Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100351

    Validate with FDA (Live)

    Date Cleared
    2010-03-16

    (32 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NEUROSCOUTTM Steerable Guidewires are intended for selective placement of microcatheters and other devices within the neuro and peripheral vasculature.

    Device Description

    The hydrophilically coated NEUROSCOUT 0.014 guidewires consist of a stainless steel wire core and a radiopaque platinum/tungsten coil on the distal tip. The basic principle of the guidewires is to act as a monorail that catheters can track over to reach a particular area of the neuro and peripheral vasculature. They have a nominal outside diameter of 0.014 inch and nominal overall length of up to 300 cm.

    AI/ML Overview

    The provided text describes the 510(k) summary for the NEUROSCOUT™ Steerable Guidewire. This is a medical device submission, and the content focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with acceptance criteria in the typical sense of AI/ML performance evaluation (e.g., sensitivity, specificity, AUC).

    Instead, the "acceptance criteria" here refer to meeting engineering performance standards and demonstrating safety and effectiveness comparable to existing devices. The "study" refers to a series of preclinical tests validating these physical and functional properties.

    Here's an analysis based on the provided text, addressing your questions where applicable and indicating when information is not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the conformance to FDA Guidance and the performance of the predicate devices. The reported device performance is that it met these criteria through preclinical testing.

    Acceptance Criterion (Implied)Reported Device Performance
    Mechanical Performance:
    Tensile StrengthTesting conducted; device performs according to description.
    Torque StrengthTesting conducted; device performs according to description.
    TorqueabilityTesting conducted; device performs according to description.
    Tip Flexibility/Linear StiffnessTesting conducted; device performs according to description.
    Kink RadiusTesting conducted; device performs according to description.
    Lubricity and Catheter CompatibilityTesting conducted; device performs according to description.
    Material/Coating Integrity:
    Coating IntegrityTesting conducted; device performs according to description.
    BiocompatibilityScreening confirmed modifications did not affect biocompatibility.
    Functionality:
    Simulated UseTesting conducted; device performs according to description.
    Adequately address user needs (via animal studies)Animal studies conducted to ensure user needs addressed.
    Overall Equivalence:
    Same intended use as predicate deviceConfirmed.
    Same operating principle as predicate deviceConfirmed.
    Same materials as predicate deviceConfirmed.
    Similar device dimensional specifications as predicate deviceConfirmed.
    Same shelf life and sterilization process as predicate deviceConfirmed.
    Similar manufacturing process as predicate deviceConfirmed.

    2. Sample Size Used for the Test Set and the Data Provenance

    This document does not describe an "AI/ML test set" in the conventional sense. The "test set" here refers to the guidewire samples used for various preclinical tests.

    • Sample Size for Test Set: Not specified in the document. The tests like tensile, torque, lubricity, etc., would have involved a sufficient number of device samples per test to generate statistically sound data, but the exact numbers are not provided.
    • Data Provenance: The preclinical testing was conducted by Codman & Shurtleff, Inc. This is prospective testing of the device itself, not data from patients or a specific country. Animal studies were also conducted, implying live subject data (animals), but specific information on the type of animal or location is not given.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable and not provided. The "ground truth" for this medical device submission involves objective physical and mechanical measurements, and direct observation of performance in simulated and animal use. It does not rely on expert interpretation of data like images for AI/ML models.


    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods like "2+1" or "3+1" are relevant for resolving discrepancies in expert interpretations (e.g., radiologists reading images). The preclinical tests described here yield objective measurements or observations of physical performance.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable and not provided. An MRMC study is relevant for evaluating the impact of AI on human reader performance, particularly in diagnostic imaging. The NEUROSCOUT Steerable Guidewire is a physical medical device, not an AI/ML product designed to assist human interpretation.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable and not provided. This device is a physical guidewire, not an algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance evaluation refers to:

    • Engineering Specifications/Standards: Performance measured against predefined acceptable ranges derived from regulatory guidance (FDA Guidance for Coronary and Cerebrovascular Guidewires, 1995) and the performance characteristics of the predicate devices.
    • Objective Measurements: Results from tests for tensile strength, torque strength, flexibility, kink radius, lubricity, coating integrity.
    • Observed Functional Performance: During simulated use and in animal studies to ensure user needs are met.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. There is no mention of a "training set" as this is not an AI/ML device. The device itself is manufactured based on design specifications; it is not "trained" on data.


    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable and not provided for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1