    Device Name
    METRO SOFT
    Date Cleared
    2010-10-28

    (274 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Metro Soft, sphere (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.

    The Metro Soft, toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.

    The Metro Soft, multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding .75 diopters and are presbyopic requiring add power of up to +4.00 diopters.

    The Metro Soft, multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eves and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add nower of up to +4.00 diopters.

    The Metro Soft, irregular cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.

    Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

    Device Description

    The Metro Soft Silicone Hydrogel Soft Contact Lenses are fabricated from efrofilcon A. which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

    The non-ionic lens material, (efrofilcon A) is a daily wear silicone hydrogel contact lens that is not surface treated and characterized by a high water content. The lens material is composed of silicone monomers cross linked with other monomers and optionally incorporates D&C Green 6 as an integrated handling tint. The lenses are made by lathe-cut for custom RX. It consists of 26% efrofilcon A and 74% water by weight when immersed in a buffered saline solution. The (efrofilcon A) name has been adopted by the United States Adopted Names Council (USAN).

    In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly bevond the limbus forming a colorless, transparent optical surface. The (efrofilcon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

    AI/ML Overview

    The provided text describes a 510(k) summary submission for a medical device, the Metro Soft Silicone Hydrogel Daily Wear Soft Contact Lens. This type of submission focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than establishing entirely new safety and effectiveness through extensive standalone clinical trials.

    Therefore, the requested information about acceptance criteria, a study proving device meets acceptance criteria, sample sizes for test/training sets, number/qualifications of experts, adjudication methods, MRMC studies, or specific ground truth for an AI/algorithm-based device is not applicable to this document. This document details a regulatory submission for a contact lens, which is a physical device, and not an AI or algorithm.

    Instead, the document outlines the following:

    1. Acceptance Criteria and Reported Device Performance:

    The "acceptance criteria" in this context are the characteristics of the device that demonstrate its substantial equivalence to predicate devices. The performance is reported by direct comparison of its physical properties and intended use.

    CharacteristicMetro Soft (efrofilcon A) - New DevicePredicate (IntelliWave3, efrofilcon A)Predicate (Metro Optics RevitalEyes, hioxifilcon B)Predicate (Coopervision Biofinity, comfilcon A)Predicate (Hydron Ltd. ActiFresh 400, lidofilcon A)
    Intended UseDaily wear for correction of ametropia, management of irregular corneal conditions.Same as new deviceSame as new deviceDaily wear for correction of ametropia.Daily wear for correction of ametropia.
    FunctionalitySame as predicate deviceFocus light rays on retinaFocus light rays on retinaFocus light rays on retinaFocus light rays on retina
    IndicationsDaily wear, Silicone Hydrogel SoftDaily wear, Silicone Hydrogel SoftDaily wear, SoftDaily wear, Silicone Hydrogel SoftDaily wear, Soft
    Production MethodLathe-Cut, custom manufacturedLathe-Cut, custom manufacturedLathe-Cut, custom manufacturedCast Molded, mass producedLathe-Cut, custom manufactured
    USAN Nameefrofilcon Aefrofilcon Ahioxifilcon Bcomfilcon Alidofilcon A
    Water Content74%74%48%48.0%74.0%
    Oxygen Permeability60 x 10-11 cm2/sec60 x 10-11 cm2/sec15.89 x 10-11 cm2/sec128 x 10-11 cm2/sec28 x 10-11 cm2/sec
    Specific Gravity1.1391.1391.1421.1421.060

    2. Study Proving Device Meets Acceptance Criteria:

    The study that "proves the device meets the acceptance criteria" in this context is the comparison demonstrating substantial equivalence to legally marketed predicate devices. The document explicitly states:

    • Clinical Data: "The clinical performance of the (efrofilcon A) lens material has been previously established, and therefore was not required for this 510(k)." This indicates that clinical studies were performed for the material in a prior submission (likely K100221 for IntelliWave3).
    • Non-clinical Testing: "A series of in vitro and in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the Metro Soft (efrofilcon A) Silicone Hydrogel Soft Contact Lenses... All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols."
    • Key Finding for Substantial Equivalence: "The Metro Soft, (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens is identical to the cleared Intelliwave3 (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens, cleared under K100221." And "The Metro Soft, (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lenses have the identical manufacturing process (lathe-cut versus lathe-cut) as the marketed Intelliwave3 (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens, cleared under K100221."

    No specific sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, or training set details are provided because this is a 510(k) submission for a physical medical device, not an AI/ML algorithm requiring such specific performance validation methods. The "ground truth" for this submission relies on established safety and performance data of the predicate devices and general standards for biocompatibility and material properties of contact lenses.