510(k) Data Aggregation

    K Number
    K101717

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-08-20

    (63 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Karam STD, Karam Mini, Karam ECN, AeroBreeze STD, and AeroBreeze Mini models are airpowered highspeed dental handpieces with intended use of preparing dental cavities for restorations, such as fillings, reducing hard tooth structure, and cleaning teeth. Karam STD, Karam Mini, AeroBreeze STD, and AeroBreeze Mini can be used with either Type B (Karam 4B) or Type A (Karam 2A) coupling. Karam ECN model connects directly to either Type A or Type B hose connectors and does not require a coupling.

    Karam45 and Dexor45 are air-powered highspeed dental handpieces with intended use of being a surgical tool for impacted third molar removal and periodontal procedures for which a conventional handpiece would be used. Both models can be used with either Type B (Karam 4B) or Type A (Karam 2A) coupling.

    Device Description

    The devices are an air-powered dental handpieces intended for use with a friction-grip bur that conforms to ISO 1797-1 standard. Recommended air pressure is 29 ~ 36 psig (gage pressure) to produce high-speed bur rotation ranging between 380,000 and 450,000 RPM. Karam Mini, Karam ECN, AeroBreeze STD, and AeroBreeze Mini models are straight-headed. Karam45 and Dexor45 models are designed with a 45-degree back angle. The devices include a water line that directs water mist/jet to the cutting area for cooling and irrigation.

    Aerobine line of handpieces employs a cartridge type air turbine with a common core design across the product line. All models employ a friction-grip push button chuck, a drive air line, a discharge air egress line, a water delivery line to supply cooling water. With the exception of Karam ECN, all models can be used with either Karam 4B or Karam 2A coupling. Karam ECN connects directly to Type A or Type B hose connectors without a coupling.

    AI/ML Overview

    The provided text describes Aerobine Line of Dental Handpieces and compares them to predicate devices for a 510(k) submission. It does not contain a detailed study with acceptance criteria and reported device performance in the format requested. The document primarily focuses on establishing substantial equivalence to predicate devices based on technological characteristics.

    Therefore, many sections of your request cannot be fully answered.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    The document lists "Minimum Performance requirements" as a technological characteristic and states that the Aerobine devices are "Identical" to the predicate devices for these requirements. However, it does not specify the quantitative acceptance criteria or the actual reported performance values for the Aerobine devices.

    Acceptance CriteriaReported Device Performance (Aerobine Line of Dental Handpieces)
    Speed rangeIdentical to predicate devices (K062740 and K972376)
    Stall torqueIdentical to predicate devices (K062740 and K972376)
    Cooling water flow rateIdentical to predicate devices (K062740 and K972376)
    Noise levelIdentical to predicate devices (K062740 and K972376)
    Bur extraction forceIdentical to predicate devices (K062740 and K972376)
    EccentricityIdentical to predicate devices (K062740 and K972376)
    Sterility (Autoclave up to 135°C 15 minutes at 134°C)Unknown (for Karam STD/Mini/ECN, AeroBreeze STD/Mini vs. K062740)
    Sterility (Autoclave up to 135°C 15 minutes at 134°C)Autoclave up to 135°C 20 minutes at 135°C (for Karam45, Dexor45 vs. K972376)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document refers to "engineering analysis results obtained in support of the sterilization validation effort (section 14)," but section 14 is not included, and no details about specific test sets, sample sizes, or data provenance are mentioned for performance criteria.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. The document does not describe a study involving expert assessment or ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/not provided. The device is an air-powered dental handpiece, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable/not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable/not provided. The document indicates that substantial equivalence was determined based on "technological profile comparisons (Table 1) and engineering analysis results."

    8. The sample size for the training set

    This information is not provided. There is no mention of a "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    This information is not provided. There is no mention of a "training set" or ground truth establishment in this context.