Goodlife AC-300 Blood Glucose Monitoring System, Goodlife AC-301 Blood Glucose Monitoring System, Goodlife AC-302 Blood Glucose Monitoring System, Goodlife AC 303 Blood Glucose Monitoring System, Goodlife AC-304 Blood Glucose Monitoring System Goodlife AC-305 Blood Glucose Monitoring System, Goodlife AC-300 Professional Blood Glucose Monitoring System, Goodlife AC-301 Professional Blood Glucose Monitoring System Goodlife AC-30
K140150 · Hmd Biomedical, Inc. · NBW · May 14, 2015 · Clinical Chemistry
Device Facts
| Record ID | K140150 |
| Device Name | Goodlife AC-300 Blood Glucose Monitoring System, Goodlife AC-301 Blood Glucose Monitoring System, Goodlife AC-302 Blood Glucose Monitoring System, Goodlife AC 303 Blood Glucose Monitoring System, Goodlife AC-304 Blood Glucose Monitoring System Goodlife AC-305 Blood Glucose Monitoring System, Goodlife AC-300 Professional Blood Glucose Monitoring System, Goodlife AC-301 Professional Blood Glucose Monitoring System Goodlife AC-30 |
| Applicant | Hmd Biomedical, Inc. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | May 14, 2015 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients. The GoodLife AC-300 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-300 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-300 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
Device Story
The GoodLife AC-300/301/302/303/304/305 series are electrochemical blood glucose monitoring systems. The device accepts fresh capillary whole blood samples via test strips. Glucose in the blood reacts with Glucose Dehydrogenase (FAD) on the strip, generating an electric current proportional to the glucose concentration. The meter measures this current, calculates the glucose level, and displays it on an LCD. The system features auto-coding upon strip insertion. Devices are available in single-patient (home) and professional (clinical) versions. Professional versions require use with single-use, auto-disabling lancing devices. The system includes an NFC module for data transmission. Healthcare providers use the displayed results to monitor diabetes control effectiveness. Patients benefit from convenient, home-based glucose monitoring.
Clinical Evidence
Bench testing only. The manufacturer performed validation testing to confirm that the modifications (NFC integration and packaging changes) met pre-determined acceptance criteria and maintained performance equivalence to the predicate device.
Technological Characteristics
Electrochemical biosensor (amperometry); Glucose Dehydrogenase (FAD) enzyme; 0.5uL sample volume; 5-second test time; 20-600 mg/dL range; auto-coding; NFC connectivity; 999-test memory; 30-day averaging; powered by battery; LCD display.
Indications for Use
Indicated for quantitative measurement of glucose in fresh capillary whole blood from fingertips. Intended for single-patient home use or multi-patient professional clinical use as an aid to monitor diabetes control. Contraindicated for diabetes diagnosis, screening, or neonatal use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- GoodLife AC-300 Blood Glucose Monitoring System (K113307)
Related Devices
- K151265 — SD GlucoNFC Blood Glucose Monitoring System, SD GlucoNFC Multi Blood Glucose Monitoring System · Sd Biosensor · Nov 24, 2015
- K132929 — SD GLUCONAVII MENTOR NFC MULTI BLOOD GLUCOSE MONITORING SYSTEM, AND SD GLUCONAVII MENTOR NFC BLOOD GLUCOSE MONITORING SYSTEM · Sd Biosensor, Inc. · Aug 5, 2015
- K123008 — RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM GM700 GE GE200 BLOOD GLUCOSE MONITORING SYSTEM RIGHTEST BLOOD GLUCOSE MONITORIN · Bionime Corporation · Jan 25, 2013
- K231192 — RIGHTEST Blood Glucose Monitoring System Max Tel · Bionime Corporation · Jan 19, 2024
- K092052 — RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, MODEL GM550 · Bionime Corporation · Jun 18, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form, with three figures facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 14, 2015
HMD BIOMEDICAL, INC. AXEL LIN QUALITY ASSURANCE MANAGER NO 181 MINSHENG ST, XINPU TOWNSHIP HSINCHU 305, TAIWAN
Re: K140150
Trade/Device Name: Goodlife AC-300 Blood Glucose Monitoring System Goodlife AC-301 Blood Glucose Monitoring System Goodlife AC-302 Blood Glucose Monitoring System Goodlife AC 303 Blood Glucose Monitoring System Goodlife AC-304 Blood Glucose Monitoring System Goodlife AC-305 Blood Glucose Monitoring System Goodlife AC-300 Professional Blood Glucose Monitoring System Goodlife AC-301 Professional Blood Glucose Monitoring System Goodlife AC-302 Professional Blood Glucose Monitoring System Goodlife AC-303 Professional Blood Glucose Monitoring System Goodlife AC-304 Professional Blood Glucose Monitoring System Goodlife AC-305 Professional Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR Dated: April 15, 2015 Received: April 16, 2015
Dear Axel Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Katherine Serrano -S
For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-300 Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-300 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-300 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-300 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-301 Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-301 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-301 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-302 Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-302 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-302 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-303 Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-303 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-303 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-304 Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-304 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-304 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
Device Name
GoodLife AC-305 Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-305 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-305 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are with the GoodLife AC-305 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-300 Professional Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.
The GoodLife AC-300 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-300 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Type of Use (Select one or both, as applicable)
본 Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{9}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-301 Professional Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.
The GoodLife AC-301 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-301 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Type of Use (Select one or both, as applicable)
본 Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{10}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-302 Professional Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.
The GoodLife AC-302 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-302 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Type of Use (Select one or both, as applicable)
본 Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{11}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-303 Professional Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.
The GoodLife AC-303 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-303 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Type of Use (Select one or both, as applicable)
본 Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{12}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-304 Professional Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.
The GoodLife AC-304 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-304 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Type of Use (Select one or both, as applicable)
본 Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{13}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140150
#### Device Name
GoodLife AC-305 Professional Blood Glucose Monitoring System
### Indications for Use (Describe)
The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of dabetes control.
The GoodLife AC-305 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-305 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Type of Use (Select one or both, as applicable)
본 Prescription Use (Part 21 CFR 801 Subpart D)
본 Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{14}------------------------------------------------
# 510(k) Summary
## 1. Submitter Information
| Company name: | HMD BioMedical Inc. |
|-----------------|------------------------------------------------------------------|
| Contact person: | Axel Lin |
| Address: | No. 181, Minsheng St., Xinpu Township, Hsinchu<br>County, Taiwan |
| Phone: | 886-3-6017000 |
| FAX: | 886-3-5885500 |
| E-mail: | axellin@hmdbio.com |
| Date Prepared: | MAY 14 2015 |
## 2. Name of Device
Trade/Proprietary Name:
For single patient use-
GoodLife AC-300 Blood Glucose Monitoring System GoodLife AC-301 Blood Glucose Monitoring System GoodLife AC-302 Blood Glucose Monitoring System GoodLife AC-303 Blood Glucose Monitoring System GoodLife AC-304 Blood Glucose Monitoring System GoodLife AC-305 Blood Glucose Monitoring System
For multiple patient use-
GoodLife AC-300 Professional Blood Glucose Monitoring System GoodLife AC-301 Professional Blood Glucose Monitoring System GoodLife AC-302 Professional Blood Glucose Monitoring System GoodLife AC-303 Professional Blood Glucose Monitoring System GoodLife AC-304 Professional Blood Glucose Monitoring System GoodLife AC-305 Professional Blood Glucose Monitoring System
{15}------------------------------------------------
| Common name: | Blood Glucose Test System |
|-----------------------|-----------------------------------------|
| Classification name: | Glucose Test System |
| Classification Panel: | Clinical Chemistry (75) |
| Regulation no.: | 862.1345 (Class II), 862.1660 (Class I) |
| Product code: | NBW, LFR |
## 3. Predicate Device
## Trade/Proprietary name:
GoodLife AC-300 Blood Glucose Monitoring System GoodLife AC-301 Blood Glucose Monitoring System GoodLife AC-302 Blood Glucose Monitoring System GoodLife AC-303 Blood Glucose Monitoring System GoodLife AC-304 Blood Glucose Monitoring System GoodLife AC-305 Blood Glucose Monitoring System GoodLife AC-300 Professional Blood Glucose Monitoring System GoodLife AC-301 Professional Blood Glucose Monitoring System GoodLife AC-302 Professional Blood Glucose Monitoring System GoodLife AC-303 Professional Blood Glucose Monitoring System GoodLife AC-304 Professional Blood Glucose Monitoring System GoodLife AC-305 Professional Blood Glucose Monitoring System Rlood Cluce Test Svste
| Common name: | Blood Glucose Test System |
|---------------|---------------------------|
| Submitter: | HMD BioMedical Inc. |
| 510(k) no.: | K113307 |
| Product code: | LFR, NBW, JJX |
- 4. Device Description
The device description of the GL GoodLife Series Blood Glucose Monitoring System is as follows.
GL GoodLife Series Blood Glucose Monitoring System consists of:
{16}------------------------------------------------
(1) Glucose Meter (Model no.: AC-300, AC-301, AC-302, AC-303, AC-304, AC-305)
(2) Glucose Test Strips (Model no.: AC) prior 510(k) clearance K113307
(3) Two levels of glucose control solutions (L1 and L2) may be purchased separately. prior 510(k) clearance K113307
(4) Check Strip
(5) Instruction for use
GL GoodLife Series Professional Blood Glucose Monitoring System consists of:
Glucose Meter (Model no.: AC-300, AC-301, AC-302, AC-303, (1) AC-304, AC-305)
(2) Glucose Test Strips (Model no.: AC) prior 510(k) clearance K113307
(3) Two levels of glucose control solutions (L1 and L2) may be purchased separately. prior 510(k) clearance K113307
- (4) Check Strip
- (5) Instruction for use
These models have the same intended use and technological characteristics except the housing of the meter. Although the different shapes of housing, the materials are the same.
## [Test Principle]
GL GoodLife Series Blood Glucose Monitoring System is an electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor. When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.
{17}------------------------------------------------
# [Control Solution]
The GL GoodLife Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) assessing the performance of the GL GoodLife Series Blood Glucose monitoring system and GoodLife Blood Glucose Test strips (AC). There are two levels of controls (L1 and L2).
# [Check Strip]
The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.
# [Device Calibration]
The device is calibrated by implicit coding process. While inserting the test strip into strip slot to perform the blood glucose test, the coding procedure is complete. The meter will apply formula including this parameter of code to calculate the glucose value.
# 5. Intended Use
For single patient use:
GoodLife AC-300 Blood Glucose Monitoring System
The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-300 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-300 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-300 Blood Glucose Meter to quantitatively measure
{18}------------------------------------------------
glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
## GoodLife AC-301 Blood Glucose Monitoring System
The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-301 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-301 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
## GoodLife AC-302 Blood Glucose Monitoring System
The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-302 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
{19}------------------------------------------------
The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-302 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
GoodLife AC-303 Blood Glucose Monitoring System
The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-303 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-303 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
GoodLife AC-304 Blood Glucose Monitoring System
The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-304 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Blood Glucose Monitoring
{20}------------------------------------------------
System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-304 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
## GoodLife AC-305 Blood Glucose Monitoring System
The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
The GoodLife AC-305 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-305 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-305 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
For multiple use-
GoodLife AC-300 Professional Blood Glucose Monitoring System
The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-300 Professional Blood
{21}------------------------------------------------
Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-300 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-300 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
GoodLife AC-301 Professional Blood Glucose Monitoring System
The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-301 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
{22}------------------------------------------------
GoodLife AC-302 Professional Blood Glucose Monitoring System
The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-302 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
## GoodLife AC-303 Professional Blood Glucose Monitoring System
The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
{23}------------------------------------------------
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-303 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
GoodLife AC-304 Professional Blood Glucose Monitoring System
The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-304 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
GoodLife AC-305 Professional Blood Glucose Monitoring System
The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid
{24}------------------------------------------------
to monitor the effectiveness of diabetes control. The GoodLife AC-305 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-305 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
{25}------------------------------------------------
- 6. Comparison of Subject Devices and predicate device
Technological Characteristics Comparison Table of GL GoodLife Series Blood Glucose Monitoring System (K113307)
| Similarities | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Item | Subject device | Predicate device |
| | GL GoodLife Series BGMS | GL GoodLife Series<br>BGMS (K113307) |
| Intended use | It is designed to quantitatively measure the<br>concentration of glucose in fresh capillary whole<br>blood. | Same |
| Detection<br>method | Amperometry: Current produced by chemical<br>reaction | Same |
| Test range | 20~600mg/dL | Same |
| Operating<br>conditions | 50~104°F (10~40°C), 20~80% | Same |
| Storage<br>Temperature | 50~104°F (10~40°C), 20~80% | Same |
| Autocoding | Yes | Same |
| Enzyme | Glucose Dehydrogenase (FAD) (Aspergillus<br>oryzae) | Same |
| Capillary<br>testing site | Fingertips only | Same |
| Sample<br>volume | 0.5ul | Same |
| Memory | 999 | Same |
| Test time | 5 sec | Same |
| Average | 30 days | Same |
| Coding<br>function | Coding function is combined in each test strip.<br>Coding will be done when insert the test strip to | Same |
| Similarities | | |
| Item | <b>Subject device</b> | <b>Predicate device</b> |
| | GL GoodLife Series BGMS | GL GoodLife Series<br>BGMS (K113307) |
| | perform the glucose test. | |
{26}------------------------------------------------
| Differences | | |
|-----------------------------------------------------------|------------------------------------------------------------------|----------------------------------------|
| Item | Subject device | Predicate device |
| | GL GoodLife Series | GL GoodLife Series (K113307) |
| Data<br>transmission | Near Field Communication (NFC)<br>module (Dynamic NFC Tag) | N/A |
| S/N Label of<br>Meter | With NFC logo | Without NFC logo |
| Bag size and<br>the label size<br>of foil packed<br>strip | Adding different packing formats for<br>15 pcs, 100 pcs/each bag | Only one format for 50 pcs/each<br>bag |
| Contents of foil<br>packed strip | 15 pcs, 50 pcs, 100 pcs/each bag | Only 50 pcs/each bag |
# 7. Discussion of Clinical Tests Performed
GL GoodLife Series Blood Glucose Monitoring System (Subject Device) results of testing performed to validate changes to the system met pre-determined acceptance criteria and demonstrate that the subject device is substantially equivalent to the predicate device.
# 8. Conclusion
The subject device was tested and fulfilled the requirements from those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.