    Intended Use

    The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

    The GoodLife AC-300 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-300 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-300 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

    The GoodLife AC-301 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-301 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

    The GoodLife AC-302 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-302 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

    The GoodLife AC-303 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-303 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

    The GoodLife AC-304 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-304 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

    The GoodLife AC-305 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-305 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-305 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

    The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.

    The GoodLife AC-300 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-300 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

    The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.

    The GoodLife AC-301 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-301 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

    The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.

    The GoodLife AC-302 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-302 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

    The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.

    The GoodLife AC-303 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-303 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

    The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.

    The GoodLife AC-304 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-304 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

    The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-305 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-305 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

    Device Description

    GL GoodLife Series Blood Glucose Monitoring System consists of: (1) Glucose Meter (Model no.: AC-300, AC-301, AC-302, AC-303, AC-304, AC-305), (2) Glucose Test Strips (Model no.: AC) prior 510(k) clearance K113307, (3) Two levels of glucose control solutions (L1 and L2) may be purchased separately. prior 510(k) clearance K113307, (4) Check Strip, (5) Instruction for use. GL GoodLife Series Professional Blood Glucose Monitoring System consists of: Glucose Meter (Model no.: AC-300, AC-301, AC-302, AC-303, AC-304, AC-305), (2) Glucose Test Strips (Model no.: AC) prior 510(k) clearance K113307, (3) Two levels of glucose control solutions (L1 and L2) may be purchased separately. prior 510(k) clearance K113307, (4) Check Strip, (5) Instruction for use. These models have the same intended use and technological characteristics except the housing of the meter. Although the different shapes of housing, the materials are the same.

    AI/ML Overview

    The provided document is a 510(k) summary for the GoodLife AC-300 series Blood Glucose Monitoring Systems. It claims substantial equivalence to a predicate device (K113307) based on meeting performance requirements.

    Here's a breakdown of the requested information, extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria and reported device performance. It only states that the device "met pre-determined acceptance criteria" and "fulfilled the requirements from those standards mentioned above." The exact standards and specific performance metrics (e.g., accuracy percentages per glucose range) are not detailed within this summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "results of testing performed to validate changes to the system met pre-determined acceptance criteria," implying a test set was used, but it does not provide any information regarding the sample size of the test set or the data provenance (country of origin, retrospective/prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not provide any information about the number of experts used to establish ground truth or their qualifications. For blood glucose monitoring systems, ground truth is typically established by laboratory reference methods, not subjective expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not provide any information about an adjudication method for the test set. For objective measurements like blood glucose, adjudication would not typically be applicable in the way it is for image interpretation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is a blood glucose monitoring system, not an AI-assisted diagnostic imaging device that involves human readers. Therefore, an MRMC comparative effectiveness study involving human readers is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, this is a standalone device. The "algorithm" (the electrochemical biosensor system) provides the measurement directly without human subjective interpretation. The document describes it as an electrochemical biosensor system that measures and displays the result directly.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    While not explicitly stated in this summary, for blood glucose monitoring systems, the ground truth is established by comparison to a recognized laboratory reference method (e.g., a YSI analyzer). The regulatory review for such devices always requires this comparison, and the phrase "met pre-determined acceptance criteria" would implicitly refer to such a comparison.

    8. The sample size for the training set

    This device is not an AI/machine learning model in the conventional sense that would require a "training set." It is an electrochemical biosensor. Therefore, the concept of a training set is not applicable. The device's performance is determined by its engineering, chemical reagents, and calibration processes.

    9. How the ground truth for the training set was established

    As the concept of a training set is not applicable for this type of device, this information is not relevant. The device's calibration is achieved by an implicit coding process when the test strip is inserted, which incorporates parameters to calculate the glucose value (page 17).