Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K140607

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-08

    (59 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The aap LOQTEQ® Olecranon Plates are indicated for fixation of fractures, osteotomies and nonunions of the olecranon, particularly in osteopenic bone.

    Device Description

    The aap LOQTEQ® Olecranon Plate consists of bone plates and bone screws, to be implanted by a surgeon in order to achieve an internal fixation of bone fragments typically after fractures. If the plates are used in conjunction with locking screws, a so called internal fixator will be realized (internal fixation).
    The aap LOQTEQ® Olecranon Plate System consists of:
    • LOQTEQ® Olecranon Plate (left and right)
    to be used with
    • LOQTEQ® Cortical Screw 2.7, small head T8, self-tapping
    • Cortical Screw 2.5, small head T8, self-tapping
    • LOQTEQ® Cortical Screw 3.5, T15, self-tapping
    • Cortical Screw 3.5, self-tapping
    • Set of Instruments aap LOQTEQ® Olecranon Plate

    AI/ML Overview

    The provided documentation describes the aap LOQTEQ® Olecranon Plate, a metallic bone fixation appliance. The study conducted to demonstrate its acceptance criteria focuses on non-clinical mechanical testing to establish substantial equivalence to predicate devices.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical performance of the aap system to be substantially equivalent to predicate devices.Customized 4-point bending tests (static and dynamic) based on ASTM F382-99 demonstrated substantial equivalence with respect to mechanical performance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the specific number of aap LOQTEQ® Olecranon Plates or predicate devices tested for the mechanical evaluation. It refers to "tests" and "test results" in the plural, implying multiple tests were conducted, but the exact sample dimension is not provided.
    • Data Provenance: The data is from non-clinical testing conducted by the manufacturer, aap Implantate AG, in Germany. This is a prospective evaluation of the device's mechanical properties.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This question is not applicable as the study described is a non-clinical mechanical testing study, not a study involving human diagnosis or interpretation where expert ground truth would be established. The "ground truth" here is the physical performance data against established ASTM standards.

    4. Adjudication Method for the Test Set

    • This question is not applicable for a non-clinical mechanical testing study. Adjudication methods are typically relevant for studies involving human interpretation or clinical outcomes where discrepancies need to be resolved.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • This question is not applicable. The provided document describes a medical device (bone plate), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human readers with and without AI assistance was not conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This question is not applicable. As mentioned above, the device is a physical bone plate, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • The "ground truth" for this non-clinical study is defined by established mechanical testing standards, specifically ASTM F382-99, and the performance of the legally marketed predicate devices. The acceptance criterion for the new device was to demonstrate "substantial equivalence" to the mechanical performance of these predicate devices.

    8. The Sample Size for the Training Set

    • This question is not applicable. There is no "training set" in the context of this non-clinical mechanical testing study. Training sets are relevant for machine learning or AI models.

    9. How the Ground Truth for the Training Set Was Established

    • This question is not applicable for the same reasons as #8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1