Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K142797

    Validate with FDA (Live)

    Device Name
    I.V. Catheter
    Date Cleared
    2015-06-18

    (262 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device, I.V. Catheter, is intended to be inserted in to a patient's vascular system for short term use (no more than 72 hours) to withdraw blood samples or administer fluid intravenously.

    Device Description

    The proposed devices, I.V. Catheters, are sterile, single use devices intended to be inserted in to a patient's vascular system for short term use to withdraw blood samples or administer fluid intravenously.

    The I.V. Catheter is a closed system, and further available in two combinations of configurations. All of them are sterilized by Ethylene Oxide to achieve a SAL of 106, and packed in a sterility maintenance package. The shelf life of the product is three years.

    AI/ML Overview

    This document is a 510(k) summary for an I.V. Catheter (K142797) submitted by Jiangxi Sanxin Medtec Co., Ltd. It declares substantial equivalence (SE) to a predicate device (K083429).

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established by compliance with recognized international and ASTM standards. The reported device performance is that it complies with these standards.

    Acceptance Criteria (Standard)Reported Device Performance
    ASTM F88-09 (Seal Strength of Flexible Barrier Materials)Device complies
    ASTM F 1140-07 (Internal Pressurization Failure Resistance of Unrestrained Package)Device complies
    ASTM F1929-98(2004) (Detecting Seal Leaks by Dye Penetration)Device complies
    ISO 11135-1:2007 (Ethylene Oxide Sterilization)Device complies (achieves SAL of 10^-6)
    USP <85> (Bacterial Endotoxin Limit)Device complies
    ISO 10993-3:2003 (Biological evaluation - Genotoxicity, carcinogenicity, reproductive toxicity)Device complies (conforms to the requirements of ISO 10993 series Standards)
    ISO 10993-4:2002 A1:2006 (Biological evaluation - Interactions with blood)Device complies (conforms to the requirements of ISO 10993 series Standards)
    ISO 10993-5:2009 (Biological Evaluation - In Vitro Cytotoxicity)Device complies (conforms to the requirements of ISO 10993 series Standards)
    ISO 10993-6:2007 (Biological evaluation - Local effects after implantation)Device complies (conforms to the requirements of ISO 10993 series Standards)
    ISO 10993-7:2008 (Biological evaluation - Ethylene Oxide Residues)Device complies (conforms to the requirements of ISO 10993 series Standards)
    ISO 10993-10:2010 (Biological evaluation - Irritation and skin sensitization)Device complies (conforms to the requirements of ISO 10993 series Standards)
    ISO 10993-11:2006 (Biological evaluation - Systemic toxicity)Device complies (conforms to the requirements of ISO 10993 series Standards)
    ASTM F756-08 (Assessment of hemolytic properties of material)Device complies
    ISO 10555-1:1995/AMD.1:1999/AMD.2:2004(E) (Sterile, single-use intravascular catheters - General requirements)Device complies
    ISO10555-5: 1996/AMD.1:1999(E) (Sterile, single-use intravascular catheters - Over-needle peripheral catheters)Device complies
    ISO594-2:1998 (Conical fittings with 6 % (Luer) taper for syringes, needles)Device complies
    ISO9626:1991/AMD-1:2001 (Stainless steel needle tubing)Device complies

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size used for each non-clinical test. The data provenance is indicated as "Non clinical tests were conducted to verify that the proposed device met all design specifications" (Page 5), suggesting these were performed by the manufacturer, Jiangxi Sanxin Medtec Co., Ltd. The country of origin for the device itself is China, and the tests were likely conducted there or by affiliated entities. The tests are prospective with respect to the submission, performed specifically to demonstrate compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This document describes a 510(k) submission for a medical device (I.V. Catheter). The "tests" here are primarily engineering and biocompatibility evaluations against recognized standards, not diagnostic performance assessments requiring expert ground truth establishment in a clinical sense. Therefore, this information is not applicable and not provided in the document. The standards themselves represent expert consensus on performance requirements.

    4. Adjudication method for the test set

    Similarly to point 3, adjudication methods like 2+1 or 3+1 are typically for diagnostic studies where there's variability in interpretation or a need to resolve discrepancies. For the non-clinical tests listed (e.g., seal strength, sterilization, biocompatibility), the results are generally objective measurements against predefined criteria in the standards. Therefore, an adjudication method is not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This document is for an I.V. Catheter, which is a physical medical device, not an AI software or a diagnostic imaging system. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is not applicable and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    As this is a physical medical device and not an algorithm, a standalone performance study in the context of an algorithm is not applicable and not discussed in the document. The non-clinical tests represent the standalone performance of the physical device.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" is defined by the specifications and limits set forth in the referenced ASTM and ISO international standards. For example, for sterility, the ground truth is "achieve a SAL of 10^-6". For biocompatibility, it's "conforms to the requirement of ISO 10993 series Standards." This is based on established scientific and engineering principles for medical device safety and performance.

    8. The sample size for the training set

    This document describes non-clinical testing of a physical medical device, not a machine learning algorithm. Therefore, the concept of a "training set" is not applicable and no information is provided.

    9. How the ground truth for the training set was established

    As there is no training set for a machine learning algorithm, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1