510(k) Data Aggregation

    K Number
    K122734

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-18

    (103 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Novosyn sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological tissues.

    Device Description

    Novosyn is a synthetic absorbable braided surgical suture which is supplied sterile. Novosyn is composed of a copolymer made from 90% gylcolide and 10% L-lactide (PGLA). The Novosyn suture is coated with 35/65 poly(glycolide-co-L-lactide) and calcium stearate. The Novosyn suture will be offered undyed or dyed with the FDA approved colorant D&C Violet No.2 in accordance with Title 21 CFR, §74.3206. The Novosyn suture will be offered in diameters ranging from USP size 8-0 through 2 and will be available in a variety of cut lengths with or without needles attached.

    AI/ML Overview

    The Aesculap Novosyn Absorbable Suture, a synthetic absorbable braided surgical suture, was tested to demonstrate its substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria (Reference Standard)Reported Device Performance
    Mechanical TestingUSP 34 (Diameter, Tensile Strength, Needle Attachment)All specifications met for tensile strength and needle attachment. Diameter specifications were not met.
    BiocompatibilityISO 10993-1Passed: Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity, Hemolysis, Genotoxicity (Chromosomal Aberration and Mouse Peripheral Blood Micronucleus, Bacterial Reverse Mutation), and Muscle Implantation (12-week).
    In-vitro ResorptionNot explicitly stated but implied to show equivalence to predicate devicesPerformed and used to demonstrate substantial equivalence.
    In-vivo ResorptionNot explicitly stated but implied to show equivalence to predicate devicesPerformed and used to demonstrate substantial equivalence.

    Note on Diameter: The document explicitly states "All specifications were met apart from diameter." However, despite this deviation, the FDA determined the device to be substantially equivalent overall. It is not specified if the deviation was deemed clinically insignificant or if additional information not provided in this summary mitigated the concern.

    2. Sample Size and Data Provenance for Test Set

    The document does not specify the exact sample sizes used for each individual test (e.g., number of sutures tested for tensile strength, number of animals for in-vivo tests). The data provenance is not explicitly mentioned (e.g., country of origin, retrospective/prospective). However, the testing was performed to support a 510(k) submission to the FDA, implying common laboratory testing standards for medical devices.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    This study does not involve human readers for diagnostic interpretation or a ground truth established by experts in the context of medical imaging or clinical assessment. The "ground truth" for this device's performance is based on established scientific and regulatory standards as outlined in USP 34 and ISO 10993-1.

    4. Adjudication Method (Test Set)

    Not applicable. This is not a study requiring adjudication of expert opinions. The performance is assessed against predefined quantitative and qualitative standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a study evaluating the physical and biological characteristics of a surgical suture, not a diagnostic imaging device requiring human reader interpretation. Therefore, no MRMC study was conducted, and no effect size for human readers improving with AI assistance is relevant.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. This is not an algorithm or AI-based device.

    7. Type of Ground Truth Used

    The "ground truth" for this device's acceptance is based on:

    • Established Standards: Mechanical properties evaluated against the United States Pharmacopeia (USP 34) standards for surgical sutures.
    • Biocompatibility Standards: Evaluated against international standards (ISO 10993-1) for biological evaluation of medical devices.
    • Comparative Performance: The overall assessment of "substantial equivalence" is the ground truth, determined by comparing its technological characteristics (design, intended use, material composition, function, range of sizes) and performance data (mechanical, biocompatibility, resorption) to legally marketed predicate devices.

    8. Sample Size for Training Set

    Not applicable. As this is not an AI/machine learning device, there is no "training set."

    9. How Ground Truth for Training Set Was Established

    Not applicable.