510(k) Data Aggregation

    K Number
    K140539

    Validate with FDA (Live)

    Date Cleared
    2014-06-09

    (97 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aero Blue* Performance Surgical Gowns are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Aero Blue* Performance Surgical Gowns meet the Level 3 requirements of the AAMI PB70: 2012 Liquid Barrier classifications. The Aero Blue* Performance Surgical Gowns are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and Ethylene Oxide (EtO) sterilization.

    Device Description

    A Spunbond/Film/Spunbond/Meltblown/Spunbond design (SFSMS) that provides AAMI Level 3 liquid barrier protection in the critical zone of the gown. The back of the gown is a SMS (spunbond/meltblown/spunbond fabric that is air-breathable and provides AAMI Level 1 liquid barrier protection.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Aero Blue* Performance Surgical Gowns

    This document outlines the acceptance criteria and the study that demonstrates the Aero Blue* Performance Surgical Gowns meet these criteria, based on the provided 510(k) summary (K140539).

    1. Acceptance Criteria and Reported Device Performance

    The Aero Blue* Performance Surgical Gowns are designed to meet the Level 3 requirements of the AAMI PB70: 2012 Liquid Barrier classifications in the critical zone, and Level 1 in the non-critical back zone. The non-clinical testing performed demonstrated compliance with these and other relevant standards.

    Standard or ReferenceTest MethodAcceptance Criteria (Implied by standard and intended use)Reported Device Performance
    Standard for the Flammability for Clothing Textiles16 CFR 1610Meet Class 1 Flammability requirements.Pass
    ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5CytotoxicityNo significant cytotoxic effects.Pass
    ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10Skin Irritation StudyNo significant skin irritation.Pass
    ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10Sensitization TestNo significant sensitization.Pass
    ISO 10993-7:2008, Ethylene Oxide sterilization residualsEO residualsEO residuals within acceptable limits for a surface device with limited contact with breached or compromised surfaces.Pass
    Laser Ignition ResistanceISO 11810-1 (2005)Demonstrate resistance to laser ignition.Pass
    Spray Impact (critical zones)AATCC 42:2007Demonstrate water resistance in the critical zones consistent with AAMI Level 3.Pass
    Liquid Barrier PerformanceAAMI PB70:2012 Level 3Critical Zone: Meet Level 3 liquid barrier performance (SFSMS design). (Implied: Resist liquid penetration under specific conditions, including spray impact and hydrostatic pressure.) Non-Critical Zone (back): Meet Level 1 liquid barrier performance (SMS fabric). (Implied: Provide minimum liquid barrier protection.)Pass
    Hydrostatic Head (critical zones)AATCC 127:2008Demonstrate water resistance under hydrostatic pressure in the critical zones consistent with AAMI Level 3.Pass
    Spray Impact (non-critical zones)AATCC 42:2007Demonstrate water resistance in the non-critical zones consistent with AAMI Level 1.Pass
    Grab Tensile, Peak Stretch, and Peak Energy - NonwovensASTM D5034 (2009)Possess adequate tensile strength for intended use.Pass
    Abrasion Resistance and Surface Bonding of SMS LaminatesWSP 20.5 (2008)Demonstrate adequate abrasion resistance and surface bonding.Pass
    Synthetic Blood PenetrationASTM-1670-08 (2008)Resist penetration by synthetic blood in the critical zone.Pass
    Mass Per Area (Basis Weight) of MaterialsD3776 (2009)Maintain consistent fabric basis weight as specified.Pass
    Water Vapor Transmission Rate Through Nonwovens and Plastic FilmsWSP 70.4 (2008)Demonstrate appropriate water vapor transmission rate for breathability.Pass
    Degree Peel Strength of Laminated Nonwovens - Raw MaterialsSTM-00197(2010)Achieve adequate peel strength for laminated components.Pass
    Resistance to Linting Dry Particle GenerationINDA WSP 160.1 (2009)Demonstrate resistance to linting and dry particle generation.Pass

    Note: The document explicitly states "All results of testing met acceptance criteria." This implies that the device performed within the established limits for each test.

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not specify the exact sample sizes used for each individual non-clinical test. It lists the standards and test methods applied. For instance, for liquid barrier and blood penetration, it refers to standards like AAMI PB70:2012 and ASTM F-1670-08. These standards typically define minimum sample sizes for testing to ensure statistical validity.

    The data provenance is retrospective, as the tests were conducted on the manufactured Aero Blue* Performance Surgical Gowns to demonstrate compliance with established standards. The country of origin of the data is not explicitly stated but is implicitly from the manufacturer's testing facilities or accredited third-party labs where these standards are tested, likely within the US given the submission to the FDA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This section is not applicable for this medical device. The "ground truth" for the performance of a surgical gown is established by objective, standardized test methods and criteria defined by recognized national and international standards organizations (e.g., AAMI, ASTM, ISO, CPSC). These standards specify the methodology, equipment, and acceptable limits for performance. There is no human expert "ground truth" to be established in the context of device performance testing for surgical gowns.

    4. Adjudication Method for the Test Set

    Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based diagnostic device evaluations where expert consensus is needed to establish ground truth. As noted above, the performance of a surgical gown is evaluated against predefined objective criteria and standardized test results. Therefore, no human adjudication method was employed or is relevant for this type of device testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study is not applicable and was not done. This type of study assesses the impact of a device (often an AI algorithm) on human reader performance, typically in diagnostic imaging. The Aero Blue* Performance Surgical Gowns are barrier garments, not diagnostic tools, and do not involve human "readers" or AI assistance in their function.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone study of an algorithm is not applicable and was not done. The Aero Blue* Performance Surgical Gown is a physical medical device, not an algorithm or software. Its performance is evaluated directly through physical and material property tests.

    7. The Type of Ground Truth Used

    The ground truth used for the device's performance claims is based on objective, quantitative measurements and qualitative assessments derived from recognized industry standards and test methods. Specifically:

    • Standardized Test Data: The "Pass" results in the "Summary of Non-Clinical Testing" indicate that the device met the precise, objective criteria set forth by each referenced standard (e.g., AAMI PB70:2012 for liquid barrier, ASTM-1670-08 for synthetic blood penetration, ISO 10993 for biocompatibility).
    • Performance Metrics: The "ground truth" is defined by specific physical and material properties (e.g., water resistance, tensile strength, flammability, cytotoxicity) measured against predefined thresholds in the standards.

    8. Sample Size for the Training Set

    A "training set" is not applicable to this device. Training sets are used in the development of machine learning algorithms. The Aero Blue* Performance Surgical Gowns are physical products whose performance is determined by their material composition, design, and manufacturing processes, and subsequently validated through non-clinical laboratory testing.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no training set for this device. The physical and chemical properties of the materials, and thus the performance characteristics of the gown, are established through material science, engineering, and validated test methods, not through an iterative process of training an algorithm on a dataset with established ground truth.