Found 1 results

510(k) Data Aggregation

    K Number
    K131319

    Validate with FDA (Live)

    Date Cleared
    2014-05-30

    (388 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The turbines of the T1 / T2 / T3 Turbine family are intended for the

    • Preparation of cavities and crowns
    • Removal of carious material
    • Removal of fillings
    • Processing of tooth and restoration surfaces
    • Reducing hard tooth structure
    Device Description

    The Sirona Dental Systems T1 / T2 / T3 turbine family consists of air powered, high speed handpieces. The handpieces are reusable, ergonomically shaped and include (T1 and T2 Turbines) a fiber optic light system. The water delivery geometry includes 4 pair of water / air spray nozzles just below the head with one way retraction valves to prevent the ingress of external water or air, when the spray system is shut down. The devices are sterilizable in an autoclave. The turbines are provided in three variants: control, boost and mini.

    In addition, the turbines are compatible with the following quick-coupling:

    • Sirona R coupling
    • Sirona F coupling
    • Sirona B coupling
    • KaVo coupling
    • W&H coupling
    • NSK MachLite coupling
    • NSK QD-J coupling
    AI/ML Overview

    This document is a 510(k) summary for a dental device, specifically the Sirona Dental Systems T1 / T2 / T3 Turbine family. Due to the nature of the device (a dental handpiece) and the regulatory submission (510(k)), the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are discussed in the context of demonstrating substantial equivalence to predicate devices, rather than through formal clinical trials with specific performance metrics such as sensitivity, specificity, or accuracy, which are typical for diagnostic or AI-driven devices.

    Here's a breakdown based on the provided text, addressing your questions where applicable and noting where the information is not relevant to this type of device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this type of device, the "acceptance criteria" are predominantly compliance with recognized international standards and demonstration of similar technological characteristics and performance to predicate devices. The "reported device performance" is a comparison of these characteristics to the predicate devices.

    Acceptance Criteria (Implied by Standards & Predicate Comparison)Reported Device Performance (Sirona T1 / T2 / T3 Turbine family)Predicate Device 1 (Nakanishi Ti-Max X)Predicate Device 2 (Morita Twin Power)
    Principle of Operation (Air-driven turbine)Turbine, Air drivenTurbine, Air drivenTurbine, Air driven
    Couplings to Treatment Chair (Compatibility with common couplings)Sirona, Kavo, NSK, W&HSirona, Kavo, NSK, W&H, BienairSirona, Kavo, NSK, W&H, Morita
    Speed Range (Comparable to predicates)250,000 - 400,000 rpm300,000 - 450,000 rpm320,000 - 400,000 rpm
    Power [W] (Comparable to predicates)22/23/20 W16-22 W18-22 W
    Chuck (Push button FG)Push button FGPush button FGPush button FG
    Light Intensity [lux] (If applicable, comparable to predicates)25,000 luxData not available25,000 lux
    Intended Use (Matching or equivalent to predicates)-Preparation of cavities & crowns-Removal of carious material-Removal of fillings-Processing of tooth & restoration surfaces-Reducing hard tooth structure...removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth....for use by authorized persons in the practice of the dentistry.
    Compliance with ISO Standards (14457, 1797, 9168, 21531)Complies with ISO 14457, ISO 1797, ISO 9168, ISO 21531Implicitly compliant for predicatesImplicitly compliant for predicates
    Sterilization Process (Validation)Validation performedNot explicitly stated for predicates, but implied by classNot explicitly stated for predicates, but implied by class
    Device Effectiveness (Validation)Validation performedNot explicitly stated for predicates, but implied by classNot explicitly stated for predicates, but implied by class
    Biocompatibility (Validation)Validation performedNot explicitly stated for predicates, but implied by classNot explicitly stated for predicates, but implied by class
    Quick Coupling (Reverse engineering)Reverse engineering conductedNot explicitly stated for predicates, but implied by designNot explicitly stated for predicates, but implied by design

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable to this type of device submission. The submission is for a mechanical dental handpiece, and its performance is evaluated through engineering tests, compliance with standards, and comparison of technical specifications to legally marketed predicate devices, not through a "test set" of patient data or clinical cases.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable to this type of device submission. There is no concept of a "ground truth" derived from expert consensus for a mechanical dental handpiece in the way it would be for an AI diagnostic device. The ground truth, in a technical sense, would be defined by engineering specifications and material properties, validated through laboratory testing.

    4. Adjudication Method for the Test Set

    This is not applicable to this type of device submission. There is no "test set" requiring adjudication in the context of a mechanical handpiece's 510(k) submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable to this device. It is a traditional mechanical dental handpiece, not an AI-assisted diagnostic or therapeutic device. Therefore, no MRMC study, AI integration, or human reader improvement metric is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a mechanical dental handpiece; there is no algorithm, AI, or software component that operates in a standalone manner.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this device, the "ground truth" is established through adherence to international standards (ISO 14457, 1797, 9168, 21531), engineering specifications, and validation through laboratory performance testing related to sterilization, device effectiveness, and biocompatibility. There is no patient-specific "ground truth" in the clinical sense mentioned (pathology, outcomes data, expert consensus). The performance is measured against objective, quantifiable physical and sterile conditions.

    8. The Sample Size for the Training Set

    This is not applicable to this device. There is no "training set" as it is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable to this device. There is no "training set."