    K121034

    Validate with FDA (Live)

    Device Name
    IOI TOTAL HIP
    Date Cleared
    2012-09-04

    (152 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONACY I-Hip system is indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthrius, theumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

    The ICONACY I-Hip system consists of femoral stem and acetabular cup (i.e., shell) porous coated components intended for cementless, press-fit fixation

    Device Description

    The ICONACY I-Hip consists of a collarless, tapered, forged titanium alloy femoral stem, mated to a cobalt chrome modular femoral head. This femoral construct articulates with an acetabular device assembly. The acetabular device assembly consists of a hemispheric titanium alloy cup (i.e., shell) coupled with a highly cross linked ultra-high molecular weight polyethylene (HXL-UHMWPE) liner. Forty percent of the femoral stem is circumferentially coated with a titanium coating designed to attain a cementless, press fit fixation. The acetabular cup is machined from forged Ti-6Al-4V ELI alloy. The cup has a threaded polar hole for insertion and two screw holes to permit the use of screws for adjunct fixation. The outer hemispheric surface of the cup has a titanium plasma spray coating for cementless, press fit fixation. A titanium locking ring is fixed into a groove on the cup to engage the groove in the HXL-UHMWPE liner. Standard instrumentation is used to implant the device.

    AI/ML Overview

    The provided text describes a 510(k) submission for the ICONACY I-Hip System, a total hip replacement device. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies. Therefore, much of the requested information about device performance, study details, and ground truth establishment is not available in the provided text.

    Here's an breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Nonclinical Testing:Outcome: Demonstrated substantial equivalency to predicate devices.
    Characterization of HXL-UHMWPEPerformed
    Testing of the locking mechanism between the cup and linerPerformed
    Range of motion analysisPerformed
    Impingement testingPerformed
    WearPerformed
    Femoral head pull-off testingPerformed
    Fatigue (endurance limits) for the femoral stemPerformed

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. No clinical test set or human subject data was used. The evaluation was based on nonclinical (bench) testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    Not applicable. No expert-established ground truth for a test set was used, as no clinical testing was performed.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set or adjudication process was described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. A multi-reader multi-case comparative effectiveness study was not performed. The submission explicitly states "Clinical Testing: Clinical testing was not required for determining substantial equivalency with the predicate devices."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical orthopedic implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for demonstrating substantial equivalence was derived from nonclinical (bench) testing results (e.g., measurements of material properties, mechanical performance under various loads, wear rates) compared against established performance characteristics of the predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. There was no "training set" in the context of an AI/ML algorithm. The device's design and manufacturing processes are informed by engineering principles and existing knowledge of predicate devices.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As noted above, there was no "training set" for an algorithm. The design and manufacturing relied on established engineering standards and performance data from predicate devices.