Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K102710

    Validate with FDA (Live)

    Device Name
    NEO-SERVO I
    Date Cleared
    2010-12-22

    (93 days)

    Product Code
    Regulation Number
    880.5400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neo-Servo i Atom Infant Incubator Model 102 is a closed-type incubator for newborns and premature neonates. It is mainly intended for temperature control in neonatal hypothermia, observation and examination in newborn nurseries, prevention of body temperature drop shortly after delivery, and pre-operative and post-operative intensive care in neonatal surgery. The incubator is provided with a function to control the infant's skin temperature.

    Device Description

    The Neo-Servo i Atom Infant incubator Model 102 consists of a hood, a main body, a control panel and is offered with or without a stand with casters. The Model 102 incorporates two (2) modes of control: air (manual control) and baby (servo control).

    The Model 102 incorporates the sensors (humidity and temperature) inside the hood monitor the air conditions inside the hood. These are connected to the Sensor module. It is equipped with an incubator air temperature control function to circulate air containing the heat generated b the heater attached to the conditioning chamber inside the hood by means of a fan in order to maintain the incubator air temperature at a fixed level. It is equipped with a skim temperature control function o maintain the infant's skin temperature at a fixed level in response to the patient's temperature as measured by the skin probe. There are optional humidity chamber and x-ray cassette tray.

    There are 2 functional modes:

    • Skin temperature control mode referred to as "Servo Control Mode" 1.
    • Air temperature control mode referred to as "Manual Control Mode" 2.
    AI/ML Overview

    The provided text describes the Neo-Servo i Atom Infant Incubator Model 102 and its substantial equivalence to a predicate device (Atom Model V-2200 – K061856). The information focuses on design, indications for use, and adherence to performance standards rather than a study proving the device meets specific acceptance criteria through a clinical trial or AI model evaluation.

    Therefore, many of the requested categories related to AI performance, sample sizes, expert ground truth, and adjudication methods are not applicable to this document. The document primarily demonstrates substantial equivalence by comparing the device's features and compliance with recognized safety and performance standards.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents a comparison to a predicate device, which implicitly defines "acceptance criteria" as meeting or exceeding the predicate's performance and complying with relevant standards.

    Feature / CriterionAcceptance Criteria (Implicit from Predicate & Standards)Reported Device Performance (Neo-Servo i Atom Infant Incubator Model 102)
    Indications for UseIntended for temperature control, observation, and critical care for newborns/premature neonates, with skin temperature control.Identical to predicate: temperature control in neonatal hypothermia, observation/examination, prevention of body temperature drop, pre/post-operative intensive care; skin temperature control.
    Environment of UseHospital, delivery suites, NICU.Identical to predicate: Hospital, delivery suites, NICU.
    Patient PopulationInfants and neonates ≤ 10 kg.Identical to predicate: Infants and neonates ≤ 10 kg.
    Technology and DesignSimilar to predicate (Atom V-2200).Identical technology and design to Atom V-2200.
    Materials (patient contact)Similar to predicate.Identical to predicate K061856.
    Skin Temperature Setting Range (Servo)34 - 37.5 °C (override 37.6 - 38.0 °C).Identical: 34 - 37.5 °C (override mode 37.6 - 38.0 °C).
    Incubator Air Temperature23 - 39 °C (including override).Identical: 23 - 39 °C (including override).
    Display Range (Skin)30.0 - 42 °C, Accuracy ± 0.3 °C.Identical: 30.0 - 42 °C, Accuracy ± 0.3 °C.
    Display Range (Air)20.0 - 42.0 °C, Accuracy ± 0.3 °C.Identical: 20.0 - 42.0 °C, Accuracy ± 0.3 °C.
    Heater Output0 - 100% in 10 increments.Identical: 0 - 100% in 10 increments.
    Warm-up Time≤ 60 min @ 25°C.Identical: ≤ 60 min @ 25°C.
    Humidity Setting Range40 - 95% RH.Identical: 40 - 95% RH.
    Humidity Display15 - 99% RH.Identical: 15 - 99% RH.
    Continuous Humidity~ 8 hours.Identical: ~ 8 hours.
    Oxygen Supply (Max Concentration)≥ 65% @ 10 Lpm.Identical: ≥ 65% @ 10 Lpm.
    Environment CO2 Concentration< 0.4% at center of mattress at 750 ml/min.Identical: < 0.4% at center of mattress at 750 ml/min.
    Safety and Performance Standards ComplianceCompliance with IEC 60601-1, IEC 60601-1-2, IEC 60061-2-19."Performance testing has been performed according to recognized standards and the proposed device met and passed all testing criteria." (Specific results not detailed in this summary).

    2. Sample Sizes Used for the Test Set and Data Provenance

    This information is not provided in the document. The submission relies on performance testing against standards, not a test set of data in the context of an AI/algorithm evaluation.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable as this is a medical device (incubator) submission based on engineering performance testing and substantial equivalence, not an AI/algorithmic device requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    Not applicable for the same reasons as #3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document does not mention any MRMC study or AI assistance.

    6. If a Standalone (Algorithm Only) Performance Study Was Done

    No. This device is a physical incubator, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to the defined performance specifications within recognized standards (IEC 60601 series) and the performance characteristics of the predicate device. The device's performance was measured against these established engineering and safety benchmarks.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reasons as #8.


    Summary of the Study Proving Acceptance Criteria:

    The "study" demonstrating the Neo-Servo i Atom Infant Incubator Model 102 meets acceptance criteria is primarily a series of engineering performance tests and assessments against recognized international standards (IEC 60601-1, IEC 60601-1-2, IEC 60061-2-19) and a direct comparison to a legally marketed predicate device (Atom Model V-2200, K061856) to establish substantial equivalence.

    The document states: "Performance testing has been performed according to recognized standards and the proposed device met and passed all testing criteria." This indicates that the device underwent rigorous bench testing to ensure it conforms to the safety and performance requirements outlined in these standards, and that its functional specifications match or are acceptably similar to the predicate device. The detailed results of these tests are not included in this summary but would have been part of the full 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1