Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140637

    Validate with FDA (Live)

    Date Cleared
    2014-03-26

    (14 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peregrine System is intended for the infusion of diagnostic and therapeutic agents into the perivascular area.

    Device Description

    The Peregrine System Infusion Catheter is a percutaneous catheter designed to deliver diagnostic and therapeutic agents through a vessel wall and into the perivascular space. The catheter contains three distal needles which are deployed using the control handle. Fluids are administered through the proximal injection lumen in the handle, which delivers the fluid through the needles at the distal end of the device. The micro-needles and the guide tubes are radiopaque for fluoroscopic visibility. The device is intended for vessels 5-7 mm in diameter and is compatible with guide catheters of at least 7F.

    AI/ML Overview

    The provided text describes the 510(k) summary for the ABLATIVE SOLUTIONS, INC. PEREGRINE SYSTEM™ INFUSION CATHETER. It focuses on the device's technological characteristics and non-clinical performance data to demonstrate substantial equivalence to predicate devices, rather than establishing acceptance criteria or reporting performance in a way that aligns with typical AI/algorithm-related studies.

    Therefore, many of the requested elements for an AI device study are not present in this document.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in the way one would for an AI algorithm. Instead, it lists various non-clinical tests conducted to demonstrate the device's functional integrity and safety.

    Acceptance Criteria CategorySpecific Tests ConductedOutcome/Reported Performance
    BiocompatibilityCytotoxicityMet ISO 10993 standards
    SensitizationMet ISO 10993 standards
    Irritation/IntracutaneousMet ISO 10993 standards
    Acute systemic toxicityMet ISO 10993 standards
    Materials mediated pyrogenicityMet ISO 10993 standards
    Hemocompatibility hemolysis (direct contact)Met ISO 10993 standards
    Hemocompatibility hemolysis (extract)Met ISO 10993 standards
    Complement activation of C3a and SC5b-9Met ISO 10993 standards
    Thrombogenicity in SheepMet ISO 10993 standards
    Functional PerformanceVisual/Dimensional InspectionFunctions as intended
    Air Ingress/Negative CollapseFunctions as intended
    Tensile StrengthFunctions as intended
    Liquid Leakage under Pressure/Leakage at HubFunctions as intended
    Flexibility and KinkFunctions as intended
    Tip StiffnessFunctions as intended
    Guide Tube Deployment ForceFunctions as intended
    Catheter TorqueFunctions as intended
    Guidewire TorqueFunctions as intended
    Corrosion ResistanceFunctions as intended
    Chemical CompatibilityFunctions as intended
    Simulated Use TestingFunctions as intended
    Device IntegrityPackaging/Simulated Shipping TestingFunctions as intended
    Sterilization validationFunctions as intended
    Clinical PerformanceIn-vivo testingFunctions as intended

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a comprehensive non-clinical testing program as outlined in the "Non-Clinical Performance Data" section.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated in terms of individual units tested for each specific test. The document mentions "testing performed on the proposed device," implying an adequate number of devices were tested per standard practice for each test type, but specific quantities are not provided.
    • Data Provenance: The tests are described as "non-clinical bench testing, simulated-use testing and in-vivo testing." This indicates the data is generated through laboratory experiments and animal (sheep) studies, not human clinical data. The country of origin for these labs is not specified. It is prospective in the sense that the tests were designed and executed to evaluate the new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the document describes the testing of a medical device (catheter), not an AI algorithm requiring expert-established ground truths on medical images or patient data. The "ground truth" here is adherence to engineering standards (ISO 10993, ISO 10555-1) and functional performance criteria. The experts involved would be engineers, scientists, and technicians conducting the tests and interpreting results against established standards. Their qualifications are not detailed.

    4. Adjudication method for the test set

    Not applicable for a device testing report of this nature. Adjudication methods like 2+1 or 3+1 typically refer to conflicts in expert labelers for AI ground truth, which is not relevant here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to a physical medical device (infusion catheter), not an AI system. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm; it's a medical device.

    7. The type of ground truth used

    The "ground truth" for this device's performance evaluation is established by:

    • Adherence to International Standards: ISO 10993 (Biocompatibility) and ISO 10555-1 (Catheter standards).
    • Functional Performance Specifications: Predefined engineering specifications for mechanical, chemical, and fluid delivery characteristics.
    • In-vivo observations: Performance in animal models (e.g., thrombogenicity in sheep) against expected physiological responses.

    8. The sample size for the training set

    Not applicable. This is not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1