Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161045
    Device Name
    Monoject
    Manufacturer
    Date Cleared
    2016-06-17

    (65 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monoject™ Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

    Device Description

    The Monoject™ Enteral Feeding Syringe with ENFit Connector is a disposable enteral feeding syringe provided in sizes of 1mL and 3mL. The device specific for low dose applications incorporates a female ENFit Syringe tip with an internal tip lumen for connection to an enteral access device with male ENFit connector.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Monoject™ Enteral Feeding Syringe with ENFit Connector. As such, it details the non-clinical testing performed to demonstrate substantial equivalence to legally marketed predicate devices, rather than a clinical study proving the device meets specific performance criteria against a ground truth.

    Therefore, many of the requested sections related to clinical studies, expert-established ground truth, and human reader performance will not be applicable or available from this document.

    Here's an analysis of the provided information based on your request:


    Acceptance Criteria and Device Performance (Non-Clinical)

    Since this is a non-clinical evaluation for a medical device (enteral feeding syringe), the "acceptance criteria" are based on the requirements of recognized industry standards. The "device performance" is reported as conformity to these standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Defined in AAMI /CN3 (PS):2014)Test Method (Defined in ISO 80369-20)Reported Device Performance
    Fluid Leakage (Clause 6.2)Annex CConforms to requirements
    Stress Cracking (Clause 6.3)Annex EConforms to requirements
    Resistance to separation from axial load (Clause 6.4)Annex FConforms to requirements
    Resistance to separation from unscrewing (Clause 6.5)Annex GConforms to requirements
    Resistance to overriding (Clause 6.6)Annex HConforms to requirements
    Disconnection by unscrewing (Clause 6.7)Annex IConforms to requirements
    Dimensional analysis for low dose ENFit connector- (AAMI /CN3 (PS):2014 Table B.2)Conforms to criteria
    Dosing Accuracy of low dose tip ENFit connector-Verified
    Biocompatibility-Demonstrated biological safety
    Stability-Properties maintained after accelerated aging (supports labeling)
    ISO 7886-1 Sterile Hypodermic Syringes (Part 1)-Conforms to requirements

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes used for each individual non-clinical test (e.g., number of syringes tested for fluid leakage). It generally states that "Testing performed on the Monoject™ Enteral Feeding Syringes with low dose tip ENFit Connectors included the items listed below..."
    • Data Provenance: The tests are non-clinical (laboratory/bench testing) performed by the manufacturer, Covidien, based in Mansfield, MA, USA. The data is thus internally generated non-clinical test data. It is neither retrospective nor prospective in the clinical sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. For non-clinical device testing verifying compliance with standards, "experts" in the sense of clinical specialists establishing a "ground truth" are not relevant. The "ground truth" is defined by the technical specifications and requirements of the referenced ISO and AAMI standards.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study requiring adjudication of expert opinions. Device performance is assessed directly against the objective criteria of the specified standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document describes a 510(k) submission for an enteral feeding syringe, which is a physical medical device. It does not involve AI, image analysis, or human readers, so an MRMC comparative effectiveness study is not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This is not a software device or an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for these tests is the set of objective, measurable requirements specified in the following international and industry standards:

    • AAMI /CN3 (PS):2014, Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications.
    • ISO 80369-20, Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods.
    • ISO 7886-1, Sterile Hypodermic Syringes for Single Use - Part 1: Syringes for Manual Use.
    • Biocompatibility standards (not explicitly listed, but implied by "Biocompatibility testing has demonstrated the biological safety").

    8. Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI-based device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set. The relevant "ground truth" for non-clinical performance evaluation is established by international and industry consensus standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1