510(k) Data Aggregation

    K Number
    K102592

    Validate with FDA (Live)

    Device Name
    SUTRAZORB
    Date Cleared
    2011-02-03

    (147 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sutrazorb Absorbable Surgical Sutures USP / EP are indicated for use in general soft tissue approximation and / or ligation in ophthalmic procedures. The safety and effectiveness of Sutrazorb sutures in cardiovascular and neurological procedures have not been established.

    Device Description

    The Sutrazorb Absorbable PGA Suture (Sutrazorb Suture) is a sterile, surgical suture composed of monofilaments and braids of polyglycolic acid. The suture is dyed violet. The violet sutures are dyed with D&C Violet #2 in accordance with CFR Title 21 Part 74.3602, reference Section 6.0.

    The Visionary Medical Supplies' Sutrazorb Suture meets all requirements established by the United States Pharmacopoeia for absorbable surgical sutures. Testing to the following USP 32:2009 Monographs for Absorbable Sutures is discussed in Section 6.0:

    • Sutures - Diameter <861>
    • Sutures Needle Attachment <871>
    • Tensile Strength <881>
    AI/ML Overview

    The provided text describes a 510(k) summary for the Sutrazorb Absorbable PGA Sutures. The document focuses on demonstrating substantial equivalence to predicate devices through performance testing.

    Here's an analysis of the acceptance criteria and the study as described in the input:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (USP Monograph)Reported Device Performance
    Sutures - Diameter <861>Meets USP standards
    Sutures Needle Attachment <871>Meets USP standards
    Tensile Strength <881>Meets USP standards
    Absorption profileMeets USP standards
    Shelf lifeMeets USP standards

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for each of the performance tests (diameter, tensile strength, needle attachment, absorption profile, shelf life). It only states that tests were conducted according to USP monograph methods.

    The data provenance is not explicitly mentioned (e.g., country of origin). The studies appear to be conducted by or for Visionary Medical Supplies, Inc. and are retrospective analyses against established USP standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable. The "ground truth" for this device is based on objective, standardized physical and chemical properties defined by the United States Pharmacopoeia (USP) monographs, not on expert interpretations of medical images or clinical data.

    4. Adjudication Method for the Test Set

    Not applicable, as ground truth is established by objective USP standards, not by expert consensus or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study (MRMC) is typically performed for AI-driven diagnostic or screening devices to assess how AI assistance impacts human reader performance. The Sutrazorb Absorbable PGA Sutures are a medical device (surgical sutures), and their evaluation focuses on physical and material properties, not diagnostic accuracy or human interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, in essence. The performance tests (diameter, tensile strength, etc.) are conducted directly on the device itself, without human interpretation as part of the primary measurement. The evaluation of the results against USP standards is a human process, but the performance data itself is intrinsic to the device.

    7. Type of Ground Truth Used

    The ground truth used is objective, standardized criteria defined by the United States Pharmacopoeia (USP) Monographs for absorbable surgical sutures. These monographs provide specific methods and acceptance limits for material properties like diameter, tensile strength, and needle attachment, as well as characteristics like absorption profile and shelf life.

    8. Sample Size for the Training Set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI. The tests are direct measurements of the product's physical and chemical properties against established standards.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device. The standards against which the device is evaluated are established by the United States Pharmacopoeia.