catheter through the stenosis. In order to facilitate advancing through the vasculature, a hydrophilic coating is present on the distal shaft and the balloon.

    The Ebony ® PTA 0.014" over the wire catheter is available in 2 usable lengths: 120 and 150 cm; and in the balloon sizes shown in Table 1. The Ebony ® PTA 0.014" over the wire catheter will be supplied sterile and is intended for one time use.

    The Ebony® PTA 0.018" over the wire Catheter is a standard catheter with a balloon near the distal tip and a Y-connector at the proximal end. The guide wire lumen (inner tubing) permits the use of guide wires to facilitate advancement of the catheter to and through the stenosis to be dilated and it ends at the tip of the catheter. Maximum guide wire diameter is 0.018" (0.46 mm).

    The balloon has radiopaque marker(s) to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

    The Ebony® PTA 0.018" over the wire Catheter is available in two catheter lengths and the balloon sizes shown in Table 2 below. The Ebony® PTA 0.018" over the wire Catheter is supplied sterile and is intended for one time use.

    AI/ML Overview

    This document describes the performance testing for the Ebony® PTA 0.014" over the wire catheter and Ebony® PTA 0.018" over the wire catheter.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary states that "all catheter test samples met the acceptance criteria for each of the performance tests." However, it does not explicitly list the quantitative acceptance criteria for each test or the specific numerical results obtained. It only provides a list of the tests performed and a general statement of compliance.

    Test PerformedAcceptance Criteria (Not explicitly stated quantitatively, but implied as "met")Reported Device Performance
    Balloon complianceMet pre-determined criteriaAll samples met criteria
    Balloon burst pressureMet pre-determined criteriaAll samples met criteria
    Balloon fatigueMet pre-determined criteriaAll samples met criteria
    Shaft resistanceMet pre-determined criteriaAll samples met criteria
    Bond strengthMet pre-determined criteriaAll samples met criteria
    Catheter dimensionsMet pre-determined criteriaAll samples met criteria
    Deflation timeMet pre-determined criteriaAll samples met criteria
    Coating verificationMet pre-determined criteriaAll samples met criteria
    Kink testingMet pre-determined criteriaAll samples met criteria
    Deployment testingMet pre-determined criteriaAll samples met criteria
    Guide wire and introducer compatibilityMet pre-determined criteriaAll samples met criteria
    BiocompatibilityPassed tests as per ISO10993 Part 1, 2, 4, 5, 10, 11, 12, ASTM F75600, and 21 CFR 58 (GLP)All materials were found to be biocompatible

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample size used for each performance test. It only states "all catheter test samples."
    • Data Provenance: The tests are described as "in vitro bench tests and analyses." The country of origin for the data is implied to be Natec Medical Ltd. in Mauritius, where the applicant is located. The tests are prospective in nature, as they involve testing the manufactured devices against pre-defined criteria.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is a medical instrument (catheter), and its performance is evaluated through objective engineering and biological bench tests, not through expert interpretation of data or images where "ground truth" would be established by human experts.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the performance is assessed through objective bench testing against pre-determined acceptance criteria, not through expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a medical device (catheter), not a diagnostic AI system or an imaging modality that would involve human readers interpreting cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This is a physical medical device. The performance tests are for the device itself, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for the performance tests is established by industry standards, engineering specifications, and regulatory requirements (e.g., ISO, ASTM, FDA guidelines) for medical device performance. For biocompatibility, it's defined by the passing criteria within the specified ISO and ASTM standards.

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/machine learning algorithm that requires a training set. The performance tests evaluate the device's physical and biological characteristics.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As this is not an AI/machine learning algorithm, there is no "training set" or corresponding ground truth establishment in that context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1