Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132841

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-01-23

    (135 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein.

    Device Description

    Zyno Medical Administration Set is a device used to administer fluids from a container to a patient's vascular system through a needle or a catheter inserted into a vein. The major modifications are: 1) extended selection of interchangeable components manufactured by additional suppliers; 2) additional contract manufacturer for the device: 3) Gamma Irradiation as additional sterilization method.

    AI/ML Overview

    The Zyno Medical Administration Set is an intravenous administration set used to deliver fluids from a container to a patient's vascular system.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Category)Reported Device Performance
    Safety & Effectiveness
    BiocompatibilitySuccessfully met acceptance criteria.
    PyrogenicitySuccessfully met acceptance criteria.
    Chemical TestingSuccessfully met acceptance criteria.
    Microbial IngressSuccessfully met acceptance criteria.
    DEHP TestingSuccessfully met acceptance criteria.
    Sterilization TestingSuccessfully met acceptance criteria.
    Functional Performance
    Bench Testing (physical & mechanical)Successfully met acceptance criteria.
    Integration Testing (with Z-800 series infusion pumps)Successfully met acceptance criteria.
    Aging TestingSuccessfully met acceptance criteria.
    Material & Design Standards
    Tubing MaterialApproved Standard PVC (Same as predicate)
    Single UseYes (Same as predicate)
    SterileYes (Same as predicate)
    ISO 8536-4:2010 CompliantYes (Same as predicate)
    ISO 10993 CompliantYes (Same as predicate)
    Operation ModeFor use with Z-800 series of infusion pumps or gravity feed (Same as predicate)

    Conclusion of Nonclinical Testing: "All nonclinical testing results successfully demonstrated that the Zyno Medical Administration Sets performed as intended. Conclusions drawn from the nonclinical testing is that the testing data demonstrated that the device is as safe, as effective, and performs as well as the legally marketed device cleared in K120685."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for each of the nonclinical tests (Biocompatibility, Pyrogenicity, Chemical, Microbial Ingress, DEHP, Sterilization, Bench, Integration, and Aging Testing). It only states that "extensive verification and validation testing" was performed and that "All results successfully met acceptance criteria."

    The data provenance is not specified regarding country of origin, nor is it explicitly stated whether the data was retrospective or prospective. Given these are bench and lab-based tests for a medical device administration set, they are inherently prospective controlled tests conducted by the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. The nonclinical tests appear to rely on established testing methods and standards (e.g., ISO standards, FDA guidance documents) rather than subjective expert consensus for establishing ground truth. Compliance with these standards is the metric, not expert opinion.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of data by multiple experts. Since no human clinical studies were performed, and the testing was nonclinical (bench, lab-based), an adjudication method is not applicable and therefore not mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Human clinical studies were deemed unnecessary to evaluate the safety or effectiveness of the Zyno Medical Administration Sets." The submission focuses on demonstrating substantial equivalence through nonclinical testing.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable to the Zyno Medical Administration Sets. This device is a physical medical administration set, not an AI algorithm or software device. Therefore, the concept of "standalone performance" in the context of an algorithm does not apply. The device's performance stands alone in its physical and functional testing.

    7. The Type of Ground Truth Used

    The ground truth for the nonclinical tests was based on established industry standards and regulatory guidance. For example:

    • Biocompatibility: Likely ISO 10993 series.
    • Sterilization: Compliance with relevant sterility assurance levels.
    • Physical/Mechanical Testing: Adherence to performance specifications to ensure proper function and safety.
    • Integration Testing: Successful operation with the Z-800 series infusion pumps, demonstrating compatibility and correct fluid delivery.
    • Pyrogenicity, Chemical, Microbial Ingress, DEHP, Aging: Compliance with specific test methods and limits defined in relevant standards and guidance documents (e.g., "Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)]" and "Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions").

    8. The Sample Size for the Training Set

    This question is not applicable. The Zyno Medical Administration Set is a physical medical device, not a machine learning or AI model that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1