510(k) Data Aggregation

    K Number
    K132258

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-11-19

    (123 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zimmer Trabecular Metal Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional or delayed healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.

    The 3.7mmD Zimmer Trabecular Metal Implants should be splinted to additional implants when used in the pre-molar region and should not be used in the molar region.
    The 4.1mmD Zimmer Trabecular Metal Implants should be splinted to additional implants when used in the molar region.
    The 4.1mmD x 8mmL Zimmer Trabecular Metal Implant should be splinted to additional implants when used in the pre-molar region and should not be used in the molar region.

    Device Description

    Trabecular Metal Implant is an endosseous dental implant composed of titanium alloy and Trabecular Metal (tantalum). The implant section is designed for ease of implantation and with greater surface area for osseointegration. The implant section surface is treated to facilitate osseointegration. In addition, the implant section is tapered with triple-lead threads.

    The Trabecular Metal Implant family is currently offered in 4.1. 4.7, and 6.0mm diameters in lengths of 8. 10. 11.5. 13. and 16mm. They include two different texturing configurations: full texture to the top of the implant and texture to 0.5mm from the top of the implant. In addition, both texturing configurations of the implant have coronal grooves on the collar to within 0.64mm of the top of the implant similar to the predicate #1 : Trabecular Metal Dental Implant. An additional Trabecular Metal implant with a new diameter of 3.7mmD will be offered in lengths of 10, 11.5, 13 and 16mmL. The implant-abutment interface platform diameter will be offered in a size of 3.5mm. The new device will feature MTX surface equivalent to existing Zimmer Dental implants. The MTX surface is used on the titanium body and is exposed on surfaces apical and coronal to the Trabecular Metal.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device, the Zimmer Trabecular Metal Implant. For medical devices, "acceptance criteria" and "device performance" are typically related to the physical and mechanical properties of the device, and the "study" proving these are usually non-clinical (laboratory) tests. Clinical studies (with human subjects) are often not required for 510(k) clearances, especially if substantial equivalence to a predicate device can be demonstrated through non-clinical testing.

    Here's the breakdown of the requested information based on the provided text, focusing on the absence of typical "AI/software algorithm" acceptance criteria and studies, as this is a dental implant:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in a quantitative table format as might be seen for software performance metrics (e.g., sensitivity, specificity thresholds). Instead, it refers to performance standards for mechanical properties.

    Acceptance Criteria CategoryReported Device Performance
    Fatigue TestingMeets standards in accordance with FDA guidance Class II Special Controls Guidance Document: Root-form Dental Implants and Endosseous Dental Implant Abutments.
    Compression TestingMeets standards in accordance with FDA guidance Class II Special Controls Guidance Document: Root-form Dental Implants and Endosseous Dental Implant Abutments.
    Mechanical StrengthDemonstrated improvements over the predicate device.
    Torque Testing (Apical Tip Strength)Indicator that the strength at the apical tip of the implant is greater than the stress the implant will see in dense cortical bone, in accordance with internal Zimmer Research Protocols.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. Non-clinical mechanical testing typically uses a certain number of samples, but the exact count is not provided in this summary.
    • Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting. There is no mention of country of origin for data as it's not patient data, nor is there a retrospective or prospective nature to the data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable for non-clinical mechanical testing of a dental implant. "Ground truth" in this context refers to established engineering standards and guidelines for mechanical performance, not expert medical consensus on diagnostic interpretations.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving discrepancies in expert interpretations, typically in clinical imaging studies. This document describes mechanical testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No. An MRMC study is a clinical study comparing human reader performance, often with and without AI assistance. This document explicitly states: "No clinical testing was performed. Non-clinical testing was used to support the decision of substantial equivalence."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    No. This device is a physical dental implant, not a software algorithm.

    7. The Type of Ground Truth Used

    For this physical device, the "ground truth" for non-clinical testing is based on established engineering principles, industry standards (e.g., ISO standards for dental implants), and FDA guidance documents (specifically, the Class II Special Controls Guidance Document: Root-form Dental Implants and Endosseous Dental Implant Abutments). The device's ability to resist fatigue, compression, and maintain torque strength constitutes the "truth" against these predefined benchmarks.

    8. The Sample Size for the Training Set

    No. This is not applicable to the development and testing of a physical medical device like a dental implant. "Training set" refers to data used to train machine learning models.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As noted above, this relates to machine learning models, not physical device engineering.