510(k) Data Aggregation

    K Number
    K122155

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-04

    (46 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PCH-2500 (PaX-i) is a digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, x-ray technicians and other professionals licensed by the law of the state in which the device is used.

    Device Description

    PCH-2500 (PaX-i) is a dental digital radiographic imaging system which is available in two different image acquisition modes. Specifically designed for dental radiography of the teeth or jaws, PCH-(PaX-i) can be equipped with four dedicated sensors for two different X-ray modalities : one panoramic (Xmaru1501CF), one cephalometric scan type (Xmaru2301CF) and two flat panel one shot ceph sensors (1210SGA and 910SGA). PCH-2500 (PaX-i) offers the digital panoramic and cephalometric X-ray modality for dental radiographs. The multi platforms of PCH-2500 (PaX-i) imaging mode provides a wide range of imaging options based on the patient diagnostic needs.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on the new component (910SGA sensor) as this is a Special 510(k) submission primarily addressing changes to an existing device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission is a Special 510(k) for PCH-2500 (PaX-i), primarily introducing a new cephalometric sensor (910SGA). The acceptance criteria for this sensor are implicitly set by its equivalence to the predicate device and the performance of other sensors within the PCH-2500 (PaX-i) system, as well as general performance standards for dental X-ray imaging.

    Acceptance Criteria CategorySpecific Criteria (Implicitly based on Predicate/Standards)Reported Device Performance (PCH-2500 (PaX-i) with 910SGA)
    Technical Specifications
    Pixel Resolution≥ 3.9 lp/mm (matching 1210SGA predicate sensor)3.9 lp/mm
    Pixel Size127 x 127 μm (matching 1210SGA predicate sensor)127 x 127 μm
    Image Receptor TypeAmorphous silicon TFT with scintillator (matching 1210SGA)Amorphous silicon TFT with scintillator
    SafetyCompliance with IEC 60601-1 and related standardsAll test results were satisfactory
    EMCCompliance with IEC 60601-1-2All test results were satisfactory
    PerformanceCompliance with IEC 61223-3-4 and IEC 61223-3-5All test results were satisfactory
    Image QualityImage quality comparable to predicate device (K113672) for diagnostic information of oral and maxillofacial anatomyExpert review of image comparisons showed substantial equivalence to predicate device.
    DICOM ConformanceCompliance with NEMA PS 3.1-3.18 (DICOM Set)Meets the provisions of NEMA PS 3.1-3.18
    Non-clinical PerformanceSatisfactory performance according to FDA Guidance for Solid State X-ray Imaging DevicesPerformed and satisfactory
    Clinical ConsiderationsSatisfactory considerations according to FDA Guidance for Solid State X-ray Imaging DevicesPerformed and satisfactory

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the "test set" (i.e., the images used for the expert review). It only mentions "image comparisons for both devices." The data provenance is not explicitly stated as retrospective or prospective, nor is the country of origin specified for the images used in the expert review.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document states an "expert review of image comparisons" was conducted, but it does not specify the number of experts used or their qualifications.

    4. Adjudication Method for the Test Set

    The document mentions an "expert review," but it does not describe an adjudication method (e.g., 2+1, 3+1). It implies a consensus or comparison against a standard, but no specific method is outlined.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, the document does not indicate that a Multi Reader Multi Case (MRMC) comparative effectiveness study was done. The evaluation primarily focused on demonstrating substantial equivalence to a predicate device through technical specifications, safety/EMC testing, and an expert review of image comparisons. There is no mention of human readers improving with AI assistance, as this is an imaging device, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    This question is not directly applicable in the context of this submission. The device is an X-ray imaging system, not an algorithm. Its performance is evaluated as an imaging system, which inherently produces images for human interpretation. The "non-clinical and clinical consideration report for the new SSXI detector, 910SGA" would assess the detector's capability to produce diagnostic images, effectively acting as its standalone performance.

    7. The Type of Ground Truth Used

    The ground truth for the "image comparisons" was established through expert review. This implies a comparison of diagnostic quality and features between images produced by the new 910SGA sensor and the predicate device's sensors. The ultimate "ground truth" for dental X-ray images is typically the clinical diagnosis or assessment derived by a qualified professional from those images.

    8. The Sample Size for the Training Set

    The document does not mention a training set or its sample size. This is a hardware submission, not a machine learning algorithm submission, so a traditional "training set" as understood in AI/ML is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set mentioned, this question is not applicable.