Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113423

    Validate with FDA (Live)

    Date Cleared
    2012-03-09

    (109 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:

    Bleomycin sulfate (15 mg/ml), Busulfan (6 mg/ml), Carboplatin (10 mg/ml), Cisplatin (1.0 mg/ml), Cyclophosphamide (20.0 mg/ml), Cytarabine HCl (100 mg/ml), Dacarbazine (10 mg/ml), Daunorubicin HCl (5.0 mg/ml), Docetaxel (10.0 mg/ml), Doxorubicin HCl (2.0 mg/ml), Epirubicin (Ellence) (2 mg/ml), Etoposide (20.0 mg/ml), Fludarabine (25 mg/ml), Fluorouracil (50.0 mg/ml), Gemcitabine HCl (38.0mg/ml), Idarubicin HCl (1.0mg/ml), Ifosfamide (50.0 mg/ml), Irinotecan HCl (20.0 mg/ml), Mechlorethamine HCl (1.0 mg/ml), Melphalan (5 mg/ml), Methotrexate (25 mg/ml), Mitomycin-C (0.5 mg/ml), Mitoxantrone (2.0 mg/ml), Paclitaxel (6.0 mg/ml), Rituximab (10 mg/ml), ThioTEPA (10.0 mg/ml), Trisenox (0.1 mg/ml), Vincristine Sulfate (1.0 mg/ml)

    The following drugs had NO breakthrough detected up to 240 minutes:

    Please note that the following drug has low permeation times of less than 60 minutes: Carmustine (3.3 mg/ml) 30.7 minutes

    Device Description

    Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves are 9.5-inch long, non-sterile, purplecolored nitrile, powder-free, ambidextrous patient examination glove that meets the specifications of ASTM D 6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

    Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves are 12-inch long, non-sterile, purplecolored nitrile, powder-free, ambidextrous patient examination glove that meets the specifications of ASTM D 6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

    Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves are 12-inch long, non-sterile, purplecolored nitrile, powder-free, ambidextrous patient examination glove that meets the specifications of ASTM D 6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the drugs listed in the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

    These gloves meet the 2008 Glove Guidance Manual recommended minimum thickness and length specifications for gloves tested for use with chemotherapy drugs.

    AI/ML Overview

    The provided 510(k) summary (K113423) describes the non-clinical testing performed for Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves (9.5" Length), Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (12" Length), and Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (12") Tested for Use with Chemotherapy Drugs. As these are patient examination gloves, the evaluation focuses on their physical properties, safety, and, for one specific type, resistance to chemotherapy drugs, rather than AI/ML performance. Therefore, many of the requested categories related to AI/ML device studies (such as MRMC studies, human-in-the-loop performance, expert qualifications for ground truth, and training set information) are not applicable.

    Here's an analysis of the acceptance criteria and study findings based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Applicable Standard)Reported Device Performance
    DimensionsASTM D 6319-10 (and 2008 FDA Glove Guidance Manual for length/thickness for chemo gloves)Meets ASTM Requirements (and 2008 FDA Glove Guidance Manual for chemo gloves)
    Physical PropertiesASTM D 6319-10Meets ASTM Requirements
    Freedom from pinholesASTM D 6319-10, ASTM D 5151-06Meets ASTM Requirements
    Powder Free (Powder Content)ASTM D 6319-10, ASTM D 6124-06Meets ASTM Requirements
    Skin Irritation Study & SensitizationISO 10993, Part 10Meets ASTM Requirements
    Systemic Toxicity StudyISO 10993, Part 11Meets ASTM Requirements
    Resistance to Permeation (for chemo gloves only)ASTM D 6978-05, ASTM F 739-07Meets ASTM Requirements (Specific drug permeation times provided)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific test (e.g., number of gloves tested for dimensions, pinholes, etc.). It generally states that testing was conducted according to the referenced ASTM and ISO standards. These standards typically define sampling plans.

    The data provenance is from non-clinical laboratory testing conducted by Kimberly-Clark Corporation, likely within the United States (as the applicant is based in Roswell, GA, and submitted to the U.S. FDA). The studies are prospective in the sense that the testing was performed specifically to evaluate the performance of these new glove models against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable for this type of device. The "ground truth" for glove performance is established by the specified ASTM and ISO standards, which dictate methodologies and acceptable limits for physical and chemical properties. Human experts are involved in conducting these standardized tests and interpreting results against predefined criteria, but not in a "ground truth" sense as understood in AI/ML validation.

    4. Adjudication Method for the Test Set

    Not applicable. The performance is measured against objective, standardized laboratory test methods, not subjective human assessments requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical glove, not an AI/ML-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance of these gloves is defined by established consensus standards (ASTM D 6319-10, ASTM D 5151-06, ASTM D 6124-06, ISO 10993 Parts 10 & 11, ASTM D 6978-05, ASTM F 739-07, and the 2008 FDA Glove Guidance Manual). These standards specify the methods and acceptable performance limits for various physical, chemical, and biological properties of examination gloves. For the chemotherapy-tested gloves, the specific breakthrough times for various drugs further contribute to the performance "ground truth."

    8. The Sample Size for the Training Set

    Not applicable. There is no AI/ML model involved; therefore, no training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no AI/ML model or training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1