Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K131325

    Validate with FDA (Live)

    Date Cleared
    2013-08-16

    (100 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cayenne Medical, Inc. Quattro® GL Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:

    Hip

    • Hip capsule repair
    • Acetabular labrum reattachment

    Shoulder

    • Capsular stabilization
    • Bankart repair
    • Anterior shoulder instability
    • SLAP lesion repairs
    • Capsular shift or capsulolabral reconstructions
    • Acromioclavicular separation repairs
    • Deltoid repairs
    • Rotator cuff repairs
    • Biceps tenodesis

    Elbow, Wrist, and Hand

    • Biceps tendon reattachment
    • Ulnar or radial collateral ligament reconstruction
    • Lateral epicondylitis repair

    Knee

    • Extra-capsular repairs
    • Medial collateral ligament
    • Lateral collateral ligament
    • Posterior oblique ligament
    • Patellar realignment and tendon repairs
    • Vastus medials obliquous advancement
    • Illiotibial band tenodesis

    Foot and Ankle

    • Hallux valgus repairs
    • Medial or Lateral instability repairs/reconstructions
    • Achilles tendon repairs/reconstructions
    • Midfoot reconstructions
    • Metatarsal ligament/tendon repairs/reconstructions
    • Bunionectomy
    Device Description

    This traditional 501(k) premarket notification is to expand the indications for use for Cayenne Medical Quattro GL Suture Anchor. The Quattro GL (LabraLink) Suture Anchor was cleared per premarket notification K112960. Cayenne seeks to expand the existing indications for use for the subject device to include the indications listed above.

    The Quattro® GL Suture Anchor is a sterile, manually operated, single procedure suture anchor. The anchor has two suture eyelets allowing for one or two sutures to be loaded through the eyelets. The suture anchor is mounted on an inserter. The Quattro GL Suture Anchor incorporates design features that facilitate suture anchor placement under arthroscopic, open, or limited access conditions in soft tissue to bone reattachment procedures. The Quattro GL Suture Anchor is only offered in one size, 2.9mm with four suture color options. The anchors are offered in two configurations, single loaded or double loaded sutures. Suture(s) used on the anchor are size # 2 non-absorbable surgical sutures. The Quattro GL inserter has a working length of 25.8 cm with an outer shaft diameter of 3.2 mm. Since the market clearance of this device, the length of the Suture Anchor was decreased from 15mm to 11.4mm. The technological characteristics of the Quattro GL Suture Anchor have not changed.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Quattro® GL Suture Anchor:

    1. A table of acceptance criteria and the reported device performance

    The provided 510(k) summary does not explicitly state specific acceptance criteria in a quantitative or qualitative manner that would be typically found for performance metrics like sensitivity, specificity, accuracy, etc., for a diagnostic device.

    Instead, the submission focuses on demonstrating substantial equivalence to existing predicate devices based on:

    Acceptance Criteria CategoryReported Device Performance (Summary)
    Mechanical Testing (Pull-Out Strength)"Testing showed that the Quattro GL Suture Anchor ultimate pull-out strength was comparable to that of the predicate device."
    Intended UseThe expanded indications for use for the Quattro GL Suture Anchor align with its intended purpose of reattaching soft tissue to bone across various anatomical locations (Hip, Shoulder, Elbow/Wrist/Hand, Knee, Foot/Ankle). This comparison demonstrates equivalence to its own previously cleared version and other predicate devices with similar indications.
    Design Features"The subject device... has the same intended use, design, technology, and materials... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
    Materials"The subject device... has the same... materials... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
    Manufacturing Processes"The subject device... has the same... manufacturing processes... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
    Sterilization Method"The subject device... has the same... sterilization method... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."
    Shelf Life"The subject device... has the same... shelf life... as the Cayenne Medical Quattro GL (LabraLink) Suture Anchor with Inserter (K112960)."

    Key takeaway: The primary "acceptance criterion" for this 510(k) appears to be demonstrating that the device is substantially equivalent to legally marketed predicate devices, particularly in terms of its mechanical properties and that the expanded indications for use do not introduce new safety or effectiveness concerns. The specific numerical values for "pull-out strength" that define "comparable" are not provided in this summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Non-clinical testing data submitted, referenced, or relied upon" including "Mechanical testing (pull-out strength)."

    • Sample Size: The specific sample size used for the pull-out strength testing is not specified in the provided summary.
    • Data Provenance: The document does not provide details on the country of origin of the data or whether the testing was retrospective or prospective. Given it's mechanical testing of physical anchors, it's inherently prospective in nature for a new batch of tested devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable to the provided document. The Quattro® GL Suture Anchor is a medical device (suture anchor) for surgical reattachment, not a diagnostic device that requires expert interpretation for establishing ground truth on a test set (e.g., for image analysis). The "ground truth" for this device lies in its physical and mechanical properties (e.g., strength, biocompatibility), which are evaluated through non-clinical laboratory testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable for the same reasons as point 3. Adjudication methods like 2+1 (two readers agree, third is tie-breaker) are used in diagnostic studies to resolve discrepancies in expert interpretation of medical images or data. The testing described for the suture anchor is mechanical in nature and does not involve human interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The Quattro® GL Suture Anchor is a surgical implant, not an AI-powered diagnostic tool or system designed to assist human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable. The device is a physical surgical implant. There is no algorithm or AI component to this device that would have standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the mechanical testing (pull-out strength), the "ground truth" would be the objective, empirically measured physical properties of the suture anchors under controlled laboratory conditions, typically using standardized testing methods (e.g., ASTM standards) on a material testing machine. This is not "expert consensus" or "pathology" in the diagnostic sense, but rather a direct measurement of physical performance.

    8. The sample size for the training set

    This section is not applicable. This device is a physical medical implant, not an AI or machine learning model that requires a "training set" of data.

    9. How the ground truth for the training set was established

    This section is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1