Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111571

    Validate with FDA (Live)

    Device Name
    VIPER SYSTEMS
    Manufacturer
    Date Cleared
    2011-07-06

    (30 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VIPER Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.

    The VIPER Systems metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.

    When used in a posterior percutaneous approach with MIS instrumentation, the VIPER Systems metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudarthrosis; and failed previous fusion in skeletally mature patients.

    The VIPER PEEK rods are only indicated for fusion procedures for spinal stenosis with instability (no greater than Grade I spondylolisthesis) from L1-S1 in skeletally mature patients.

    Device Description

    The VIPER Systems are 5.5mm rod systems offered in both titanium and stainless steel materials. The systems consist of monoaxial screws, polyaxial screws, uni-planar screws, and extended tab implants. They are available in various geometries and sizes to accommodate patient anatomy. The VIPER Systems 8mm to 12mm diameter favored angle polyaxial screws are not intended to be used with the VIPER PEEK rods.

    AI/ML Overview

    The provided text outlines the 510(k) premarket notification for the DePuy Spine VIPER Systems, focusing on its substantial equivalence to predicate devices and performance data. It does not describe an AI medical device or a study involving acceptance criteria with sensitivity, specificity, or human reader performance. Instead, it describes a medical device (Pedicle Screw Spinal System) and its mechanical performance testing.

    Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance related to AI, sample sizes for test sets or training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as these concepts are not addressed in the provided document.

    The document describes Performance Data for the VIPER Systems, which are mechanical tests rather than clinical or AI performance evaluations:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test TypeAcceptance CriteriaReported Device Performance
    Mechanical Performance (ASTM F 1798)The device must demonstrate mechanical properties (static and dynamic cantilever bending, axial pull-out, torsion, and static axial slip) that are substantially equivalent or superior to the identified predicate devices, ensuring it can withstand the expected mechanical loads in vivo without failure or compromise to its intended function. The specific quantitative criteria (e.g., stiffness thresholds, fatigue limits, pull-out force minimums) are not explicitly detailed in the summary but would be defined in the full test protocol.The performance data submitted "characterize the subject VIPER Systems components." The Conclusion states: "Both the Performance Testing and Substantial Equivalence Justification demonstrate that the device is as safe, as effective, and performs as well as the predicate device." This implies the tests met the established criteria for substantial equivalence to the predicate devices.
    BiocompatibilityBiocompatibility requirements per ISO 10993 (or equivalent) for implantable devices.While not explicitly detailed as "reported performance" in this summary, the use of "ASTM F 138 implant grade stainless steel, ASTM F 139 implant grade stainless steel, and ASTM F 136 implant grade titanium alloy" for manufacturing indicates compliance with accepted biocompatibility standards for these materials.

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not specified in the provided text. Mechanical testing typically involves a set number of samples per configuration (e.g., 5-10 samples per test condition) to satisfy statistical requirements for device qualification.
    • Data provenance: The performance data was generated through "Performance data per ASTM F 1798" and submitted by DePuy Spine, Inc. This indicates laboratory testing of physical device components, not clinical data from patients or data from specific countries.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable. The "ground truth" for mechanical testing is based on the physical properties and performance of the device components as measured by test equipment, not on expert interpretation of results in the way clinical diagnostic accuracy studies require.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable. Adjudication methods are used in studies involving human interpretation or clinical outcomes, not for standard mechanical testing results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. This document describes a medical device (spinal pedicle screw system) and its mechanical performance, not an AI medical device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as the device is a physical pedicle screw system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for mechanical testing is based on objective, quantifiable measurements (e.g., load at failure, displacement, fatigue cycles) obtained from laboratory test equipment, adhering to ASTM F 1798 standards. There is no expert consensus, pathology, or outcomes data used to establish ground truth for this type of performance evaluation.

    8. The sample size for the training set:

    • This is not applicable as the document describes a physical medical device and its mechanical testing, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • This is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1