Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101572

    Validate with FDA (Live)

    Date Cleared
    2011-08-01

    (423 days)

    Product Code
    Regulation Number
    866.5750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.

    Device Description

    IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk. The allergens are covalently bound to a soluble polymer/co-polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support. Incubation Cycles: 2 × 30 minutes.

    AI/ML Overview

    The provided document describes the Siemens Healthcare Diagnostics Inc. IMMULITE® 2000 3gAllergy™ Specific IgE Assay, which is an in vitro diagnostic device for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The document focuses on demonstrating the substantial equivalence of nine additional specific allergens for use with this existing assay.

    Here's an analysis of the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria values for each performance characteristic (e.g., "Total CV must be < 10%"). Instead, it presents the results of the studies and concludes that the analytical performance is "acceptable" and the assay "compares well with clinical documentation." However, we can infer some criteria based on industry standards for similar assays and the presented data.

    Performance CharacteristicAcceptance Criteria (Inferred from CLSI standards & acceptability)Reported Device PerformanceConclusion
    Precision (Total CV)Typically < 15% for quantitative assays, with smaller CVs for lower concentrations and closer to the medical decision point.%CV values for all allergens and positive samples generally range from 3.65% to 9.29% (see detailed table in document).Accepted. All reported Total CVs are within generally accepted analytical precision limits.
    Linearity (Regression)Regression equation close to Y=X (slope ~1, intercept ~0), with narrow 95% Confidence Intervals for slope and intercept.Slopes range from 0.966 to 1.011. Intercepts range from -0.0099 to 0.1233. 95% CIs for slopes generally include 1; for intercepts, they generally include 0 or are small.Accepted. Demonstrated linearity within assay limits.
    Specificity (Inhibition)Target % inhibition of 50% met by relevant inhibitor extract in a concentration-dependent fashion.The inhibition study demonstrated that the allergens tested are inhibited by the relevant inhibitor extract in a concentration dependent fashion. The target % inhibition of 50% was met.Accepted. Confirmed specificity for each allergen.
    Specificity (Cross-reactivity - Inhibition using Negative Controls)Results below a certain percentage (e.g., < 20%) when tested with unrelated allergens.Results for all specific allergens (Basil, Cacao, Oregano, Parsley, Pine Nut, and Vanilla) were below 19.3% for inhibition by unrelated allergens.Accepted. Demonstrated no significant cross-reactivity with unrelated allergens.
    Clinical SensitivitySufficiently high to identify true positive cases. (No specific threshold given, but relative to predicate).50.9% (Lower Conf. 44%, Upper Conf. 57%)Accepted, within the context of an aid to diagnosis and to be used with other clinical findings. This Sensitivity is applied to the overall clinical data agreement, not a standalone diagnostic sensitivity.
    Clinical SpecificitySufficiently high to identify true negative cases. (No specific threshold given, but relative to predicate).100.0% (Lower Conf. 100%, Upper Conf. 100%)Accepted. High specificity, indicating few false positives against clinical data.
    Clinical AgreementSufficient overall agreement with clinical data. (No specific threshold given, but relative to predicate).86.5% (Lower Conf. 84%, Upper Conf. 89%)Accepted, supporting the intended use when combined with other clinical information.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision and Linearity: The document does not explicitly state the total number of individual patient samples used for precision and linearity.
      • Precision: For each allergen, "three positive samples and one negative sample" were tested. Positives 1-3 were assayed in two aliquots, two runs/day over 20 days. Positive 4 was assayed in two aliquots, two runs/day over 10 days.
      • Linearity: For each allergen, "two samples were diluted in 2-fold serial dilutions to 5 levels." N refers to the number of data points for regression, which ranges from 8 to 12.
    • Specificity (Inhibition): A "single serum sample or pool of sera" was used for competitive inhibition. A "negative sample" was used for background. For cross-reactivity, "one positive sample with three unrelated allergen extracts" was used.
    • Clinical Performance:
      • Test Set Size: 828 total samples (228 clinically positive, 600 clinically normal/negative).
      • Data Provenance: Not explicitly stated regarding country of origin. The study appears to be retrospective, as results were compared to "accompanying clinical information" and "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity," implying existing patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts who established the "clinical data" or "clinical documentation" used as ground truth for the clinical performance study. It refers generally to "clinical findings and other laboratory tests" and "signs, symptoms and other diagnostic evidence of allergen sensitivity."

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (e.g., 2+1, 3+1) for establishing the clinical ground truth. It simply refers to "clinical data" and "clinical documentation."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is an in vitro diagnostic assay, not an imaging device or AI-powered diagnostic that involves human readers interpreting results in a "multi-reader multi-case" scenario. Therefore, an MRMC comparative effectiveness study involving human readers with/without AI assistance is not applicable and was not performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is a standalone diagnostic assay (IMMULITE® 2000 3gAllergy™ Specific IgE Assay) which provides quantitative measurements of IgE. Its performance is presented as the "algorithm only" or "device-only" performance, as it reports a numerical result. The assay is intended to be used "in conjunction with clinical findings and other laboratory tests," indicating that a human (clinician) interprets the result within the broader clinical context, but the assay itself is not a human-in-the-loop system in the sense of AI-assisted image interpretation.

    7. The Type of Ground Truth Used

    • Analytical Studies (Precision, Linearity, Specificity): The ground truth for these studies is based on established analytical methods and reference materials (e.g., known concentrations for linearity, relevant inhibitor extracts for specificity).
    • Clinical Performance Study: The ground truth used was clinical documentation and findings, including "case histories of suspected clinical reactions to the specific allergen or allergy" for allergic patients, and classification as "non-atopic individuals" for controls. This falls under outcomes data or expert clinical consensus based on patient history, symptoms, and other diagnostic evidence, rather than a single definitive pathology report for each case.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of machine learning or AI development. This is a traditional immunoassay, not an AI/ML algorithm that requires a separate training set. The data presented demonstrates the analytical and clinical performance of the manufactured assay.

    9. How the Ground Truth for the Training Set was Established

    As this is not an AI/ML device, there is no "training set" in that sense, and therefore no ground truth was established for a training set. The assay's design and parameters would have been developed through R&D processes, likely involving internal validation with characterized samples.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1