Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110393

    Validate with FDA (Live)

    Date Cleared
    2011-03-04

    (21 days)

    Product Code
    Regulation Number
    864.5620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hemo-Control Hemoglobin Measurement System is indicated for the quantitative determination of hemoglobin in arterial, venous, or capillary blood.

    The microcuvettes part number 3000-3012-0765 are indicated for use in the Hemo-Control Hemoglobin Measurement System and compatible measurement systems. The microcuvettes are intended to be used only once and must be disposed of after use as potentially infectious waste.

    Estimation of hematocrit as a function of Hemoglobin is performed for normal hemoglobin ranges only (120 to 180 g/liter or 12.0 to 18.0 g/deciliter). The estimated hematocrit is not indicative of disease states such as anemia and abnormal values will not be reported.

    For In Vitro Diagnostic Use Only

    Device Description

    The Hemo-Control Hemoglobin Measurement System is comprised of a Hemo-Control Hemoglobin Measurement Photometer and Hemo-Control Microcuvettes. The scope of this 510(k) is limited to a modification of the microcuvettes.

    The Hemo-Control Microcuvettes are single-use microcuvettes filled with dry reagents. A modified azide methemoglobin method is used. The use of microcuvettes with short light pathways makes it possible to analyze undiluted blood. The filled microcuvette is inserted into the Hemo-Control Hemoglobin Measurement Photometer, the color produced by the chemical reaction in the microcuvette is measured, and the Hb level is calculated and displayed. Light emitting diodes (LED's) are used as light sources with a photodiode to detect the light.

    The plastic microcuvette consists of a clear body with a cavity which takes up approximately 10 µL of blood which combines with the dry reagent chemistry. The optical distance between the microcuvette walls is fixed and permits photometric determination of the hemoglobin in undiluted blood samples using the Lambert-Beers Law. The microcuvette optical and chemical characteristics are unchanged by the modification.

    The Hemo Control Hemoglobin Measurement System with the microcuvette modification employs the identical fundamental scientific technology as the predicate device(s).

    AI/ML Overview

    The provided documentation describes the acceptance criteria and a study demonstrating that the modified Hemo_Control Microcuvettes meet these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The study primarily focuses on precision and correlation with a reference method. The acceptance criteria can be inferred from the reported precision (low CV%) and strong correlation (high R^2).

    Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied)Reported Device Performance (Hemo_Control modified microcuvette)
    PrecisionWithin-Run CV%Low CV% (e.g., < 2%)0.7 % (8.0 g/dL), 0.6 % (11.8 g/dL), 0.5 % (15.7 g/dL)
    Total CV%Low CV% (e.g., < 2%)1.5 % (8.0 g/dL), 1.4 % (11.8 g/dL), 1.1 % (15.7 g/dL)
    Between-Day CV%Low CV% (e.g., < 2%)1.4 % (8.0 g/dL), 1.5 % (11.8 g/dL), 1.1 % (15.7 g/dL)
    Accuracy/CorrelationR^2High (e.g., > 0.95)0.9955 (compared to CLSI H15-A3 reference method)
    Linearity/BiasRegression slopeClose to 1.0 (e.g., 0.95-1.05)1.0086
    Regression interceptClose to 0.00.2929

    Note: The acceptance criteria are "implied" as specific numeric thresholds are not explicitly stated as "acceptance criteria" but rather as "results obtained" that demonstrate substantial equivalence.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Test Set: 100 samples
    • Data Provenance: The document does not explicitly state the country of origin. The test used venous blood samples. It is a retrospective analysis in the sense that the samples were already collected and then tested. The study design appears to be a laboratory evaluation (non-clinical test).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This study is a non-clinical laboratory test comparing a device's performance to a reference method, not an interpretative study requiring human expert ground truth.

    • Number of Experts: Not applicable, as the ground truth was established by a reference method.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    Not applicable. This was a direct comparison to a reference method, not an interpretative study requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a non-clinical laboratory evaluation of a modified medical device against a reference standard, not an AI-assisted diagnostic tool involving human readers.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The precision and correlation data presented represent the performance of the modified microcuvette with the Hemo_Control device, without human interpretation as part of the measurement process. It's a "device only" performance evaluation.

    7. Type of Ground Truth Used

    The ground truth was established by a reference method, specifically the CLSI H15-A3 standard for the quantitative determination of hemoglobin in blood.

    8. Sample Size for the Training Set

    Not applicable. This device is a measurement system, and the study described is a validation study for a device modification, not a machine learning model that requires a training set. The "device" already has its algorithm embedded.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for a machine learning model was used.