    K093916

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-08-23

    (244 days)

    Product Code
    Regulation Number
    862.1055
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neobase Non-derivatized MSMS reagent kit (for use on the PerkinElmer TQD MSMS Screening System) is intended for the measurement and evaluation of amino acids, succinylacetone, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. Quantitative analysis of these analytes (Table 1) and their relationship with each other is intended to provide analyte concentration profiles that may aid in screening newborns for metabolic disorders.

    Device Description

    The measurement of amino acids, succinylacetone, free carnitine, and acylcarnitines with the NeoBase assay involves extraction of dried blood spots from newborns with a solution containing stable-isotope labeled internal standards and analysis using a tandem mass spectrometry (MSMS) system. The each analyte relative to their response of internal stable-isotope labeled corresponding standard is proportional to analyte concentration.

    AI/ML Overview

    The device being described is the NeoBase Non-derivatized MSMS Kit, intended for the measurement and evaluation of amino acid, succinylacetone, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. This quantitative analysis aids in screening newborns for metabolic disorders.
    The study presented is a non-clinical study comparing the performance of the NeoBase Non-derivatized MSMS kit on the PerkinElmer TQD Triple Quadrupole Mass Spectrometer System (PerkinElmer TQD platform) against its predicate devices, the MS2 and PerkinElmer Quattro Micro platforms (QMicro). The goal was to demonstrate substantial equivalence.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as a set of predefined thresholds. Instead, it demonstrates equivalence to predicate devices by showing comparable performance characteristics. The key performance metrics evaluated were:

    • Precision (Imprecision Percent Coefficient of Variation - %CV): Lower %CV indicates higher precision.
    • Recovery (Mean % Recovery and 95% Confidence Interval): Indicates the accuracy of analyte measurement.
    • Measurable Ranges: The range over which the device can accurately quantify analytes, ensuring coverage of clinically significant levels.
    • Method Correlation (Ratio of Measured Concentration): Comparing the TQD platform with predicate devices (MS2 and QMicro). A ratio of 1.0 indicates equivalent concentration measurements.
    • Clinical Correlation (Percent Agreement in clinical determinations): How well the TQD platform agrees with predicate platforms in classifying samples above or below clinical cutoffs.
    • Detection of True Positive Samples: The ability of the device to correctly identify known positive cases.

    Here's a summary of the reported device performance, focusing on the TQD platform's comparison to predicate devices:

    Performance CharacteristicAcceptance Criteria (Implied: Comparable to Predicate)NeoBase Non-derivatized MSMS Kit (TQD Platform) Performance (as compared to MS2/QMicro)
    Precision (Average Total Imprecision %CV)Should be comparable to or better than predicate devices.Amino Acids: Generally around 10-18% for TQD, similar to or slightly better than MS2/QMicro (e.g., ALA 10, ARG 10, MET 18, TYR 8, VAL 12). Carnitines/Acylcarnitines: Not explicitly provided for all, but overall implied to be adequate based on predicate comparison.
    Recovery (Mean % Recovery)Should be comparable to predicate devices.Amino Acids: Ranges from 57% (SA) to 104% (C0) for TQD, generally comparable to MS2/QMicro. 95% CI also presented.
    Measurable RangesShould cover all clinically significant ranges.For all analytes, the TQD range (µM) includes or extends beyond the "Cutoff Range (µM)", demonstrating sufficiency for clinical use.
    Method Correlation (Mean Ratio of Measured Concentration)Ratios close to 1.0 (indicating statistical equivalence).MS2/TQD: Ratios ranged from 0.89 to 1.09, with small variation, indicating statistical equivalence. QMicro/TQD: Ratios ranged from 0.92 to 1.08, with small variation, indicating statistical equivalence.
    Clinical Correlation (% Agreement)High percentage agreement with predicate devices.All Analytes: Ranged from 99.2% to 100.0% agreement between TQD and MS2/Sciex platforms in clinical determinations.
    Detection of True Positive Samples100% agreement with predicate devices in detection.100% agreement for all 17 true positive samples (representing 14 disorders) between TQD and MS platforms.

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-clinical (Analytical) Test Set:
      • Precision and Recovery: The tables provided (5.3, 5.4, 5.5) show averaged data, but the explicit number of samples/replicates isn't detailed for each specific test. However, the method correlation section states that "enriched samples (five levels) was analyzed (as singlicates of each level) for 16 runs to provide a total of 80 individual measurements" for each analyte on each platform.
      • Method Correlation: Data from "dried blood spots enriched with the analytes of interest," specifically "5 levels times 5 runs per analyte" resulted in 25 means per platform for each analyte (80 individual measurements total, as above).
    • Clinical Test Set:
      • Clinical Correlation (Percent Agreement): 2499 random newborn screening specimens (presumptive negative data set) and 17 specimens with true positive diagnoses. Some analytes specify 2598 total observations (2499 + 80 individual measurements from enriched samples for other tests?), while others specify 2518* (2499 presumptive negatives + 19 true positives, including newly acquired NKH and H-ALA samples mentioned in footnote).
      • True Positive Samples: 17 samples with true positive diagnoses representing 14 disorders.
    • Data Provenance: The document does not specify the country of origin for the samples. It mentions "newborn heel prick blood samples dried on filter paper," which is a standard collection method. The data is retrospective in the sense that these were pre-existing biological samples used for evaluation. It's not a prospective collection of new patients for this specific study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This study focuses on diagnostic device performance (quantitative measurements of analytes) rather than interpretive tasks that would typically require human expert adjudication of images or complex clinical scenarios.

    • Analytical Performance (Precision, Recovery, Measurable Ranges, Method Correlation): Ground truth is established by the known concentrations of analytes in the spiked/enriched samples, and the quantitative measurements determined by specialized laboratory equipment (Mass Spectrometers). No human experts are involved in establishing this type of ground truth.
    • Clinical Correlation and True Positive Samples: The "ground truth" for the 17 true positive samples is referred to as "true positive diagnoses." The document does not specify how these diagnoses were established (e.g., whether by pathology, genetic testing, or clinical consensus) nor does it mention the number or qualifications of experts involved in these initial diagnoses.

    4. Adjudication Method for the Test Set

    Not applicable. This is an analytical/quantitative device performance study rather than an interpretive study requiring human adjudication. The "agreement" for clinical correlation refers to the concordance between the numerical results of the TQD platform and the predicate platforms against established clinical cutoffs, not human expert consensus on a diagnosis.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This study is evaluating the analytical performance and clinical correlation of a laboratory diagnostic assay, not a device that requires human interpretation of outputs.

    6. If a Standalone Study (algorithm only without human-in-the-loop performance) was done

    Yes, this is essentially a standalone (algorithm only) study. The "device" is a reagent kit used on an automated mass spectrometry system. The study compares the quantitative results generated by the TQD platform (with the NeoBase kit) directly to the predicate MS2 and QMicro platforms (also using the NeoBase kit) without a human-in-the-loop interpretation step being evaluated as part of the primary outcome for device clearance. The output is a numerical concentration.

    7. The Type of Ground Truth Used

    • Analytical Performance: Ground truth is based on known concentrations in control and enriched samples. These are prepared by spiking analytes into a matrix (dried blood spots) at specific, verifiable concentrations.
    • Clinical Correlation and True Positive Samples:
      • For the 2499 random newborn screening specimens, the "ground truth" for clinical "agreement" is whether the analyte concentrations fall above or below their respective clinical cutoffs (as determined by the predicate device).
      • For the 17 true positive samples, the ground truth is established clinical diagnoses of metabolic disorders ("true positive diagnoses").

    8. The Sample Size for the Training Set

    The document describes evaluation of the device performance, not the training of an AI algorithm requiring a specific "training set." This device is a reagent kit for a mass spectrometry system, not an AI/ML software device in the typical sense. Therefore, there is no explicit "training set" in the context of an AI model.

    Historically, the predicate devices (MS2 and QMicro) and their associated kits would have undergone extensive validation and optimization (which could be conceptually analogous to a training phase, but for analytical chemistry rather than AI). The current study is demonstrating the equivalence of the NeoBase kit on a new platform (TQD) to these previously validated systems.

    9. How the Ground Truth for the Training Set was Established

    As there is no distinct "training set" for an AI algorithm in this context, this question is not applicable. The methods used in developing and validating the NeoBase kit and its use on mass spectrometry platforms would involve many iterations of experiments to establish linearity, accuracy, precision, and other analytical specifications. This process relies on robust analytical chemistry principles and reference materials with known concentrations, rather than a "ground truth" established by human experts for AI training.

    Ask a Question