510(k) Data Aggregation

    K Number
    K140437

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-13

    (112 days)

    Product Code
    Regulation Number
    878.3500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Su-Por® Surgical Implants in block, sheet, and anatomical shapes are intended for non-weight bearing applications of craniofacial reconstruction/cosmetic surgery and repair of craniofacial trauma. Su-Por® Surgical Implants are also intended for the augmentation or restoration of contour in the craniomaxillofacial skeleton.

    Device Description

    The Su-Por® Surgical Implants are marketed as single use sterile implants with various shapes and sizes for different areas of the craniofacial skeleton. The applications include non-load bearing augmentation and/or reconstruction of the craniofacial skeleton. The raw material used for the Su-Por® Surgical Implants is high-density polvethylene when molded into the implants becomes a porous high-density polyethylene. Polvethylene has a long history of use in surgical implantable products. The interconnecting open pore structure of the Su-Por® Surgical Implants allow for tissue in growth. The material used to manufacture the Su-Por® Surgical Implants has been utilized in reconstruction and soft tissue repair for many years. There is a long history of the use of porous polyethylene implants for enucleation and evisceration, as well as for many applications in craniofacial reconstruction and augmentation, with a history of safety and performance. The implants are single use and provided sterile by ethylene oxide (EO) terminal sterilization.

    AI/ML Overview

    Here's a summary of the device's acceptance criteria and the study that proves it meets them, based on the provided text:

    Device: Su-Por® Surgical Implants


    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Su-Por® Surgical Implants are established by demonstrating substantial equivalence to legally marketed predicate devices. This means that the Su-Por® implants must be as safe, as effective, and perform as well as or better than the predicate devices. The performance is assessed across several categories, including material properties, manufacturing processes, packaging, and biocompatibility.

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance (Su-Por® Surgical Implants)
    BiocompatibilityCytotoxicityPassed (demonstrates substantial equivalence to predicate devices)
    ISO Systemic ToxicityPassed (demonstrates substantial equivalence to predicate devices)
    ISO Intracutaneous StudyPassed (demonstrates substantial equivalence to predicate devices)
    USP Pyrogen StudyPassed (demonstrates substantial equivalence to predicate devices)
    ISO Muscle Implantation StudyPassed (demonstrates substantial equivalence to predicate devices)
    SterilitySterilization Validation (Ethylene Oxide - EtO)Completed sterilization validation, demonstrating they are sterile devices for implantation, equivalent to predicate devices.
    Mechanical PropertiesImpact TestingCompleted. (Implies performance similar to predicate devices, contributing to substantial equivalence.)
    Flexural TestingCompleted. (Implies performance similar to predicate devices, contributing to substantial equivalence.)
    Tensile Strength TestingCompleted. (Implies performance similar to predicate devices, contributing to substantial equivalence.)
    Material PurityPurity Testing per USPCompleted. (Implies performance similar to predicate devices, contributing to substantial equivalence.)
    Material StructurePorosity TestingCompleted. (Implies comparable porous structure to predicate devices, allowing for tissue ingrowth as intended.)
    Overall EquivalenceIntended Use & Indications for UseSame as predicate devices.
    Technological Characteristics & Principles of OperationSame as predicate devices.
    Raw Materials, Manufacturing Processes, PackagingSame as predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a "test set" in the context of a statistical study with a specific sample size. Instead, the evaluation focuses on bench testing and biocompatibility testing of the Su-Por® Surgical Implants themselves.

    • Sample Size for Testing: Not specified as a number of "cases" or "patients." The testing involved samples of the device material and manufactured implants for various physical, chemical, and biological assessments.
    • Data Provenance: The document does not specify the country of origin for the data or if it was retrospective or prospective. It describes laboratory and material testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of information (number of experts, qualifications, etc.) is typically associated with clinical studies or evaluations where human assessment of medical images or conditions defines the ground truth. This document describes the premarket notification (510(k)) process for a medical implant, which primarily relies on bench testing and established material science standards to demonstrate substantial equivalence to existing devices.

    Therefore, there is no mention of experts establishing ground truth in this context, as it's not a diagnostic AI device or a clinical outcome study requiring such a setup.


    4. Adjudication Method for the Test Set

    As there is no "test set" in the sense of clinical cases requiring expert adjudication, no adjudication method is mentioned or applicable.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the performance of AI algorithms when used by human readers (e.g., radiologists, pathologists) for diagnostic tasks. The Su-Por® Surgical Implant is a physical implant, not a diagnostic tool or an AI-assisted interpretation system.


    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone algorithm performance study was not done. This is not an AI device. The evaluation is for a physical medical implant.


    7. The Type of Ground Truth Used

    The "ground truth" in this context is based on established material science standards, biocompatibility regulations, and predefined physical/chemical properties that define the safety and performance of porous polyethylene implants. The "truth" is whether the material and finished device meet these engineering and biological specifications, and whether they are comparable to predicate devices with a history of safe and effective use.

    • Bench Test Results: Passing criteria for impact, flexural, tensile strength, purity, and porosity.
    • Biocompatibility Standards: Meeting ISO and USP standards for cytotoxicity, systemic toxicity, intracutaneous reaction, pyrogenicity, and muscle implantation.
    • Sterilization Validation: Demonstrated sterility via Ethylene Oxide (EtO).

    8. The Sample Size for the Training Set

    This concept is not applicable here. There is no AI algorithm being "trained." The Su-Por® Surgical Implant is a manufactured device.


    9. How the Ground Truth for the Training Set Was Established

    This concept is not applicable here, as there is no training set for an AI algorithm.