Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K130506

    Validate with FDA (Live)

    Device Name
    IMSE P-TLIF
    Date Cleared
    2013-05-13

    (75 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMSE P-TLIF is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). IMSEP-TLIF is to be used with autologous bone graft and implanted via an open transforaminal or posterior approach. IMSEP-TLIF implants are to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

    Device Description

    Institute of Musculoskeletal Science and Education, P-TLIF is a device for interbody fusion of the anterior column of the spine. These cages are hollow so that bone can grow through the device, fusing the adjacent bony surfaces. IMSE, P-TLIF is a hollow device with texture on two opposing convex sides, and is offered in various lengths, widths, heights and shapes. Institute of Musculoskeletal Science and Education designed the IMSE, P-TLIF to be placed through a transforaminal or posterior approach and to address vertebrae in the lumbosacral region of the spine.

    AI/ML Overview

    The provided documentation describes the IMSE P-TLIF device, an intervertebral fusion device. The focus of the submission is to demonstrate substantial equivalence to predicate devices, primarily through nonclinical testing.

    Here's an analysis of the acceptance criteria and the study results based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Performed)Reported Device Performance
    Static and Dynamic Compression Test (per ASTM F2077)Met the acceptance criteria.
    Static and Dynamic Compression Shear (per ASTM F2077)Met the acceptance criteria.
    Subsidence Test (per ASTM F2267)Met the acceptance criteria.
    Wear Debris (per ASTM F2077 and ASTM F1877)Met the acceptance criteria.
    Static Expulsion TestMet the acceptance criteria.
    Sterility Assurance Level (SAL) of 10^-6 (per ISO 17665) for Autoclaveable ImplantsValidation testing conducted (using the half-cycle method) to achieve this SAL.
    Sterility Assurance Level (SAL) of 10^-6 (per ISO 17665) for Autoclaveable Instruments and CaseValidation testing conducted (using the half-cycle method) to achieve this SAL.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided summary does not specify the sample size for the nonclinical tests (e.g., number of devices tested for compression, shear, etc.). It only states that these tests were performed to demonstrate substantial equivalence.

    Similarly, the data provenance (e.g., country of origin, retrospective or prospective) is not mentioned for these nonclinical tests. As these are laboratory tests, provenance in the clinical sense is not applicable.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the studies performed were nonclinical (mechanical and material testing), not clinical studies requiring expert ground truth for interpretation of patient data.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies performed were nonclinical, not clinical studies requiring adjudication of outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC comparative effectiveness study was not performed. The document explicitly states: "No clinical tests were performed."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable as the device (IMSE P-TLIF) is a physical intervertebral fusion device, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The ground truth for the nonclinical tests was based on standardized mechanical and material testing protocols (e.g., ASTM F2077, ASTM F2267, ASTM F1877, ISO 17665). The "ground truth" in this context refers to the specified performance parameters and thresholds defined within these standards, which the device had to meet.

    8. The Sample Size for the Training Set

    This question is not applicable as the device is a physical medical implant, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1