510(k) Data Aggregation

    K Number
    K140809

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-23

    (83 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be used for aspiration during flexible endoscopy.

    Device Description

    The Cliq Aspirator model DV-300 is an AC-powered high-vacuum / low-flow suction unit used for endoscopic aspiration. The DV-300 creates a negative pressure (vacuum) that draws fluids through disposable tubing, and into a collection container. The fluids are trapped within the collection container for proper disposal. The device is comprised of a maintenance-free pump unit; power cord, on/off switch, pressure relief valve, pressure adjustment knob, pressure gauge, microbial filter, and intermediate tubing.

    The Cliq Aspirator: model DV-300 requires 110-120 V and 60 Hz. It has Class II with Type BF applied part protection against electric shock. The DV-300 suction pump operates via a piston pump type, and has a maximum vacuum pressure of 620 mm Hg. The pump produces a flow rate of up to 18 Liters per minute.

    AI/ML Overview

    The Cliq Aspirator Model DV-300 underwent performance testing to establish substantial equivalence to the predicate device, Olympus Suction Pump Model KV-5. The evaluation focused on electrical safety, electromagnetic compatibility, and overall performance, comparing key specifications between the two devices.

    1. Acceptance Criteria and Reported Device Performance

    CriterionAcceptance Criteria (Predicate Device: Olympus Suction Pump Model KV-5)Reported Device Performance (Cliq Aspirator Model DV-300)Meets Acceptance Criteria
    Electrical Requirements100-120 V, 50/60 Hz110-120 V, 60 HzYes
    Protection against electric shockClass I with Type BF applied partClass II with Type BF Applied partNo (Different Class)
    Maximum Vacuum Pressure638 mm Hg620 mm HgYes (Comparable)
    Vacuum Pump TypePistonPistonYes
    FlowUp to 20 L/minUp to 18 L/minYes (Comparable)
    Sound LevelNot known (for predicate)53 dBANA (No direct comparison)
    Weight12.7 kg3.5 kgNo (Significant Difference)
    Dimensions31 x 25.5 x 22 cm30 x 16.5 x 19 cmYes (Comparable)
    Operating Environment (Temperature)10 - 40 °C0 - 40 °CYes (Wider Range)
    Operating Environment (Humidity)0 - 95 %0 - 90 %Yes (Comparable)
    Operating Environment (Atm. Pressure)70 - 106 kPa70 - 106 kPaYes
    Storage Environment (Temperature)-40 - 70 °C-20 - 50 °CNo (Narrows Range)
    Storage Environment (Humidity)0 - 95 %0 - 95 %Yes
    Storage Environment (Atm. Pressure)23.5 - 106 kPa50 - 106 kPaYes (Comparable)
    FilterMicrobial, hydrophobicMicrobial, hydrophobicYes
    Suction Tube length2 m1.8 mYes (Comparable)
    Filter Tube length900 mm250 mmNo (Significant Difference)
    Power CableYesYesYes
    Suction Jar Volume2 liter (one liter optional)1.5 L or 2.4 LYes (Comparable)
    Performance Standards adherenceEN IEC 60601-1, EN IEC 60601-1-2, EN ISO10079-1, UL 2601-1, CAN/CSA Std. No. C22.2 No.601.1-M90EN IEC 60601-1, EN IEC 60601-1-2, EN ISO10079-1Yes (subset met)

    Note on "Meets Acceptance Criteria": The statement "Yes (Comparable)" or "Yes (Wider Range)" indicates that the Cliq Aspirator's performance is either directly equivalent or within an acceptable range compared to the predicate device, supporting substantial equivalence. Differences in Class protection, weight, and filter tube length while noted, were evidently not deemed critical enough by the FDA to prevent substantial equivalence, given the overall conclusion. The "Not known" for predicate sound level means no direct comparison could be made for that specific metric.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state a specific "test set" sample size for a particular number of devices or clinical cases. The testing reported is related to general device performance, electrical safety, and electromagnetic compatibility standards. These types of tests typically involve a sample of manufactured devices (e.g., a few units from a production run) rather than patient data.

    • Data Provenance: The testing was conducted in the context of regulatory submission for a device manufactured by Ding Hwa Co., Ltd in Taiwan. The tests are likely laboratory-based and performed by the manufacturer or a contracted testing facility to meet international standards. No information is provided regarding country of origin of data in terms of patient population or clinical studies, as this is not a clinical performance study.
    • Retrospective or Prospective: Not applicable, as this is not a clinical study involving retrospective or prospective data collection from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. The "ground truth" for this type of device (an aspirator) is its adherence to predefined engineering specifications and safety standards. This is objectively measured and verified according to scientific and engineering principles, not through expert consensus on clinical findings. Therefore, no experts in a clinical sense were used to establish ground truth for a test set.

    4. Adjudication Method for the Test Set:

    Not applicable. As noted above, the "tests" refer to measurements against engineering and safety standards, not clinical assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No. This document describes the regulatory submission for a medical device (an aspirator pump), not an AI/CAD system. Therefore, an MRMC comparative effectiveness study, which is relevant for evaluating human reader performance with and without AI assistance in interpretation tasks, was not performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    No. This is a physical, electromechanical device, not an algorithm or AI. Standalone performance as typically understood for AI was not applicable or performed.

    7. The Type of Ground Truth Used:

    The ground truth used was adherence to established international and national engineering standards for medical devices, specifically:

    • EN ISO 10079-1:2009: Electrically powered suction equipment safety requirements.
    • IEC/EN 60601-1-2:2007 (3rd Ed.): Electromagnetic Compatibility.
    • IEC 60601-1: General basic safety and essential performance of medical electrical equipment.
    • Performance Verification Testing: Testing against predefined functional specifications (e.g., maximum vacuum pressure, flow rate).

    The comparison against the predicate device (Olympus Suction Pump Model KV-5) also served as a basis for evaluating substantial equivalence in performance characteristics.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/machine learning device that utilizes a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As this device does not use a training set, the establishment of ground truth for such a set is irrelevant.