Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K141253

    Validate with FDA (Live)

    Date Cleared
    2015-01-09

    (240 days)

    Product Code
    Regulation Number
    888.3070
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The New Era Orthopaedics Polyscrew Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), spinal tumor, pseudarthrosis and failed previous fusion.

    The New Era Orthopaedics Polyscrew Pedicle Screw System is also intended for non-cervical pedicle screw fixation for the following indications: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It is also intended for the following indications: trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The New Era Orthopaedics Polyscrew Pedicle Screw System is a multiple component, posterior spinal fixation system which consists of pedicle screws, rods and cross links. All of the components are available in a variety of sizes to more closely match the patient's anatomy.

    Materials:

    Ti-6Al-4V per ASTM F136 CoCr per ASTM F1537

    AI/ML Overview

    The provided document describes the New Era Orthopaedics Polyscrew Pedicle Screw System, a Class II medical device. As a Class II device approved through the 510(k) pathway, its approval hinges on demonstrating "substantial equivalence" to a predicate device, not necessarily on meeting specific numerical acceptance criteria in the same way a novel device might.

    Therefore, the "acceptance criteria" for this device are implicitly tied to the performance of its predicate devices, as demonstrated through non-clinical testing. The study provided focuses on proving this substantial equivalence through mechanical testing.

    Here's the breakdown of the information requested based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: For 510(k) submissions, explicit numerical "acceptance criteria" and "reported device performance" in the typical sense of a novel device study are often not stated. Instead, the performance is reported as meeting or being equivalent to the predicate device, which implies satisfying the predicate's established safety and effectiveness.

    Acceptance Criteria (Implicit)Reported Device Performance
    Mechanical Performance: The device system (New Era Orthopaedics Polyscrew Pedicle Screw System) must demonstrate equivalent or superior mechanical performance in terms of static and dynamic compression, static torsion, and dissociation compared to its predicate devices, as evaluated using recognized ASTM F1717 standards. This implies that the device should withstand similar forces and exhibit comparable stability and durability to established pedicle screw systems without premature failure or degradation.Non-Clinical Test Summary: The New Era Orthopaedics Polyscrew Pedicle Screw System underwent the following tests: - Static and dynamic compression per ASTM F1717 - Static torsion per ASTM F1717 - Dissociation testing The results of this testing indicate that the New Era Orthopaedics Polyscrew Pedicle Screw System is equivalent to predicate devices.
    Material Compatibility: The materials used (Ti-6Al-4V per ASTM F136; CoCr per ASTM F1537) must be biocompatible and suitable for implantation in the human body, consistent with materials used in predicate devices.Materials: Ti-6Al-4V per ASTM F136 and CoCr per ASTM F1537 were used, which are standard, biocompatible materials for such implants and are consistent with those found in predicate devices.
    Functional Equivalence: The device's design, intended use, and general operational principles must be substantially equivalent to the predicate devices.Substantial Equivalence Claim: The New Era Orthopaedics Polyscrew Pedicle Screw System is claimed to be substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety, and performances.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a numerical sample size for the non-clinical mechanical tests (static/dynamic compression, static torsion, dissociation). For mechanical testing of medical devices, sample sizes are typically determined by statistical requirements to achieve confidence in the measurements, but are not usually reported as a single "test set sample size" in the same way as a clinical study. Instead, multiple units of each component or assembly configuration would be tested to derive statistical data (e.g., mean, standard deviation, and comparisons to predicate device performance limits).
    • Data Provenance: The data provenance is from non-clinical laboratory testing. The document does not specify the country of origin for the testing, but it is typically conducted in controlled laboratory environments in accordance with international (ASTM) standards. This is not retrospective or prospective in a clinical sense, as it refers to laboratory experiments.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: This question is not applicable to this type of submission. Ground truth, in the context of device performance, is established through standardized engineering tests (e.g., ASTM F1717) and comparison to established predicate device performance and regulatory standards, not through expert consensus on qualitative data.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: This question is not applicable. Mechanical tests are quantitative and objective; their outcomes are measured directly based on physical properties and responses to applied forces, not through an adjudication process like one might find in a clinical trial or image interpretation study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret results, comparing their performance with and without AI assistance. The New Era Orthopaedics Polyscrew Pedicle Screw System is a physical implant, not a diagnostic tool requiring human interpretation of results in this manner.
    • Effect Size of Human Readers: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: This question is not applicable. This device is a physical pedicle screw system, not an algorithm or AI software. Therefore, there is no "standalone performance" in the context of an algorithm. Its performance is its mechanical integrity and function as a physical implant.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for this device's performance is established by objective mechanical performance standards (e.g., ASTM F1717) and comparison to the established performance of legally marketed predicate devices. The implicit ground truth is that the device must be safe and effective, similar to existing devices, as demonstrated by meeting standardized mechanical testing requirements and being made of recognized biocompatible materials.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: This question is not applicable. This device is not an AI/algorithm that requires a "training set." The engineering and design process for such a physical device involves iteration and testing, but not in the sense of a machine learning training set.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: This question is not applicable as there is no "training set" for a physical implant device in the context of this submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1