    Device Name
    INSERTION TOOL
    Date Cleared
    2014-04-16

    (29 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insertion Tool is used to facilitate the introduction of a guidewire during general intravascular procedures.

    Device Description

    The Insertion Tool is a sterile accessory device indicated to facilitate the introduction of a guidewire into the lumen of a guiding catheter, without damage to the guidewire's distal tip.

    It is designed to accommodate guidewires with diameters from 0.010" to 0.018" and is composed of a rigid funnel known as the hub which is attached to a stainless steel hypotube. The interior of the hub is designed to allow a guidewire with multiple bends at the distal tip to be directed easily into the smaller diameter stainless steel hypotube.

    AI/ML Overview

    This document is a 510(k) summary for an "Insertion Tool" and does not relate to an AI/ML device. Therefore, the requested information about acceptance criteria, study details, human-in-the-loop performance, and training/test set specifics for an AI/ML model is not applicable and cannot be extracted from the provided text.

    The document describes a medical device, its intended use, and comparison to predicate devices, focusing on design verification and validation through bench tests. There is no mention of AI/ML algorithms or their performance in this submission.