Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103662

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-15

    (121 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The system is indicated for primary or revision surgery in skeletally mature individuals for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

    The system is intended for use either with or without bone cement in total hip arthroplasty.

    Device Description

    The proposed Continuum and Trilogy IT Acetabular System Shells are a line extension to the predicate modular acetabular shells, with a smaller 40 mm outer diameter. The Longevity IT Elevated Liner in this submission is offered in a size to be used with the 40 mm Continuum and Trilogy IT Shells above and has a 22 mm articulation diameter. The Longevity IT Offset and Oblique Liners are modular acetabular cup liners intended to be used with the Continuum and Trilogy IT Acetabular Systems. Offset Liners have a head center that is offset 7 mm beyond the acetabular shell center of rotation and the Oblique Liners have a liner face and head center tilted by 10° relative to the acetabular shell rim.

    AI/ML Overview

    The provided document describes the Zimmer Continuum® and Trilogy® IT Acetabular Systems & Longevity® IT Highly Crosslinked Polyethylene Elevated, Offset, and Oblique Liners. This is a medical device for total hip arthroplasty. The document details the non-clinical performance and conclusions, but explicitly states that clinical data and conclusions were not needed for this device. This means there was no study involving human subjects or expert readers to evaluate its performance against specific acceptance criteria in a clinical setting as would typically be done for AI/diagnostic devices.

    Therefore, many of the requested elements for an AI/diagnostic device's acceptance criteria and study design are not applicable to this submission.

    Here's an analysis based on the provided text, addressing the applicable points:

    1. A table of acceptance criteria and the reported device performance

    Since this is a non-clinical evaluation of a hip prosthesis, the "acceptance criteria" are related to mechanical and physical performance rather than diagnostic accuracy. The document lists the specific non-clinical tests conducted, implying that the device "met performance requirements." However, the specific quantitative acceptance criteria for each test (e.g., "liner push-out force must exceed X Newtons") and the actual numerical results are not explicitly detailed in this summary.

    Acceptance Criteria Category (Implied)Reported Device Performance
    Liner Push-Out EvaluationMet performance requirements
    Liner Lever-Out EvaluationMet performance requirements
    Liner Torque-Out EvaluationMet performance requirements
    Rim Deformation TestingMet performance requirements
    Liner Locking Mechanism Strength AnalysisMet performance requirements
    Anatomic Fatigue TestingMet performance requirements
    Liner Wear PerformanceMet performance requirements
    Liner Durability and Backside WearMet performance requirements
    Temperature Effects on Liner AssemblyMet performance requirements
    Interaction of MRI with ImplantsMet performance requirements
    Range of Motion EvaluationMet performance requirements
    Shell Fatigue and Deformation EvaluationMet performance requirements

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided because the study was non-clinical (mechanical testing of physical devices), not a clinical study involving patients or data. The "test set" would refer to the number of physical liners and shells subjected to each mechanical test. This specific quantity is not disclosed in the summary document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth for mechanical testing is established by engineering specifications and measurements, not expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are relevant for subjective human assessments, not objective mechanical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical hip prosthesis, not an AI diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For non-clinical performance, the "ground truth" implicitly refers to engineering specifications, material science standards, and biomechanical requirements for an implantable device to function safely and effectively within the human body. For example, a "liner push-out" test would have an engineering specification for the minimum force required to dislodge the liner, and the measured force would be compared against that.

    8. The sample size for the training set

    Not applicable. This is a physical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical device, not an AI model.

    In summary: The provided document is a 510(k) premarket notification for a medical device (hip prosthesis) based entirely on non-clinical performance testing. It states that "Clinical data and conclusions were not needed for this device" because it is a line extension of previously cleared predicate devices and it met performance requirements through various mechanical and material tests. Therefore, questions related to clinical studies, expert readers, AI performance, and AI training/ground truth are not relevant to this specific submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1