    Intended Use

    The Dual Luer Lock Cap is indicated for use as a cap for male or female Luer ports on medical devices such as manifolds, stopcocks or sets.

    Device Description

    The subject of this submission is a sterile Dual Luer Lock Cap which will be used to cover male or female Luer ports on medical devices. The proposed Dual Luer Lock Cap will replace the existing Dual Luer Lock Cap and will be sold as a cap to cover an open Luer port after the Luer port has been accessed and is no longer in use. The Dual Luer Lock Cap consists of an integrated design with a male Luer lock connection on one end and a female Luer lock connection on the other end.

    AI/ML Overview

    Below is the requested information about the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided text.

    Acceptance Criteria and Device Performance Study

    This device, the Dual Luer Lock Cap, is an IV Administration Set component. The performance testing focused on meeting recognized international standards for Luer fittings.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test)StandardReported Device Performance
    Air leakageISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
    Separation ForceISO 594-2:1986Met (implied by "All test results meet the acceptance criteria")
    Unscrewing torqueISO 594-2:1986Met (implied by "All test results meet the acceptance criteria")
    Ease of assemblyISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
    Stress crackingISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
    Liquid leakage testISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
    Luer gaugingISO 594-1:1986Met (implied by "All test results meet the acceptance criteria")
    Resistance to overridingISO 594-2:1986Met (implied by "All test results meet the acceptance criteria")
    BiocompatibilityISO 10993-1Shown to be biocompatible and appropriate for intended use (prolonged duration, external communicating, indirect blood path testing)

    Note: The document states that "All test results meet the acceptance criteria," indicating successful performance without providing specific quantitative results for each test.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the performance tests (e.g., how many Dual Luer Lock Caps were tested for each criterion). The data provenance is also not specified in terms of country of origin or whether it was retrospective or prospective, beyond stating that "Baxter Healthcare Corporation conducts risk analyses and design verification tests."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable. The device is a physical medical component, not an AI/diagnostic software. Ground truth for its performance is established by objective engineering tests against international standards (ISO 594-1 and ISO 594-2), not expert consensus on interpretations.

    4. Adjudication Method

    This section is not applicable, as there is no expert interpretation or consensus-based ground truth for this type of device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This section is not applicable. The device is a physical medical component, not an AI system intended to assist human readers in interpreting cases. Therefore, a MRMC study comparing human readers with and without AI assistance was not performed.

    6. Standalone Performance Study

    This section is applicable in the sense that the device's performance was evaluated in a standalone manner (i.e., the physical device itself was subjected to engineering tests without human intervention in the performance of the device's function). The "Performance Data" section directly addresses the results of these standalone tests against ISO standards.

    7. Type of Ground Truth Used

    The ground truth used for performance validation was based on objective engineering standards ("Bench performance testing requirement according to ISO 594-1:1986" and "ISO 594-2:1986") and biocompatibility standards ("ISO 10993-1").

    8. Sample Size for the Training Set

    This section is not applicable. The device is a physical medical component and does not utilize a training set in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable, as there is no training set for this type of device.