Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120881

    Validate with FDA (Live)

    Date Cleared
    2012-04-18

    (26 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BridgePoint Medical Mantaray™ Guidewires are intended to facilitate placement of balloon dilatation catheters or other intravascular devices during percutaneous transluminal angioplasty (PTA). The Mantaray™ Guidewires are not to be used in the cerebral blood vessels.

    Device Description

    The additional models and modified Mantaray™ Guidewires are conventionally constructed, single use, disposable guidewires that consist of a full-length stainless steel shaft with proximal PTFE coating where the distal portion of the stainless steell core is taper ground to provide distal flexibility. The distal portion also includes a coaxially positioned coil constructed of platinum/tungsten material for visibility under fluoroscopy. The coil is fixed to the stainless steel core wire via silver alloy solder and is coated with hydrophilic coating. The distal tip of the guidewire is supplied with an angled geometry which transitions to a conventional rounded tip. A short extension with an approximate diameter of 0.0035"-0.0065" (which is a monolithic extension of the core wire) extrands approximately 0.007" distal of the rounded tip.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the BridgePoint Medical Mantaray™ Guidewire, focusing on modifications to an already cleared device. This submission does not include a study that proves the device meets acceptance criteria in the sense of a clinical trial or a performance study using human or image data.

    Instead, this 510(k) is for a modified version of an existing guidewire, and the acceptance criteria and supporting "study" are based on comparative bench testing and biocompatibility testing to demonstrate substantial equivalence to the predicate device. The core idea is that if the modified device performs similarly to the predicate device in key functional and safety aspects, and the changes are minor, then it is considered substantially equivalent.

    Here's an analysis of the provided information based on your requested points, recognizing that the "device" here is a guidewire, not a diagnostic or AI-driven system, so some of your points are not directly applicable.


    1. A table of acceptance criteria and the reported device performance

    The document lists performance characteristics that were evaluated for the modified Mantaray™ Guidewire and states that the device "met the established performance criteria." However, it does not provide specific numerical acceptance criteria or quantitative performance results in a table. It only lists the types of tests performed and a general statement of successful completion.

    Acceptance Criteria (General)Reported Device Performance
    Mechanical Performance:
    Tensile StrengthMet established criteria
    Dimensional AccuracyMet established criteria
    Guidewire Insert & WithdrawalMet established criteria
    FlexibilityMet established criteria
    Fatigue ResistanceMet established criteria
    Coating IntegrityMet established criteria
    Torque ResponseMet established criteria
    Surface Defect AbsenceMet established criteria
    Tip Memory RetentionMet established criteria
    RadiopacityMet established criteria
    Biocompatibility:
    CytotoxicityMet established criteria
    Kligman SensitizationMet established criteria
    IrritationMet established criteria
    Prothrombin Time AssayMet established criteria
    Complement Activation AssayMet established criteria
    In Vivo ThrombogenicityMet established criteria
    In Vitro HemocompatibilityMet established criteria
    Unactivated Partial Thromboplastin TimeMet established criteria
    PyrogenicityMet established criteria
    Acute Systemic CytotoxicityMet established criteria

    The "established performance criteria" are defined by the predicate device's performance and the general requirements for guidewires of this type. The study demonstrates that the modified device performs comparably to the predicate and meets these safety and performance benchmarks.


    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated in terms of number of guidewires tested for each category. For bench testing, typical "sample sizes" would refer to the number of units tested per configuration/material, which is not provided.
    • Data Provenance: The testing was "in vitro bench testing" and "biocompatibility tests," implying laboratory settings. There is no mention of country of origin for data.
    • Retrospective or Prospective: Not applicable for bench and biocompatibility testing. These are controlled laboratory tests performed to demonstrate device performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This is a physical medical device (guidewire), not a diagnostic algorithm or imaging system requiring expert interpretation for "ground truth." The performance criteria are objective, physical measurements and biological responses.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. No expert adjudication was required for the bench and biocompatibility test results.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a guidewire, not an AI-assisted diagnostic tool.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a guidewire, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on established engineering specifications, material science principles, and biological safety standards. Bench tests (e.g., tensile strength, flexibility, radiopacity) have objective, measurable outcomes compared against predetermined ranges or relative to the predicate device's performance. Biocompatibility tests compare material interactions against known safe biological responses.


    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning algorithm that requires a "training set."


    9. How the ground truth for the training set was established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1