510(k) Data Aggregation

    K Number
    K141618

    Validate with FDA (Live)

    Date Cleared
    2014-07-08

    (21 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Arrow VPS Stylet PLUS and VPS Console (together the VPS System) is to quickly and accurately guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction.

    The CG+ Arrow PICC Catheters powered by Arrow VPS Stylet PLUS permit venous access to the central circulation through a peripheral vein, with the guidance of the VPS system.

    The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffess for use in placement of the cather , intravascular capability for ECG delection and intravascular ultrasound for catheter guiding and positioning. The VPS Style, when used with the VPS Console, provides real-time catherer in location information by using the patient's physiological (cardiae electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location.

    The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.

    Limiting but not contraincied situations for this technique are in patients where allerations of cardias the presentation of the P-waye as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the Pwave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

    The Pressure Intectable PICC with Chlorag+ard Aniithombogenic Technology is indicated for short-lerm or long-term peripheral access to the central venous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure of pressure of pressure injection equipment used with the pressure injectable PICC catheter may not exceed 300 psi.

    Chlorag+ard Technology treatment on the external surface of the entire fluid patiway of the catheler has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheler surfaces. Antimicrobial and antitlersmoogenic effectiveness were evaluated using in viro and in vive test methods and no correlation between these test methods and clinical outcome has currently been asseriained. It is not intented of existing infections or vein thrombosis.

    The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffess for use in placement of the catheter, intravascular capability for ECG detection and intravascular ultrasound for catherer guiding and positioning. The VPS Style, when used with the VPS Console, provides real-time catheter ip location information by using the patient's physiological (cardial activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location.

    The VPS System is indicated for use as an allemative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location.

    Limiting but not contraindicated situations for this technique are in patientions of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachyeardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the Pwave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

    Device Description

    The Arrow Vascular Positioning System (VPS) Stylet PLUS (hereafter referred to as Arrow VPS Stylet PLUS) is designed for use with a VPS Console to guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavoatrial junction. The CG+ Arrow PICC powered by Arrow VPS Stylet PLUS provides the user with an Arrow VPS Stylet PLUS aiready loaded into a central catheter. (The Arrow VPS Stylet Plus and the Arrow VPS Stylet Plus included in the CG+ Arrow PICC powered by Arrow VPS Stylet Plus are exactly the same stylet.)

    The subject device, the Arrow VPS Stylet PLUS has the following characteristics:

    • 6 ft overall length .
    • < 0.021" outer diameter over working length of polyimide with polytetrafluoroethylene . (PTFE) heat shrink
    • Intravascular electrocardiogram (ivECG) signal sensing stainless steel wire with exposed . portion at the distal end
    • Doppler transducer connected to coaxial cable at the distal end .
    • Coaxial cable and ivECG wire attached to connector at the proximal end to be plugged in . to VPS Console or extension cable (that in turn connects to the VPS Console)
    • Tuohy-Borst adapter ●
    • Marking accessory .

    The Arrow VPS Stylet PLUS is the stylet portion of a vascular positioning system designed to be used with the VPS Console and a market-available catheter. The Arrow VPS Stylet PLUS is a polyimide tube with a fluoropolymer (PTFE) heat shrink. The tubing contains a Doppler sensor on a coaxial cable and an intravascular electrocardiogram (ivECC) signal sensing stainless steel wire. The Doppler sensor and the exposed portion of the ivECG are located at the distal end of the stylet and are used to detect and transmit physiological information to the VPS Console for analysis. The proximal end contains a connector to the VPS Console or to an extension cable that in turn connects to the VPS Console.

    The CG+ Arrow PICC Powered by Arrow VPS Stylet PLUS is the Arrow VPS Stylet PLUS preloaded into a 4.5 Fr 1-Lumen, 5.5 Fr 2-Lumen, and 6 Fr 3-Lumen, 40-55 cm pressure injectable antimicrobial and antithrombogenic Peripherally Inserted Central Catheter (PICC).

    The CG+ Arrow PICC is a short-term or long-term, single use catheter designed to provide access to the central venous system. It consists of a non-tapered, radiopaque polyurethane extruded catheter body with a softer, contoured Blue Flex Tip. The catheter can be used for the injection of contrast media. The external catheter body and the internal fluid path of the device are treated with Chlorag+ard, a Chlorhexidine-based coating technology.

    AI/ML Overview

    This document describes a 510(k) submission for the Arrow VPS Stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS. It is a submission for substantial equivalence to previously cleared devices, meaning the focus of the study is to demonstrate that the new device performs as well as the predicate device(s), rather than establishing new clinical effectiveness. Therefore, the information provided below is extracted based on the nature of this type of submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    For a 510(k) submission seeking substantial equivalence for a medical device that has minor changes from existing predicate devices, the "acceptance criteria" are typically defined by demonstrating that the modified device performs comparably to the predicate device and meets established engineering and biocompatibility standards. Clinical performance is generally not evaluated with new efficacy endpoints unless the changes introduce new risks or intended uses. In this case, the acceptance criteria are largely implied by the PASS results in the nonclinical testing, indicating the device meets the set standards for each test.

    Acceptance Criteria (Implied by Test)Reported Device Performance
    Mechanical strength (Tensile, Torque, Tip Flexibility, Flexing, Fracture)PASS
    Corrosion resistancePASS
    Component compatibility & grip strength (Marking accessory, Tuohy-Borst Adapter)PASS
    Leak integrity (Tuohy-Borst)PASS
    Physical characteristics metPASS
    X-ray DetectabilityPASS
    Electrical Performance (Hi-Pot, Continuity, Doppler Bandwidth, Capacitance, Sensitivity, Sensitivity Stability)PASS
    Biocompatibility (Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Hemocompatibility)PASS
    Electrical Safety and Electromagnetic CompatibilityPASS
    Catheter Compatibility (Simulated Use Insertion/Removal, Force to Remove stylet)PASS
    Substantial Equivalence to Predicate DeviceAchieved

    2. Sample Size Used for the Test Set and Data Provenance

    Since this is a 510(k) submission for substantial equivalence based on nonclinical testing, there is no "test set" of clinical data in the typical sense (i.e., a set of patient data used to evaluate clinical accuracy or performance metrics against a ground truth). The testing conducted was primarily bench testing and laboratory testing on the device itself.

    • Sample Size: The document does not specify exact sample sizes for each nonclinical test (e.g., how many stylets were tested for tensile strength). These are typically determined by engineering and quality control standards.
    • Data Provenance: The data is generated from laboratory and bench testing performed on the Arrow VPS Stylet PLUS and the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS. The origin of the data is the manufacturer's internal testing. It is not patient or country-specific. The testing was performed after preconditioning with Et.O sterilization, ISTA simulated environmental conditions, and one year of accelerated aging.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable to this submission. "Ground truth" in the context of expert consensus is typically relevant for studies evaluating diagnostic accuracy or clinical outcomes where human experts interpret data. This 510(k) submission focuses on demonstrating the device's technical specifications and safety through nonclinical testing, not on clinical interpretation or an AI algorithm's performance against human experts.

    4. Adjudication Method for the Test Set

    This section is not applicable as there was no clinical test set requiring adjudication by experts. The "adjudication" in this context would be the successful completion of each nonclinical test according to predefined engineering and safety standards, as indicated by the PASS results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed to evaluate the impact of an AI system on human reader performance, which is not the focus of this 510(k) submission for mechanical and electrical device changes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device described (Arrow VPS Stylet PLUS) is a physical medical instrument used to guide catheters and provides real-time information to a "VPS Console." While the VPS Console likely contains algorithms to process the ivECG and Doppler signals, this submission focuses on the stylet itself. Therefore, a "standalone algorithm performance" study in the context of AI models is not relevant or described in this document. The electrical performance tests (Hi-Pot, Continuity, Doppler Bandwidth, Capacitance, Sensitivity, Sensitivity stability) assess the functionality of the device's signal-sensing components, which are essential for the console's operation.

    7. The Type of Ground Truth Used

    The "ground truth" for the nonclinical testing performed can be defined as:

    • Engineering Specifications and Standards: Each test (e.g., tensile strength, electrical performance, biocompatibility) has predefined, objective acceptance criteria (e.g., a minimum tensile strength, a specific range for electrical resistance, absence of cytotoxic effects) based on international standards (e.g., ISO, ASTM, IEC) and the manufacturer's own design specifications.
    • Predicate Device Performance: A key aspect of "substantial equivalence" in a 510(k) is demonstrating that the new device performs "as well as" the predicate device(s). Therefore, the performance characteristics of the legally marketed predicate devices serve as an implicit "ground truth" or benchmark for comparison in the nonclinical testing.

    8. The Sample Size for the Training Set

    This section is not applicable. This submission describes a physical medical device and its nonclinical testing to demonstrate substantial equivalence, not the development or training of an AI algorithm. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable as there is no training set for an AI algorithm.