Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K130061

    Validate with FDA (Live)

    Date Cleared
    2013-02-27

    (48 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Piccolo Composite Diaphyseal Plate is indicated for the fixation of various long bones, such as the humerus, femur and tibia, including osteopenic bone, osteotomies, and nonunions or malunions in adult patients.

    These plates are also indicated for fracture fixation of diaphyseal areas of long bones in pediatric patients.

    Device Description

    The Piccolo Composite Plate System comprises implants (plates and screws), and a set of instruments.

    The plates are made of carbon fiber reinforced polyetheretherketone (CFR-PEEK), and are marked with a tantalum thread, to provide for their visualization under fluoroscopy.

    The Piccolo Composite Diaphyseal Plate is either 4.5mm, or 5.3mm thick; 14.5mm or 17.5mm wide; and is provided in lengths of 82 - 220mm (4 - 13 holes), depending on plates thickness and width.

    The screws are made of titanium alloy. Both locking (04.0mm and Ø5.0mm) and nonlocking (04.5mm) screws are available in a range of lengths.

    AI/ML Overview

    This document describes the mechanical testing performed to demonstrate the substantial equivalence of the Piccolo Composite® Plate System.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (ASTM F382)Reported Device Performance (Compared to Predicate Devices)
    Single Cycle (Static) 4-Point BendingMeets or exceeds performance of predicate devicesComparable to predicate devices
    Dynamic (Fatigue) 4-Point BendingMeets or exceeds performance of predicate devicesComparable to predicate devices
    Selected Screws CharacteristicsMeets or exceeds performance of predicate devicesPerformed and determined to be safe and effective

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the mechanical testing. However, it indicates that "Performance characteristics... were evaluated per ASTM F 382 - Standard Specification and Test Method for Metallic Bone Plates." This implies that the sample sizes used would have conformed to the requirements of ASTM F382 for such testing.

    The data provenance is from laboratory testing conducted by CarboFix Orthopedics Ltd. There is no information regarding country of origin of data in terms of clinical data, as this study refers to mechanical testing. This was a prospective study in the sense that the tests were specifically designed and executed to evaluate the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This study involves mechanical performance testing of a medical device against an ASTM standard and predicate devices, not the evaluation of clinical data requiring expert review for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. This is mechanical testing, not a clinical study requiring adjudication of expert opinions. The "adjudication" is met by comparing test results against the established ASTM standard and predicate device data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This document describes the mechanical testing of a bone plate system, not an AI-assisted diagnostic or therapeutic device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This document describes the mechanical testing of a bone plate system, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this study is defined by:

    • ASTM F382 - Standard Specification and Test Method for Metallic Bone Plates: This standard provides the methodologies and acceptable performance benchmarks for bone plates.
    • Performance of Predicate Devices: The measured performance of existing, legally marketed devices (Piccolo Composite® Plate System K102597, K120409, and Synthes Locking Compression Plate System e.g., K000682, K082807) served as the comparative "ground truth" to establish substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This is mechanical testing of a physical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, this is mechanical testing, not machine learning.