510(k) Data Aggregation

    K Number
    K132652

    Validate with FDA (Live)

    Date Cleared
    2013-12-20

    (116 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VIA™ Microcatheter is intended for the introduction of non-liquid interventional devices (such as stents/flow diverters) and infusion of diagnostic (such as contrast media), or therapeutic agents into the neuro, peripheral and coronary vasculature.

    Device Description

    The VIA™ Microcatheter and VIA™ PLUS Microcatheter are designed to be introduced over a stcerable guidewire into the vasculature. The physician inserts the catheter into the vein or artery through the skin (percutaneous) using a sheath or guidewire. The device can then be navigated to the treatment site. Navigation is aided by the coated surface of the catheter which assists with manipulation while in the vasculature. Throughout the procedure the physician can obtain the position of the catheter by the tip marker using fluoroscopic techniques. Interventional devices and infusion of diagnostic and therapeutic agents can be delivered through the lumen of the catheter to the treatment site.

    The VIA™ Microcatheter is a sterile single lumen device with a single distal tip marker designed to aid the physician in accessing distal vasculature when used with a guide catheter and steerable guidewire. Variable shaft stiffness ranging from a flexible tip to a semi-rigid proximal section aid the physician in tracking over selectively placed guidewires. The proximal end of the catheter incorporates a standard luer adapter to facilitate attachment of accessories. A single radiopaque marker positioned at the distal tip facilitates fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic coating which reduces friction during manipulation in the vessel. The inner lumen of the catheter has a PTFE liner which assists with delivery of interventional devices, such as an intraluminal flow diverter.

    The VIA™ and VIA™ PLUS Microcatheter are available in effective lengths of 154 cm and 133 cm and inner diameters of 0.27 inches and 0.33 inches, respectively. For commercialization purposes the 0.027 inch diameter will be named the VIA™ Microcatheter and the 0.033 inch diameter will be named the VIA™ PLUS Microcatheter.

    The VIA™ Microcatheter is presented in a tyvek pouch and is sterile, single-use only and nonpvrogenic.

    Accessories: Each VIA™ Microcatheter is provided with a shaping mandrel to facilitate distal tip shaping.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the VIA™ Microcatheter based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestMethodologyApplicable International Standard and/or Sequent Test Method (TM)Acceptance CriteriaReported Device PerformanceConclusion
    Visual and DimensionalTests conformance to specified microcatheter dimensions and visual criteria.ISO 10555-1:2009Conformance to specified dimensions and visual criteriaMet performance specifications per standard and test methodComplete conformance to standard
    Kink ResistanceMeasures diameter at which microcatheter shaft sections and junctions will kink.BS EN 13868:2002Not explicitly stated (implied by "Met performance specifications")Met performance specifications per standard and test methodConformance to standard with the following deviations:Kink diameter determined based on mechanical kink (drop in compressive force) instead of 50% reduction in water flow. This method is appropriate as the Via is primarily used to deliver non-liquid devices.
    Tip BucklingTests force required for tip to buckle.Internal test methodNot explicitly stated (implied by "Met performance specifications")Met performance specifications per test methodComplete conformance to Sequent TM
    Tracking ForceTests force required to advance an interventional device through the microcatheter lumen.Internal test methodNot explicitly stated (implied by "Met performance specifications")Met performance specifications per test methodComplete conformance to Sequent TM
    Steam Shaping and Shape RetentionTests that microcatheter can be steam shaped to a clinically relevant angle and can maintain a minimum % of the initial angle after simulated use.Internal test methodClinically relevant angle maintained, minimum % of initial angle retainedMet performance specifications per test methodComplete conformance to Sequent TM
    Shaft TensileMeasures the ultimate tensile strength of all Pebax and Vestamid junctions along the length of the catheter shaft.ISO 10555-1:2009Not explicitly stated (implied by "Met performance specifications")Met performance specifications per standard and test methodComplete conformance to standard
    Hub-Shaft TensileMeasures the ultimate tensile strength of the hub to shaft junctionISO 10555-1:2009Not explicitly stated (implied by "Met performance specifications")Met performance specifications per standard and test methodComplete conformance to standard
    BurstMeasures peak pressure before microcatheter burst/liquid leakage.ISO 10555-1:2009Not explicitly stated (implied by "Met performance specifications")Met performance specifications per standard and test methodComplete conformance to standard
    Coating Friction and Coating IntegrityMeasures the average peak coating friction/lubricity. Coating integrity uses dye to test that coating remains adhered to catheter after simulated use.Harland Medical Systems Coating Friction and Dye Test Methods FDA Guidance Document, Class II Special Controls for PTCA Catheters (section 12) (Issued 2010)Not explicitly stated (implied by "Met performance specifications")Met performance specifications per standard and test methodComplete conformance to supplier test methods. In line with FDA guidance document on coating integrity.
    Coating Adherence/ParticulateMeasures particulate generated from the hydrophilic coating on exterior of microcatheter, as well as particulate generated from advancing an interventional device through the inner lumen of the microcatheter.Internal test method FDA Guidance Document, Non-Clinical Engineering Tests for Intravascular Stents and Associated Delivery Systems (section 12) (Issued 2010) FDA Guidance Document, Class II Special Controls for PTCA Catheters (section 13) (Issued 2010)Not explicitly stated (implied by "Met performance specifications")Met performance specifications per standard and test methodComplete conformance to Sequent TM. In line with FDA guidance documents on particulate testing.
    Hub PerformanceTests hub liquid and air leakage, as well as that the hub can withstand adequate forces. Tests that the hub meets general requirements for conical fittings.ISO 594-1:1986 ISO 594-2:1998 Internal test methodNot explicitly stated (implied by "Met performance specifications")Met performance specifications per standard and test methodConformance to standard with the following deviations:Used an alternative ISO 594-2:1998 fitting to test hubs for Separation Force and Unscrewing Torque. The fitting used was measured to have a minor diameter greater than called for in the standard. This was deemed as worst case for these tests, therefore acceptable to use.Only short term stress cracking was inspected for on hubs. ISO 594-2:1998 calls for inspection after 48 hours. Intended use of Via microcatheter only requires RHV or syringes to be connected for short durations during delivery of implant, therefore long term testing is not applicable.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text only mentions "bench tests performed" and "simulated use testing in animals." It does not specify the sample size for the test set (number of catheters tested for each bench test) or the data provenance (e.g., country of origin, retrospective or prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device is a microcatheter, and the evaluation relies on physical and functional bench testing, and simulated animal use, not on human interpretation of images or data requiring expert consensus or ground truth in the typical clinical AI context.

    4. Adjudication Method for the Test Set

    Not applicable, for the same reasons as #3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size

    No MRMC study was done. This is a medical device (microcatheter), not an AI algorithm for diagnostic interpretation. The evaluation focuses on the device's physical and functional performance, not on improving human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable to a physical medical device like a microcatheter. The "standalone" performance here relates to the microcatheter's ability to perform its intended functions (e.g., track, burst pressure, kink resistance) as per its design specifications. The performance data presented in the table represents the standalone performance of the device against predefined technical specifications and standards.

    7. The Type of Ground Truth Used

    The ground truth for the performance testing is based on:

    • International Standards: ISO 10555-1:2009, BS EN 13868:2002, ISO 594-1:1986, ISO 594-2:1998.
    • FDA Guidance Documents: Class II Special Controls for PTCA Catheters (sections 12 & 13) and Non-Clinical Engineering Tests for Intravascular Stents and Associated Delivery Systems (section 12).
    • Supplier Test Methods: Harland Medical Systems Coating Friction and Dye Test Methods.
    • Sequent Medical Internal Test Methods (Sequent TM): For tests like Tip Buckling, Tracking Force, Steam Shaping and Shape Retention, Coating Adherence/Particulate, and Hub Performance.

    Essentially, the "ground truth" is defined by the technical specifications, performance limits, and methodologies outlined in these various standards and internal protocols.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set. The term "training set" is relevant for AI models, not for traditional medical devices undergoing bench and simulated use testing.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.