Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K142128

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-02-23

    (203 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kangaroo™ Enteral Feeding Syrinqe with ENFit Connector delivers nutritional formula to the gastrointestinal system of a patient who is physically unable to eat and swallow. The feeding syringes are intended to be used in clinical or home care settings by users ranging from laypersons to clinicians (under the supervision of a clinician), to administer enteral nutrition. The syringe is intended for all age groups.

    Device Description

    The Kangaroo™ Enteral Feeding Syringe with ENFit Connector consists of disposable enteral feeding syringes that deliver formula to provide nutrition for those who do not have the ability to orally ingest food. The device incorporates a female ENFit connector for connection to an enteral access device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Kangaroo™ Enteral Feeding Syringe with ENFit Connector. It is a submission for a device that is substantially equivalent to a predicate device, focusing on non-clinical testing to demonstrate safety and effectiveness.

    Here's an analysis of the provided information against your request:

    1. A table of acceptance criteria and the reported device performance

    The document lists several tests conducted in accordance with ISO 80369-3 and ISO 80369-20. The "acceptance criteria" are implied by the "Requirement is in:" columns, referring to specific clauses in ISO 80369-3, and the "reported device performance" is generally stated as "The testing demonstrates the proposed devices conform to the requirements of ISO 80963-3." Specific quantitative performance values are not provided in this summary.

    TestAcceptance Criteria (ISO 80369-3 Clause)Reported Device Performance
    Fluid LeakageClause 6.2Conforms to requirements of ISO 80963-3
    Stress CrackingClause 6.3Conforms to requirements of ISO 80963-3
    Resistance to separation from axial loadClause 6.4Conforms to requirements of ISO 80963-3
    Resistance to separation from unscrewingClause 6.5Conforms to requirements of ISO 80963-3
    Resistance to overridingClause 6.6Conforms to requirements of ISO 80963-3
    Disconnection by unscrewingClause 6.7Conforms to requirements of ISO 80963-3
    Dimensional analysisTable B.2 of ISO 80369-3Conforms to Table B.2 of ISO 80963-3
    Minimum through diameter flow testing(Implied performance objective)Exceeds the performance of the predicate device
    Biocompatibility testing(Implied biological safety standards)Demonstrated the biological safety of indirect contact parts
    Stability testing(Implied shelf-life/aging standards)Evaluated properties after accelerated aging
    Risk associated with misconnection(Implied acceptable risk level)Assessed and captured in various reports
    Usability and human factors testing(Implied usability/safety standards)Conducted and captured in a validation study

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for any of the non-clinical tests (e.g., number of syringes tested for fluid leakage, stress cracking, etc.). It also does not explicitly state the data provenance (country of origin, retrospective/prospective). However, given that these are non-clinical engineering and bench tests, "data provenance" in the context of clinical studies (like patient demographics or retrospectivity) is not directly applicable. These tests would be conducted in a laboratory setting by the manufacturer, or a certified testing lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This request is not applicable to the provided document. The study described is a non-clinical performance evaluation of a physical medical device (enteral feeding syringe) against engineering standards (ISO 80369-3/20). There is no "ground truth" in the sense of expert medical diagnosis or interpretation that would require a panel of experts. The "ground truth" here is adherence to the specified technical standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this involves non-clinical engineering testing against predefined standards, not a diagnostic or interpretive task requiring adjudication by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a study involving human readers, AI, or diagnostic imaging. It is a submission for a physical medical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests is adherence to the specified international standards, specifically ISO 80369-3: Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications and ISO 80369-20: Common test methods. This is equivalent to established, recognized engineering and safety standards.

    8. The sample size for the training set

    Not applicable. There is no concept of a "training set" for this type of non-clinical device testing.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set, this question is not relevant.

    In summary, the provided document describes a 510(k) submission for a physical medical device (enteral feeding syringe). The "study" refers to a series of non-clinical engineering and bench tests designed to demonstrate conformity to ISO standards for small-bore connectors. The questions regarding expert consensus, clinical ground truth, training/test sets, and AI/human reader performance are not applicable to the nature of this particular device approval process.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1